Training Program 15 January 2008. Contents. The chem-ID Mission Introduction Instrument Operation Identification Database/s Remote Operation User menu Cleaning Decontamination How to read the Graph Support. The Mission. What is a chem-ID?. The chem-ID is a field gas analyzer
E N D
Training Program 15 January 2008
Contents • The chem-ID Mission • Introduction • Instrument Operation • Identification Database/s • Remote Operation • User menu • Cleaning • Decontamination • How to read the Graph • Support
What is a chem-ID? The chem-ID is a field gas analyzer • Basically a sniffer • Analyzes gases, liquids with a vapor component, and solids with a vapor component • Core Technology: Two-dimensional gas chromatography with thermal conductivity detectors • Gas chromatography (GC) is the gold standard of chemical laboratories • Before now it has been limited to the laboratory or to field systems that require serious training and expertise • The chem-ID is designed to be easy to use with almost no training required
When would you use a chem-ID? • To analyze unknowns: • Tanker spill • Complaints of odors or people getting sick • Analyze unknown contents of containers, rooms, or vehicles
The chem-ID replaces many tests that require: $100,000 of equipment & 12 feet of bench space A trained analytical chemist
What the chem-ID does well: • Excellent at analyzing individual chemicals in complex clouds • Can analyze chemicals in extremely small amounts (parts-per-billion) • Can analyze and differentiate a large number of different chemicals • Intuitive and easy to use • Doesn’t require a trained chemist to operate • Don’t have to have a specially-trained team to operate and maintain • Allows testing in the field that traditionally required sending samples to the lab • Removes variations caused by high and low temperatures, humidity, wind, altitude
Technology Restrictions • Target chemical has to have a vapor component – does not do solids • Explosives (nitrate-based) typically have very little vapor signature • Very light gases not separated from air • Chlorine, helium, hydrogen, nitrogen, fluorine • Acids
chem-ID Features • Features: • 16 lbs • 16” x 8” x 9” • Performs an analysis by pushing one button • All analysis results are saved in non-volatile memory (won’t erase even if the battery dies) • Can be operated completely by remote control • Chemical analysis can be quickly emailed anywhere • Expanding database can be easily upgraded • Can be decontaminated
Sniffer Port Liquid Analyzer Kit Sniffer Nozzle
What does a chem-ID need to operate? Battery (7+ hours) Reference Chemical (2+ weeks) Helium Tank (2+ days)
Safety Issues • The chem-ID is not Intrinsically safe • Mil Spec 28.8V lithium Battery • Do not Store above 140ºF (60ºC), crush, mutilate, reverse polarity, disassemble, or dispose of in fire. • 2200 psi Helium tank • DoT approved for transport • Tank will change colors if exposed to high temperatures and weakened • Warning label on tank • Reference chemical (aka dopant) is 1,4 dichlorobenzene • Common • Present in very small amount • Wash hands after replacing
Consumables Location Battery Reference Chemical Helium Tank
Consumables: Battery • Battery • 7 hour battery life w/ continual testing, longer if idle or processing data • Rapid battery replacement is provided through the rear door • Rechargeable Li Ion • 14.4 VDC, 8.0 Ahrs • Manual shows installation in step-by-step pictures
Consumables: Helium • Ultrapure He Carrier Gas • Lasts for months in storage • When unit is operational 1 tank can last up to 24 hours of continuous operation • Rapid replacement is provided through the rear door • Can be ordered/ refilled through Laurus Systems • Manual shows installation in step-by-step pictures
Consumables: Reference Chemical • Reference Chemical (dopant) • Small dose in every sample • Used by the chem-ID to automatically calibrate variations of temperature and pressure • Easy replacement in the field • Manual shows installation in step-by-step pictures
Liquid Analyzer Kit Carrying Straps Spare Pre- concentrator CD: Software & Manual PC Bluetooth Radio Kit Battery Helium Tank Storage Reference chemical Recharger Carrying Case and Contents
