An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. A Thermo Electron Corporation új XSeriesII ICP-MS készülékének és környezetvédelmi alkalmazásainak bemutatása Száraz Sándor Unicam Magyarország Kft. 1144 Budapest, Kőszeg u. 29. unicam@unicam.hu

  2. XSeriesII ICP-MS Brief introduction to ICP-MS

  3. Characteristics of ICP-MS • Wide element coverage • Low limits of detection • Elemental and isotopic information • Wide dynamic range • Wide variety of samples

  4. 3. Ion Analysis Separation by m/z ratio 2. Ion Focusing / Optional Collision Cell 1. M+ Species Generated in the Plasma 4. Ion Detection by electron multiplier M+ M+ M+ Detector Quad Cell Ar Plasma Sample Gas ICP-MS Process • 4 Basic Stages 1. Sample Introduction and Ion Generation 2. Ion Focusing 3. Separation of Analyte Ions in Quadrupole Mass Filter 4. Ion Detection

  5. ICP-MS - Sample Introduction • Samples are normally analyzed as solutions • Sample is introduced into an argon plasma as a fine aerosol, via a peristaltic pump, nebulizer and spray chamber • Within the plasma (< 6000K) the solvent is evaporated and the sample species are decomposed into their constituent atoms and ionised - Ionisation process is extremely efficient in the plasma, and contributes to the high sensitivity of ICP-MS As the solution passes through the Ar plasma, 4 main processes will occur

  6. ICP-MS - Sample Introduction • Normally via concentric nebulizer ~1ml/min uptake • Typical sample volume - 20 elements using 3 x 2 secs. integrations - 4 - 5ml sample required • Optional low flow nebulizer for μl. sample analysis • Optional laser ablation for direct solids analysis

  7. MO+ M+ M++ Cooler Hotter ICP-MS - Sample Introduction • Maximize M+ • - Minimize MO+ • - Minimize M++ • Sample Uptake - Nebulizer dependent • Plasma Gases - Cool • Auxiliary • Nebulizer dependent • Torch Position - x, y, z control

  8. Ion Sampling and Focusing • Ions formed in the plasma are extracted through a sample and skimmer cone arrangement • Ion Lenses focus & optimize ion transmission to quadrupoleanalyser • Sample treatment is important to minimize deposition on cones:- - Dissolve and dilute to ~ 0.2% w/v - up to ~ 5% nitric acid preferred - up to ~ 5% hydrochloric acid - ~ 1% hydrofluoric acid (using an Inert Sample Introduction Kit), phosphoric acid or sulfuric acid Slide Valve Skimmer Cone Ion Lens Sample Cone DA Extraction Lens Increasing vacuum

  9. Ion Focusing Mechanism • Optimizes ion transmission to quadrupole analyser • Conventional ion focusing devices - Complex lens configuration and often incorporate a photon stop - Backgrounds typically 10-20 cps • XSeriesII: New high efficiency ion guide - Simple design with reduced number of lenses - Incorporate an innovative chicane deflector and off-axis quadrupole - Backgrounds of <0.5 cps - Upgradeable to collision cell technology (CCT)

  10. Ion Analysis – Quadrupole MS • Quadrupole consists of 4 rods mounted equidistant to each other around the circumference of a circle • Separates Ions according to their mass to charge (m/z) ratio • Principles of operation - Alternating RF / DC potentials are applied to the quadrupole rod pairs - Ions move in a spiral motion down the quadrupole axis - Majority of masses are put into an unstable trajectory and are rejected - However under a given condition, ions of a specific m/z will pass through the analyser and reach the detector

  11. Ion Detection • Ion detection is made with an electron multiplier • Measures the number of Ions arriving at the detector - proportional to the concentration of that isotope in the original solution • Detector operates in pulse counting (ion counting) and analogue (ion current) modes • With automatic cross calibration measures major, minor and trace analytes in a single acquisition -2Kv -2Kv

  12. Data Acquisition Modes • Scanning - Qualitative analysis and Quantitative analysis - SemiQuantitative analysis: Post run data retrieval on un-calibrated elements • Peak Jumping - Selected isotopes - Optimized dwell times - Improved detection limits • Time Resolved Analysis - Transient signal analysis - Applicable to laser ablation ICP-MS and chromatographic studies

  13. ICP-MS Full Mass Spectrum • Simple spectra (primary M+ ions) - Facilitates simple interpretation • Very high signal to background - Low detection limit capabilities V ICP-MS Spectra - Vanadium 10mg/L V ICP-AES Spectra - Vanadium 10mg/L

  14. Calibration Techniques • Semi Quantitative - Analyte sensitivity can be determined from the instrument response built from the response to other known analytes. • Relative Sensitivity Factors (RSF) apply corrections for ionization and sample processing effects, these can be pre-determined for a given matrix to improve accuracy • Full Quantitative - Multi element external calibration standards - Standard additions calibration - Isotope ratio determination (for isotope dilution measurement)

  15. Semi-Quantitative Analysis • External standards not required for each analyte • Calibration via RSFs from Response Curve • Full elemental and isotopic information • Major to trace concentration determined • No prior knowledge of sample required Semi Quantitative Calibration Plot

