1 / 7

Problems

D m = 0.0256 amu. Problems. The ratios e/m of 4 He + and D 2 + ions are very close:. m D 2 = 4.0282 amu m He = 4.0026 amu. A high resolution mass spectrometer should be required in order to separate the two peaks. D 2 concentration » expected 4 He concentration

franz
Download Presentation

Problems

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Dm= 0.0256 amu Problems • The ratios e/m of 4He+ and D2+ ions are very close: mD2 = 4.0282 amu mHe = 4.0026 amu • A high resolution mass spectrometer should be required in order to separate the two peaks. • D2 concentration » expected 4He concentration • Capability of on line analysis during the electrolytic experiment

  2. Conventional techniques • Dynamic analysis technique (differential pumping or diffusion through large impedance circuits): • alter the original composition of the gas mixture; • exhibit low sensitivity; • do not allow for on line analysis. • Static analysis technique (Cryosorption pumps): • may erratically trap helium in the condensates; • require a long cycle time and are not suitable for on line analysis. • Unable to pump hydrogen isotopes to high degree

  3. Innovative solutions • Non evaporable getter (NEG) pumps: • remove to a high degree all non inert components of the gas mixture (especially hydrogen isotopes); all inert gases are obviously preserved; • exhibit very high pumping speeds (especially for hydrogen); • the pumping speed for other active gases depends on the operating temperature of the getter alloy; • do not require frequent regeneration, allowing periodic analysis of gas samples during the experiment (roughly every 40 minutes). • A storage circuit has been purposely designed, which: • allows to periodically sample the gas mixture without influence on the electrolytic process; • eliminates any alteration of the gas mixture composition; • allows to easily compare the results of the analysis of each sample.

  4. High resolution Mass Spectrometer (Balzers QMA 410). Two NEG pumps are used: SAES Capacitorr B 1300 (600 g of ST185 TiV alloy) operating at high temperature (300 - 400°C) SAES GP200 MK4 W (170 g of ST707 alloy) operating at RT. Dynamic pumping: Main pump: Pfeiffer TMU 261 turbo molecular drag pump Backing pump: Pfeiffer TMU 071 + MD4 No cryosorption pumps are used. Automatic operation (Field-Point + LabView). Pressure stabilized storage circuit. Main features

  5. Gas mixture inert gases Thermostatic Box GV5 GV3 QMA UHV chamber n.o. n.c. HT NEG pump RT NEG pump Sample volume n.o. GV1 Storage circuit Cell UHV containment vessel n.o. = normally open valve n.c. = normally closed valve Schematic layout of the gas analysis system

  6. High resolution spectra of a 4He-D2 mixture. dwell time 1 s, scan width 10 amu, scan time for each cycle 10s.

  7. 6 40Ar+ 5 Ion current (x10-8 A) 4 3 2 40Ar++ 1 0 0 5 10 15 20 25 30 35 40 45 50 Mass (amu) Analog spectrum of an air sample

More Related