2103-390 ME Exp and Lab I

2103-390 ME Exp and Lab I. Basics for Physical Quantities and Measurement Physical Quantity Measured Quantity VS Derived Quantity Some Terminology Physical Principle/Relation of An Instrument and Sensor and Sensing Function f s Measurement System Model

donald
Download Presentation

2103-390 ME Exp and Lab I

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. 2103-390 ME Exp and Lab I • Basics for Physical Quantities and Measurement • Physical Quantity • Measured Quantity VS Derived Quantity • Some Terminology • Physical Principle/Relation of An Instrument and Sensor and Sensing Function fs • Measurement System Model • Input-Output relation: y = f ( x ; …) • Important: Identify MS, input x, and output y clearly

  2. Theoretical Input-Output Relation and Theoretical Sensitivity • Linear Instrument VS Non-Linear Instrument • Some Common Mechanical Measurements • Calibration • Static Calibration • Calibration Points and Calibration Curve • Calibration Process VS Measurement Process • Some Basic Instrument Parameters • Range and Span • Static Sensitivity K • Resolution • Some Common Practice in Indicating Instrument Errors

  3. Basics for Physical Quantities and Measurement

  4. Describing a physical quantity q • Dimension • Numerical value with respect to the unit of measure • Unit of measure Physical QuantityDescribing A Physical Quantity Physical quantity • A quantifiable/measurable attribute we assign to a particular characteristic of nature that we observe.

  5. Measured Quantity VS Derived Quantity The Determination of The Numerical Value of A Physical Quantity q must be either through • Measurement with an instrument Measured quantity or • Derived through a physical relation  Derived quantity (and by no other means) Because of existing physical relations/laws, we don’t want anybody to make up any number for a physical quantity.

  6. Some Terminology Measurement / Measure • The process of • quantifying, or • assigning a specific numerical value corresponding to a specific unit (of measure) to a physical quantity q of interest in a physical system. Measurand / Measured Variable • The physical quantity q that we want to measure, e.g., velocity, pressure, etc. Instrument / Measuring Instrument / Measurement System • The physical tool that we use for quantifying the measurand, e.g., thermometer, manometer, etc.

  7. Often, our desired physical quantity x – measurand – cannot be measured directly (in its own dimension and unit). • Physical Principle/Relation of An Instrument ( fs ) • and • Sensor and Sensing Function fs

  8. pb pressure at surface b Dh pa pressure at surface a rm Class Discussion What is the pressure difference (pa – pb)? Do we measure the pressure difference (pa – pb)directly in the unit of pa with a U-tube manometer?

  9. pb pressure at surface b Dh pa pressure at surface a Dh (pa – pb) rm Input measurand x Measurement System (MS) y = fs ( x ; …) (sensor stage) Output y • What is the pressure difference (pa – pb)? • We do not measure (pa – pb) directly. • Instead, we measure Dh. • Then, determine the desired measurand (pa – pb) from the physical principle/relation (from static fluid)

  10. Input measurand x Output y Dh (pa – pb) y Dh x (pa – pb) • Often, our desired physical quantity x – the measurand – cannot be measured directly (in its own dimension and unit). • We need to determine/derive its numerical value from • another physical quantity y, which is more easily measured, and • a physical relation/principle. Measurement System (MS) (sensor stage)

  11. pb pressure at surface b Dh pa pressure at surface a rm Physical Principle/Relation of An Instrument [and Sensing Function fs] The physical principle of a U-tube manometer is static fluid (fluid in static equilibrium). • Physical Principle of An Instrument [and Sensing Function fs] • The physical principle that allows us to determine the desired measurand x with dimension [x] in terms of another physical quantity ys with different dimension [ys]. • We refer to the underlying physical relation as sensing functionfs.

  12. pb pa Dh rm Principle: Fluid Static Dh (pa – pb) Input measurand x Measurement System (MS) Output y Physical Principle: Fluid Statics fsis the sensing function.

  13. Measurement System Model

  14. Sensor (or sensor-transducer) Output ys: ys = f (x;…) Measurement System / Instrument Process/System Signal Conditioning stage Output stage Sensor stage Transducer stage Signal path Input x Output y Control stage Thermometer Sensor/Transducer • employs physical phenomena • to sense the desired physical quantity x • in terms of another more easily measured quantity ys . Thermal expansion Temperature T Length L(scale)

  15. Measurement System Model Input x Sensor stage (Sensing element) Signal Modification Stage Output Stage Output y (Numerical value ) Measurand q in a physical system f f s m Measurement System Model • Physical Principle of The Instrument and Sensing Function ( fs ) • We shall refer to • the physical principle that allows the sensor to sense the desired measurand x with dimension [x] in terms of another physical quantity yswith different dimension [ys] as the physical principle of the instrument, and • the corresponding underlying physical relation as the sensing functionfs.