Selecting Test Concentration • Determine the concentration level needed for the test: • Standard (default) or parts-per-billion (ppb) • Standard test level is sufficient in almost all situations • If you do a ppb test in a high concentration environment, the unit may need multiple manual clean cycles to finally clean out • Perform tests at standard concentration unless… • You are analyzing a chemical at ppb levels • A chemical presence is strongly suspected without any overt evidence (i.e. no liquids, no odors)
Checklist • Test Checklist • Turn on Unit • Open front and back vents • Open Helium valve • Check Helium gauge for at least 300 psi • Press Analyze button
1. Turn On Unit On/Off Rocker Switch
2. Open Vents Front Twist Knobs to Point at Open Back
3. Open Helium Valve Open Panel Twist Helium Knob all the way open (in the direction of the arrow on the knob)
4. Check Pressure Gauge Pressure Gauge Need 300 PSI or more
5. Press Analyze Button Button Labels Press Middle Button to Start Analysis
The chem-ID will tell you… • If the battery is too low to begin an analysis • If the helium is too low to begin an analysis • If you forget to open the front or back vents • If you need to add more reference chemical
What the unit is doing during Analyze • Automatically goes through the following phases: • Warm-up • Draws an air sample • Prepares the sample • Performs the analysis • Displays raw data in real-time • After analysis is complete, replaces raw data with processed data and shows identifications • Purges and cleans • Cool down to prepare for next test Overall Status Status of this Phase
Gas Chromatographs During Analyze, this graph will be slowly drawing Raw Data is not very useful, but shows the unit is busy After the raw graph is done, it will be processed and replaced with this graph This is the real chromatograph
Database • Early focus is on TICs • Expanding database: • More TICs • CWAs • Hydrocarbons • Common interferents • Meth Lab variant
Chemical Warfare Agents Hydrogen Cyanide (AC) Soman (GD) Lewisite (L) Cyanogen Chloride (CK) Mustard Gas (HD) Tabun (GA) Sarin (GB) Nit. Mustard (HN1,2,3) Interferents Antifreeze Clorox Kerosene Burning cardboard Diesel White vinegar Burning cotton Gasoline Windex Meth Manufacturing Chemicals a-benzyl-N-methylphenethylamine Dibenzyl ketone Nitroethane Acetic acid Diethylmalonate Phenyl-2-propanone Aluminum oxide Dimethylformamide Phenyl acetic acid Amphetamine Ephedrine phosphorous Benzyl chloride Ethyl acetate Sodium carbonate Benzyl cyanide Mercury Sodium hydroxide Benzyl methyl ketone benzylimine Methylamine Tri-(-phenylisopropyl) amine Di-(1-phenylisopropyl) amineN,N-dimethylamphetamine Future Library/Category Expansions
Identification Confidence • When identification of an unknown chemical shows several possibilities, the confidence bar shows which identification is closest to the database Unknown chemical could be Xylene or Ethyl Benzene Confidence Bar
What can be done with the chem-ID Manager? • Control the chem-ID remotely up to 100m away • Download test results to the PC • Study and compare test results on a PC (bigger screen, easier controls, more comfortable environment) • Convert chromatographs into CSV format (read by Microsoft Excel) • Update clock on chem-ID easily • Update firmware on chem-ID easily • Update identification database easily
You will need • A PC with Windows XP or Vista • In the chem-ID kit • chem-ID Manager Software • USB Bluetooth radio and installation CD • Installation • Bluetooth radio • Install as per the enclosed instructions • chem-ID Manager • Put CD in drive and open PC Manager directory
Turn on Bluetooth radio in the chem-ID • Under the Menu button: • Menu • Settings • Communication Port: Bluetooth Bluetooth Logo means the radio is on
First get Windows Bluetooth talking to chem-ID Bluetooth • 1. In Windows, select Bluetooth icon and tell it to search for Bluetooth devices nearby (or search Bluetooth neighborhood). • 2. Select chem-ID and tell it to connect
Windows Bluetooth talking to chem-ID Bluetooth cont. It will ask for a passkey. Passkey is always 00000000 (8 zeros) When the connection is established, Windows will tell you the port (com4 for example)
Connect to chem-ID Manager Connect Menu
Connected Connected
PC Manager Software Chemical Analysis Files on the PC Chemical Analysis Files on the chem-ID Remote Control Panel