  16. Fully Quantitative Analysis External Calibration • External calibration standards measured prior to unknown samples • Element or isotope responses are plotted against concentration • Concentration of unknown samples calculated from the calibration graph Standard Additions • Known sample is spiked with known concentrations of analytes • Standard addition calibration plot provides accurate low level concentration data from the x-axis intercept • Eliminates need for matrix matched calibration standards

  17. As(III) DMA AsC + AsB As(V) MMA 8.1 6.0 10.3 3.8 Time (min) Time Resolved Analysis • Enables time resolved data to be acquired both qualitatively and quantitatively from transient signals • Wide applicability including Laser Ablation ICP-MS and chromatographic studies

  18. XSeriesII ICP-MS Routine Environmental Applications

  19. Environmental Applications – Sample types • Wide variety of environmental sample types • Drinking water • Ground water • Waste water / effluent • Leachates • Soil / contaminated land digests • Sediment digests • Plant / animal tissue digests • Generally high sample numbers (>100 per day in many laboratories)

  20. Enviromental applications:Myths and misconceptions in ICP-MS • Considered in the flame AA, furnace AA and ICP-OES communities that: • Not possible to measure high ppm concentrations (>100ppm) by ICP-MS • Interferences are a significant problem • Instrumentation is complex and difficult to use • Very high cost

  21. Memory effects vs wide dynamic range: Monitored Sample Uptake and Wash • Intelligent sample monitoring to ensure stability prior to data acquisition. • Monitored washout eliminates sample carry over.

  22. XSeriesII - Xt Interface • Optimized response for Environmental Analysis • Maximum productivity through long-term matrix tolerance and excellent signal stability • Minimum recalibrations between samples • Extended dynamicrange • EnablesLinear determination of Na,Mg,K,Ca to 200 mg/L • >> 1000mg/L when used in conjunction with high resolution measurement

  23. DC M-1 M M+1 RF Reaching high ppm concentrations:Variable resolution – what does it mean in practice? • Allows the user to extend the dynamic range even further • The dynamic range limitation then becomes the 0.2% total dissolved solids upper level imposed by the interface cones • Can be adjusted down to peak widths lower than 0.3 amu • Can be adjusted on-the-fly to allow standard and ‘high’ resolution measurements to be made analyte by analyte

  24. Na, Ca, calibration – Xt interface, standard resolution Na: blank to 300 ppm Cr: blank to 100 ppb Sensitivity = 0.9 Mcps/ppm BEC = 107 ppb Sensitivity = 1.3 Mcps/ppm BEC = 0.2 ppb Standard XSeriesII configuration (no CCT), using In as the internal standard Peak width (at 10% peak height) = 0.75 amu

  25. Reaching even higher ppm concentrations:Na calibration – Xt interface, increased resolution Na: blank to 2000 ppm 2000 ppm Sensitivity = 0.5 Mcps/ppm BEC = 38 ppb Standard XSeriesII configuration (no CCT), using Ga as the internal standard Peak width (at 10% peak height) = 0.2 amu

  26. Long Term Drift of X Series with Xt Interface (Borehole Water Matrix) Li = 21 µg l-1 RSD = 1.6 % In = 10 µg l-1 RSD = 1.1 % U = 4 µg l-1 RSD = 1.5 %

  27. Comparison with ICP-OES R2 = 0.984 R2 = 0.997 n = 188

  28. Conclusions for XSeriesII standard with environmental work • Hardware and software advancements of the XSeriesII ICP-MS allow rapid, accurate analyses of environmental samples with turnkey operation. • Great improvement in sample throughput. • Linear range extended and matrix tolerance improved with Xt interface. • Protocol compliance even without CCT.

  29. XSeriesII ICP-MS Performance Options for Environmental Research Applications

  30. Peltier Controller option – do we need it? • Rapid, precise temperature control of spray chamber for optimum performance - constant aerosol formation independent of the laboratory temperature control • Typical operation at 2oC - reduced solvent loading into the plasma - lower oxides and other interferences - improved MDLs • Can be used with the standard quartz or inert spray chamber • Rapid changeover between aqueous and organic matrices using the organics kit Variable temperature Peltier (thermo-electric) block

  31. CCTED – When do we need it? • Research Applications • Ultimate Detection Limits - well below regulatory levels • Analysis of some complex or pristine environmental matrices such as seawater or snow / ice. • Common interferences can not be eliminated using conventional quadrupole ICP-MS technology • Interference correction equations • Matrix removal • CCTED enables direct analysis without the need for matrix removal leading to improved DLs and enhanced productivity

  32. Introduction to the use of CCT with environmental analysis:Principles of Collision Cell Technology Iron Argon Oxygen Helium

  33. Introduction to the use of CCT with environmental analysis • Some analytes have few interference problems and perform best with the cell unpressurised, i.e. standard ICP-MS mode, e.g. Pb • The ideal multi-element analysis would result in several analyses under different conditions, e.g. • Standard mode • H2 mode • NH3 mode • A more efficient method is to switch modes “in-sample” • is this viable in terms of speed, stability and accuracy?