  16. Measurement System (MS) Input measurand x (physical quantity) Output y (physical quantity) • We are then interested inthe input-output relation Input - Output Relation: How to find the output-input relation - Theoretical - Actual  Static Calibration

  17. Input measurand x Measurement System (MS) Output y Important: Identify MS, input x, and output y clearly • When considering measurement system (or subsystem) characteristics • Measurement System: Identify the measurement system (MS) clearly (physically as well as functionally), from input x to output y. • Input Measurand x: Identify the physical quantity that is the input measurand x and its dimension/unit. • Output y: Identify the physical quantity that is the output y and its dimension/unit. • Calibration Curve: Find and draw the calibration curve ( y VS x ) for the system • Note: • It helps to identify the dimensions of the input and output physical quantities clearly. Is it length, pressure, velocity, or voltage, etc? • Recognize that if there is no output indicator, we cannot yet know the numerical value. • For example, the output of the pressure transducer is voltage output, but without a voltmeter or an output indicator, we cannot yet know the numerical value of this voltage output. • In this regard, e.g., when perform uncertainty analysis, the output indicator must be accounted for as part of the measurement system.

  18. Measurement System Output Input First-Order System: Measurement System Model: First-Order System • The ODE has the solution where the complementary solution is given by and is the time constant. • The particular solution depends on the input forcing function.

  19. = Normalized time 1/e = tn First-Order System: Step Forcing Input

  20. Theoretical Input-Output Relation and Theoretical Sensitivity

  21. pb pressure at surface b Define The Measurement System MS (Define the input and the output quantities clearly.) The Theoretical Input-Output Relation and Sensitivity Dh (output) Dh pa pressure at surface a pa –pb (input measurand) rm The input-output relation: Input measurand x ? Output y ? MS: U-Tube Manometer (pa –pb) [pressure] Dh [Length] The Theoretical Input-Output Relation and Theoretical Sensitivity forU-Tube Manometer

  22. slope • Linear Instrument • Output y is a linear function of input measurand x. • The slope K is constant throughout the range • General Non-Linear Instrument • Output y is not a linear function of input measurand x. • The slope K is not constant throughout the range Theoretical Input-Output Relation and Graph, andLinear VS Non-Linear Instrument

  23. Outputy Input measurandx Measurement System (MS) Sensitivity Sensitivity K • If K is large, small change in input produces large change in output. • The instrument can detect small change in input measurand more easily.

  24. Example 1: Define MS, Input x, Output y Clearly Note: Here, we define a measurement system in a more general term, based on the interested functional relation.

  25. + + 2 1 rm= density of manometer fluid ra= density of fluid a rb= density of fluid b pa = static pressure at center a pb = static pressure at center b p1 = static pressure at 1 p2 = static pressure at 2 ha = elevation at center a hb = elevation at center b h1 = elevation at free surface 1 h2 = elevation at free surface 2 Dh = h2 – h1

  26. Measurement System 1 (MS1) + + Input measurand x ? Input measurand x ? Output y ? Output y ? U-Tube Manometer U-Tube Manometer + + pa –pb [pressure] p1 –p2[pressure] Dh [Length] Dh [Length] Measurement System 2 (MS2) Example 1 Determine the theoretical input-output relations for the two measurement systems[See Appendix A for the derivation]

  27. Example 2: Redefine our measurement system for convenience

  28. + + 2 1 Measurement System 1 (MS1) h [Length] Dh [Length] pa –pb [pressure] pa –pb [pressure] U-Tube Manometer U-Tube Manometer + + 2 Measurement System 2 (MS2) Equilibrium position 1 Example 2Redefine our measurement system for convenience • MS1: It is not convenient to measure the change from the two interfaces. • We may redefine our output/system. • MS2: Here, it is more convenient to measure the change with respect to one stationary reference point.

