An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Analyze-Immediately Parameters pH

  2. Analyze Immediately Within 15 minutes of collection: • pH • Residual Chlorine • Temperature • Chlorine Dioxide • Dissolved Oxygen (probe) • Sulfite note sample collection & analysis times!!!

  3. Topics covered today: • pH • Temperature • Dissolved Oxygen (DO) • Residual Chlorine (TRC or CPO)….

  4. pH: what we will cover…. • Significance of pH. (data quality) • Points to Remember. • Lab Chemicals. • Good Lab Practices (GLP). • Review NJAC Regs. as they pertain to the recording & reporting of data.

  5. What exactly is pH??? • Negative log of Hydrogen ion activity. • pH above 7: alkaline or basic. • pH below 7: acidic. • pH = 7 is considered neutral. • Di-ionized water saturated with CO2 has pH = 5.65

  6. Data Quality : “a piece a cake!” • Properly working instrument & clean apparatus. • Fresh buffers and reagent rinse water. • Hold times met. • Correctly calibrated. • Proper sample handling. • GLP

  7. Data Quality: Calibration Grab: • Daily, within 3 hrs. of sample analysis. • 2 buffers, bracketed. • Mid-range check. • Record time, concentration and actual reading of each. • For initial calibration, reading must be +/- 0.05 of true value. • Recalibrate if difference is greater than +/- 0.2 (applies to 3 hr. check). • Discard pH buffer aliquots after each use.

  8. Data Quality: Calibration Continuous: • Weekly, unless otherwise stated by SA. • Indirect Calibration - basically a grab (as close to electrode as possible); measure within 15 minutes with lab pH meter; compare readings of continuous monitor with lab pH meter. Record info in logbook, sign & date. • Direct Calibration (as in grab procedure). - accurate to 0.1, per method (SM 4500-H B). • Record time, concentration and actual reading of each. • Discard pH buffer aliquots after each use.

  9. Make sure……. • Grab pH: meter in good working order (field or hand-held models vs.lab or bench-top models). • Continuous monitor: keep probe clean. • Calibrate as per method. • Maintain calibration & sample collection log. • Maintain field data notebook.

  10. Make Sure (continued)- • Discard buffers IMMEDIATELY following use - DO NOT re-use! • Discard expired buffers! • Electrode is OK - not damaged. • Stir or swish sample during measurement. • Record time of analysis.

  11. Make Sure (continued)- • Meters must have accuracy of +/- 0.05 and +/- 0.05 resolution • ...since pH is temperature-dependent • Must be temperature-compensated… • or use stand-alone thermometer…

  12. Lab Chemicals…... • FRESH buffers! (purchase ready-made or make yourself). • Reagent water. • Electrode filling solution (if necessary).

  13. Hold Times • Pertains to grab pH and indirect calibration for continuous measurement. • 0 - 15 minutes - note time of collection & time of analysis!

  14. Good Lab Practices (GLP) • Data signed, initialed & dated. • Instrument ID & serial number • NO pencil! • NO white-out! • Single-line, date & initial all errors.

  15. Analyze-Immediately parameters Records • NJAC 7:18-8.5 - This section pertains specifically to Analyze-Immediately parameters… • GENERAL: retain raw data records, quality control data records, chain-of-custody forms, laboratory reports… • Storage of data - data must be filed & maintained in an accessible location on lab premises for 1 year after analysis date… • Long-term storage: records of analysis must be stored for 5 years after analysis date.

  16. Retain as Records of Analysis • Assigned lab ID # or other form of ID • date & time of analysis • name & signature of person(s) who collected sample • name & signature of person(s) who analyzed sample • type of analysis performed & DSAM* used (*DSAM=Dept. Sanctioned Analytical Method) • results of analysis & raw data generated from analysis

  17. Reporting Results • Certified environmental lab name & NJ lab ID number • date, time & location of sample collection and sample analysis • type of analysis performed and analytical method employed • results from analysis • name & signature of environmental lab manager or designee per NJ regulations (NJAC 7:18-2.11(a)1iii)

  18. Summary • Don’t underestimate the importance of pH. • Adhere to NJAC Regulations and approved method. • Document-document-document!!!