An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. CVD single crystal diamond detectors for UV and soft X-Ray plasma diagnostic at JET tokamak Presented by Maurizio Angelone Associazione EURATOM-ENEA sulla Fusione ENEA, C.R. Frascati

  2. Contributors Università di Roma “Tor Vergata”, (Italy) S. Almaviva M. Marinelli E. Milani G. Prestopino A. Tucciarone C. Verona G. Verona Rinati EFDA-JET, Culham, UK I. Coffey, A. Murari Ass. EURATOM/ENEA Frascati, Italy M. Angelone M. Pillon

  3. Outline • Motivation • SCD detectors • Device characterization • Results at JET • Conclusions

  4. UV and X diagnostic at JET • Plasma diagnostics: most relevant emissionsinextreme-UV and soft-X-ray spectral range. • Information on: • temperature • density • particle and energy confinement timescale • impurities dynamics • atomic collision rates • plasma-wall interaction • plasma dynamics • characterization of the electron fluid • atomic structure of highly ionized atoms Required features for detectors: • Radiation hardness • Low sensitivity to neutrons and -rays • Fast response • Visible blind • Good sensitivity in the EUV/soft-X spectral range

  5. Single Crystal Diamond detector Layered structure produced with two separate steps UV or X-Ray The intrinsic layer is 2-200 mm thick depending upon the use active area is about 6 mm2 Al or Pt contact electrode disc-shaped CVD intrinsic Ag contact JET DETECTORS • The Al contact is used for UV (100 nm thick). The intrinsic layer was 5 mm thick. • For X-rays (Al contact) the intrinsic layer was 22 mm thick A CVD B-dopd HPHT substrate • The diamond photodetector operated in transversal configuration. • The detector can work with no applied bias voltage (photovoltaic regime). • The photocurrent is measured between the p-type diamond layer and ground. No signal due to secondary electrons.

  6. Single Crystal Diamond detector The UV detector has been calibrated in the range 20 - 80 nm using He/Ne gas discharge radiation sources and a toroidal grating monochromator Responsivity to UV The X-rays detector has been absolute calibrated in the range 6-20 keV at the synchrotron facility Diamond Light Source, Harwell (UK) + Vbias Responsivity to X-ray X-ray Transmission

  7. Devices characterization: I-V curve Dark current The best signal to dark current ratio (SDR) performance is obtained at zero bias voltage The dark current is very low Our devices show a photocurrent response even at zero voltage bias, exploiting the internal junction electric field

  8. Devices characterization SCD X-rays - detector SCD UV - detector Temporal Linearity Vbias = 0 Volt He UV

  9. Devices characterization - UV He and He-Ne spectra measured by the SCD detector • All spectral lines are clearly observed with high signal to noise ratios, demonstrating the good photodetection capabilities of CVD single crystal diamond in the extreme UV spectral region. • The low intensity lines of the He spectrum in the wavelength range 22 – 27 nm are easily resolved.

  10. Measurements at JET Two CVD diamond detectors were permanently installed at JET in KS6 chamber and connected to the main on-line data acquisition system They are continuously recording the JET shots. The electronic consists in variable-gain low-noise current amplifiers FEMTO model DLPCA-200 and a 1040KHz sampling rate ADC VUV and X broadband spectroscopy – direct line of sight of the plasma

  11. The data acquired at JET have evidenced: fast response time, good signal to noise ratio, insensitivity to background radiation such as neutrons and gammas, good reliability during long lasting operation. Diamond results compared with JET standard UV (microchannel Plates) and X-ray detectors (silicon). The UV detector was used for measuring Edge Localized Modes (ELMs) activity as well as impurities influx during laser blow-off experiments The Soft-X detector seems promising for MHD studies. Improved performances (time resolution) by using a 40 KHz sampling rate ADC Measurements at JET: Results

  12. Measurements at JET: Results Neutron (in)sensitivity Shot #75057 Neutrons JET shot acquired by keeping closed the UV entrance shutter

  13. Measurements at JET : UV Shot #74355: Nickel ablation experiment

  14. Measurements at JET – X-ray ZOOM Shot #74663 Shot #74663

  15. Conclusions • CVD single crystal diamond UV and X detectors were permanently installed at Joint European Torus during the 2008-2009 experimental campaigns and connected to the main on-line data acquisition system. • The detectors can operate at zero bias voltage and show fast response and good signal to noise ratio. • First data acquired at JET have evidenced fast response time, good sensitivity, stability and extremely low sensitivity to neutron background. Analysis still in progress • Some JET research groups are currently using the data produced with diamond photodetectors to study the dynamic of impurities in the plasma as well as the ELMs.