  29. Example 3: Find the theoretical input-output relation and the theoretical sensitivity

  30. + + 2 (pa - pb) Dh Measurement System (MS) 1 The theoretical input-output relation is Example 3Find the theoretical input-output relation and the theoretical sensitivity

  31. Example 4: Theoretical sensitivity and how to increase sensitivity in the design of an instrument

  32. Example 4 How can we increase the sensitivity of the manometer?Differential Pressure Measurement - Inclined Manometer Principle: Static fluid Fox et al, 2010, Example Problem 3.2 pp. 59-61. • Appropriate sensitivity K can be chosen by changing d/D and sin q, e.g., • Smaller q Higher K

  33. Some Common Mechanical Measurements

  34. Some Common Mechanical Measurements • Temperature • Pressure • Velocity • Volume Flowrate • Displacement • Velocity • Acceleration • Force • Torque

  35. Example 5: Differential Pressure MeasurementInclined Manometer Principle: Fluid Static From Dwyer http://www.dwyer-inst.com/PDF_files/Priced/424_cat.pdf

  36. L = 0 L = 0 L (mmW) Input measurand x ? Output y ? Balance position: apply pa = pb Measure position: apply pa > pb Inclined Manometer Principle: Fluid Static pa –pb [pressure] (at the free surfaces) L [Length, mmW] From Dwyer http://www.dwyer-inst.com/PDF_files/Priced/424_cat.pdf

  37. Example 6: Differential Pressure MeasurementPressure Transducer: Capacitance Principle: Capacitance From Omega: http://www.omega.com/ppt/pptsc.asp?ref=PX653&ttID=PX653&Nav=

  38. Input measurand x ? Output y ? Pressure Transducer (alone) of the pressure transducer alone • Without output stage such as voltmeter or output indicator, however, we cannot yet know the numerical value of the output (voltage). • This is not yet a complete measurement system – no output stage. Pressure Transducer Principle: Capacitance pa –pb [pressure] (at the ports) V [Voltage, Vdc] From Omega: http://www.omega.com/ppt/pptsc.asp?ref=PX653&ttID=PX653&Nav=

  39. Input measurand x ? Output y ? Pressure Transducer + Output Indicator • Specification/Characteristics (of the measurement system, e.g., accuracy, etc.) • Need to take into account the characteristics (e.g., accuracy, etc.) of the output stage – i.e., output indicator – also. Pressure Transducer + Output Indicator Principle: Capacitance pa –pb [pressure] (at the ports) V [Voltage, Vdc] From Omega: http://www.omega.com/ppt/pptsc.asp?ref=PX653&ttID=PX653&Nav=

  40. Output Indicator (alone) From Omega http://www.omega.com/ppt/pptsc.asp?ref=DP24-E&Nav=

  41. Calibration • Static Calibration • Calibration Points and Calibration Curve • Some Basic Instrument Parameters

  42. CalibrationStatic Calibration Calibration is the act of applying a known/reference value of input to a measurement system. Objectives of Calibration Process • Determine the actual input-output relation of the instrument. • Quantify various performance parameters of the instrument, e.g., range, span, linearity, accuracy, etc. The known value used for the calibration is called the reference/standard. Static Calibration: • A calibration procedure in which the values of the variables involved remain constant. • That is, they do not change with time.

  43. Calibration points: yM(x) Reference: Known value Static calibration curve (fit): yC(x) If it is linear, yC(x) = Kx + b. Calibration Points and Calibration Curve • Calibration Points: We first have a set of calibration points. • Calibration Curve: For convenience in usage, we fit the curve through calibration points and use the fitted equation in measurement.

  44. Static calibration curve (fit): yC(x) Reference: Known value Reference: Known value Measurement process Calibration process Calibration Process VS Measurement Process

  45. ymax ro =ymax - ymin ymin ri = xmax - xmin xmax xmin Some Basic Instrument ParametersRange and Span Calibration Range: Input range: Output range: Span: Input span: Output span: Full - scale operating range (FSO) = Output span =

  46. Static calibration curve (fit): yC(x) If it is linear, yC(x) = Kx + b. Calibration points: yM(x) Some Basic Instrument ParametersStatic Sensitivity K Static Sensitivity: • is the slope of the static calibration curveyC(x)at that point. • In general, K = K(x). • If the calibration curve is linear, K = constant over the range.

  47. Indicated output displays are finite/digit Continuous t Some Basic Instrument ParametersResolution (DRy) [Output] Resolution (DRy) is the smallest physically indicated output division that the instrument displays or is marked. Note that • while the input may be continuous (e.g., temperature in the room), • the indicated output displays are finite/digit (e.g., ‘digital’ or numbered scale on the output display). DRy

  48. Output Resolution: DRy Linear static calibration curve: yC(x) = Kx + b. Indicated output displays are finite/digit, the input resolution is correspondingly considered finite. Some Basic Instrument ParametersResolution and Static Sensitivity KThe smallest change in input that an instrument can indicate. DRx: Input Resolution

  49. Some Common Practice in Indicating Instrument Errors

  50. Some Common Practice in Indicating Instrument Errors • Error e in % FSO: • Error e in % Reading: • Error e in [unit output]/[unit input]: