Needs version 1.1.3.8A. Step 1: Check Settings. „Enable SampleTrack“ is unchecked „Manually destination selection“ is checked „Clean ... “ is unchecked. Open procedure editor dialog. Two main procedures will be defined:

Download Presentation

Needs version 1.1.3.8A

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Needs version 1.1.3.8A

  2. Step 1: Check Settings • „Enable SampleTrack“ is unchecked • „Manually destination selection“ is checked • „Clean ... “ is unchecked

  3. Open procedure editor dialog Two main procedures will be defined: 1. Fill syringe with the total solvent amount 1200µl for 12 tubes. Name: „Load Syringe DMSO 1200.gsp“ 2. Dispense partial solvent amount 100µl into one destination. Name: „Dispense Volume 100.gsp “ (with sub procedure „Sub Disp Volume 100.gsp “)

  4. Step 2: Edit procedure „Load Syringe DMSO 1200.gsp“ 1. Check the file name„Dispense Volume 100.gsp “ 2. Select Source Type: „Solvent in Tray“. • 3. Select your own values for: • - Solvent name • Volume (Speed ...) • Hint: Never forget „Update“ and „Save“ after modifications.

  5. Step 3a: Edit dispense sub procedure „Sub Disp Volume 100.gsp “ Press the button „Edit“ to edit or create a procedure of the type „Dispense“.

  6. Step 3a: Edit dispense sub procedure „Sub Disp Volume 100.gsp “ Check the file name „Sub Disp Volume 100.gsp“. • Select your own values for: • Volume • Z offset height (Speed ...).

  7. Step 3b: Edit dispense main procedure „Dispense Volume 100.gsp “ Check the file name „Dispense Volume 100.gsp “ Select the Source Type: „Gas Gap“ Select your Dispense procedure. Volume = 0

  8. Step 4: Start Automation to fill sample loop • Select procedure • „Load Syringe DMSO 1200.gsp“. Start Automation. Hint: Solvent not used in procedure! 2. Select one dummy source and destination position (not used inside procedure!). 3. Create Order and continue with „OK“.

  9. Step 5: Create Orders to fill rack matrix. Open Dialog „Create new Orders“. Do NOT start automation! Syringe would be homed zero. Procedure finished OK: solvent filled in Sample loop.

  10. Step 6: Create Orders to fill rack matrix • Select procedure • „Dispense Volume 100.gsp“. Hint: Solvent not used in procedure! 2. Select 12 source and destination positions (destinations will be used). 3. Create Orders and close dialog with „OK“.

  11. Step 7: Start preparation to fill rack matrix Do NOT start automation! Syringe would be homed to zero. Use menu „Service Tools“. Sub menu: „Prepare all Orders“.

  12. Step 7: Running preparation