An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Sector 9 Comparison • First measurements done on December 19, 2000 • flow - 1.2 l/m • inlet - 20.4C • front rtd 25.5C • rear rtd 25.3C • pwr - 8.7W/module • Irradiate at LLNL from December 21, 2000 to March 7, 2001 to 11 Mrads • Second measurements done on March 29, 2001 • flow - 1.2 l/m • inlet - 21.0C • front rtd 25.6C • rear rtd 26.2C • pwr - 8.7W/module

  2. Before After

  3. Smoke Test • Power at 8.7W per module(at room T). • No cooling. • See T rise plot. • Note power drops by 20% at max T, R change. • Image at max T • Turn on cooling and reimage both sides

  4. Before smoke test After smoke test