  34. Uses for CCT - Analytes, Interferents & Gases

  35. StandardMode CCTMode Spectra for UPW Water (35-65 amu)

  36. CCT for environmental applications: Instrument • XSeriesII ICP-MS with CCTED and Xt interface • CCT connected to two gases: • 8% H2 in He • 1% NH3 in He • 2 computer controlled MFCs allow gases to be changed “in-sample”

  37. CCT for environmental applications: Experimental - considerations Wide range of typical environmental analytes were measured: - 30 analytes, 55 isotopes Many analytes have associated interference problems….

  38. CCT for environmental applications:Experimental - Timing Li Be Na Rb Sr Rb Mo Ag Cd Sn Cs Ba Tl Pb U Mg Al K Ca Cr Fe Mn Ni Cu Zn Ga As Se V Cr Analytes Uptake 25s NH3/He 3x1.6s reps Settle Delay 30s H2/He 3x18s reps Settle Delay 30s Standard Mode 3x18s reps Wash 25s Time Profile Total Time Per Sample (3 repeats) = 3 minutes, 45 seconds

  39. CCT for environmental applications:Results - detection limits (mg/L) Based on 3s on 12x10replicates of blank, each from a new calibration

  40. CCT for environmental applications:Results – stability of real sample in different modes Sample diluted 1+9 and spiked to 2% HCl

  41. CCT for environmental applications:Conclusions • The XSeriesII allows rapid settings changes, allowing two different cell gases and normal ICP-MS mode to be used in a single sample analysis • 3¾ minutes per sample, 55 isotopes, 3 measurements/isotope, • Excellent stability is retained in each mode • RSDs of typically <1% over a 12 hour duration for a signal of ~50,000cps • Mode switching allows the optimum conditions for each analyte to be utilised, resulting in ultimate performance: • DLs in the ppt or sub-ppt range for almost all elements • The technique results in freedom from many interferences • Accuracy of within ±5% for the vast majority of analytes, even after spiking to 700ppm chloride

  42. Xs Interface option – do we need it? • Xs- extraction provides enhanced sensitivity whilst retaining the extremely low background characteristics of the XSeriesII • For research and ultra trace applications • Actinides • Isotope ratios • Small spot laser applications • Interchangeable with the Xi interface - User interchangeable ~ 2 minutes • Typical sensitivity >200Mcps/ppm for mid –high mass elements (In – U)

  43. X Series Environmental Methods ICP-MS Productivity Pack • Supplied to customers on instrument delivery: • PlasmaLab Productivity Method Template • Detailed instructions on instrument set-up, solution preparation and sample analysis • A Productivity Method Template that can be modified to a laboratory’s own working method • All calibration and quality control solutions required to run each protocol • A printed file containing the Productivity Method Template

  44. Environmental Protocols • US EPA has developed a series of methods for the analysis of drinking water, waste water and other environmental samples: - • OW 200.8 drinking water 1991 • OSW 6020 waters, wastes, soils, etc 1994 • OSW 6020A waters, wastes, soils, etc 1998 • CLP ILM05.2D waters, wastes, soils, etc 2001 These are challenging, QC intensive, multiple analyte, multiple concentration methods with tough specifications for accuracy, precision and DLs

  45. Environmental Protocols - Typical Requirements • Methods are ‘Prescriptive’ with specific rules on various analytical practices that must be followed e.g. • Specified Elements: • Up to 23 elements, 7-orders of magnitude range (ppt to high ppm) • QC Checks – post calibration: • Separate source calibration verification • Interference check - High Matrix • Interference check - High Matrix + Spike • QCChecks – every 10 samples: • Continuing calibration verification • Reference material • Detection limit check standard • Sample QCs - every 20 samples: • Duplicate • Serial Dilution and Spike Recovery

  46. XSeriesII - Interference Correction Equations X Series interference correction equations derived for use with EPA protocols

  47. XSeriesII MDLs vs. ILM05.3 CRQLs- EPA ILM 05.3 Environmental water and waste water

  48. XSeriesII Environmental Productivity Pack Supplied to customers on instrument delivery: • PlasmaLab Productivity Method Template on CD • Detailed instructions on instrument set-up, solution preparation and sample analysis • Productivity Method Template can be modified to a laboratory’s own SOP • All calibration and quality control solutions required to run each protocol • A printed file containing full instructions of the operating procedure Environmental Productivity Pack - Contents

  49. XSeriesIIICP-MS – Environmental Analysis Summary • Routine • MDLs - XSeriesIIICP-MS easily provides protocol compliant detection limits • QC checks: XSeriesII consistently produces accurate results on QCs and samples • Interference checks : XSeriesII design offers excellent freedom from interference and stable correction where necessary • Fastest sample throughput with protocol compliance • Productivity Pack uniquely offers: • Proven, reliable off-the-shelf EPA methods for the - XSeriesII ICP-MS • Unbeatable rapidity of start-up after installation • Research • Peltier cooled spray chamber for improved long term stability and MDLs • CCTED for interference removal / improved accuracy and detection limits in complex matrices • Xs interface for special applications requiring the ultimate signal/background • LC / GC packages for speciation studies