An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. ME 322: InstrumentationLecture 24 March 24, 2014 Professor Miles Greiner

  2. Announcements/Reminders • This week: Lab 8 Discretely Sampled Signals • Next Week: Transient Temperature Measurements • HW 9 is due Monday • Midterm II, Wednesday, April 2, 2014 • Review Monday • Extra Credit Opportunity, Friday, April 18, 2014 • Introduction to LabVIEW and Computer-Based Measurements Hands-On Seminar • NI field engineer Glenn Manlongatwill walk through the LabVIEW development environment • 1% of grade extra credit for actively attending • Time, place and sign-up “soon”

  3. Transient Thermocouple Measurements • Can a the temperature of a thermocouple (or other temperature measurement device) accurately follow the temperature of a rapidly changing environment?

  4. Lab 9 Transient TC Response in Water and Air • Start with TC in room-temperature air • Measure its time-dependent temperature when it is plunged into boiling water, then room temperature air, then room-temperature water • Determine the heat transfer coefficients in the three environments, hBoiling, hAir, and hRTWater • Compare each h to the thermal conductivity of those environments (kAir or kWater)

  5. Dimensionless Temperature Error T Environment Temperature TF Initial Error EI = TF – TI Error = E = TF – T ≠ 0 TI TI t t = t0 • At time t = t0 a thermocouple at temperature TIis put into a fluid at temperature TF. • Error: E = TF – T • Theory for a lumped (uniform temperature) TC predicts: • Dimensionless Error: • (spherical thermocouple) TF T(t)

  6. Lab 9 Measured Thermocouple Temperature versus Time • From this chart, find • Times when TC is placed in Boiling Water, Air and RT Air (tB,tA,tR) • Temperatures of Boiling water (maximum) and Room (minimum) (TB, TR) • Thermocouple temperature responds more quickly in water than in air • However, slope does not exhibit a step change in each environment • Temperature of TC center does not response immediately • Transient time for TC center: tT ~ D2rc/kTC

  7. Type J Thermocouple Properties • State estimated diameter uncertainty, 10% or 20% of D • Thermocouple material properties (next slide) • Citation: A.J. Wheeler and A.R. Gangi, Introduction to Engineering Experimentation, 2nd Ed., Pearson Education Inc., 2004, page 431. • Best estimate: • Uncertainty: • tT ~ D2rc/kTC; = ?

  8. TC Wire Properties (App. B)

  9. Dimensionless Temperature Error • For boiling water environment, TF = TBoil, TI = TRoom • During what time range t1<t<t2does decay exponentially with time? • Once we find that, how do we find t?

  10. Data Transformation (trick) • Where , and b = -1/t are constants • Take natural log of both sides • Instead of plotting versus t, plot ln() versus t • Or, use log-scale on y-axis • During the time period when decays exponentially, this transformed data will look like a straight line

  11. To find decay constant b using Excel • Use curser to find beginning and end times for straight-line period • Add a new data set using those data • Use Excel to fit a y = Aebx to the selected data • This will give b = -1/t • Since t, • Calculate (power product?), ? • Assume uncertainty in b is small compared to other components • What does convection heat transfer coefficient depend on?

  12. Thermal Boundary Layer for Warm Sphere in Cool Fluid Thermal Boundary Layer • h increases as k increase and object sized decreases • = Dimensionless Nusselt Number TF T r D Conduction in Fluid

  13. Lab 9 Sample Data • http://wolfweb.unr.edu/homepage/greiner/teaching/MECH322Instrumentation/Labs/Lab%2009%20TransientTCResponse/LabIndex.htm • Plot T vs t • Find TBand TR • Calculate q and plot vs time on log scale • In Boiling Water, TI = TR, TF = TB • In Room Temperature air and water, TI = TB, TF = TR • Select regions that exhibit exponential decay • Find decay constant for those regions • Calculate h and wh for each environment • For each environment calculate • NuD • BiD

  14. Fig. 4 Dimensionless Temperature Error versus Time in Boiling Water • The dimensionless temperature error decreases with time and exhibits random variation when it is less than q < 0.05 • The q versus t curve is nearly straight on a log-linear scale during time t = 1.14 to 1.27 s. • The exponential decay constant during that time is b = -13.65 1/s.

  15. Fig. 5 Dimensionless Temperature Error versus Time t for Room Temperature Air and Water • The dimensionless temperature error decays exponentially during two time periods: • In air: t = 3.83 to 5.74 s with decay constant b = -0.3697 1/s, and • In room temperature water: t = 5.86 to 6.00s with decay constant b = -7.856 1/s.

  16. Lab 9 Results • Heat Transfer Coefficients vary by orders of magnitude • Water environments have much higher h than air • Similar to kFluid • Nusselt numbers are more dependent on flow conditions (steady versus moving) than environment composition

  17. Air and Water Thermal ConductivitiesAppendix B • kAir (TRoom) • kwater(TRoom, TBoiling)

  18. Lab 9 Extra Credit • Measure time-dependent heat transfer rate Q(t) to/from the TC (when TC is placed into boiling water) • 1st Law • Differentiation time step • Sampling time step • Integer m • What is the best value of m?

  19. Measurement Results • Choice of DtDis a compromise between eliminating noise and responsiveness

  20. What Do We Expect? Expected for Uniform Temperature TC TF = TB Ti = TR Measured t0

  21. What do we measure? Expected for uniform temperature TC Q Measured tT t ti

  22. Sinusoidally-Varying Environment Temperature • For example, a TC in a car exhaust line • Eventually the TC will have • The same average temperature and unsteady frequency as the environment temperature • However, its unsteady amplitude will be less than the environment temperature’s. TTC TENV

  23. Heat Transfer from Fluid to TC Fluid Temp TF(t) Q =hA(T – T) • Environment Temperature: TE = M + Asin(wt) • Divided by hA and • Let the TC time constant be (for sphere) • 1st order, linear differential equation, non-homogeneous T D=2r

  24. Solution • Solution has two parts • Homogeneous and non-homogeneous (particular): • T = TH+TP • Homogeneous solution • Solution: • Decays with time, not important as t∞ • Particular Solution to whole equation • Assume ) • ) • Plug into non-homogeneous differential equation to find constants C, D and E

  25. Particular Solution • Plug in assumed solution form: • Collect terms: • + • Find C, D and E in terms of A and M • C = M =0 =0 =0

  26. Result • ) • where

  27. Compare to Environment Temperature tD • Same mean value • If 1 >> = • Then • Minimal attenuation and phase lag • Otherwise • ( T

  28. Example • A car engine runs at f = 1000 rpm. A type J thermocouple with D = 0.1 mm is placed in one of its cylinders. How high must the convection coefficient be so that ATC = 0.5 AENV? • If the combustion gases may be assumed to have the properties of air at 600C, what is the required Nusselt number?

  29. Can we measure time-dependent heat transfer rate, Q vs. t, to/from the TC?1st Law Differential time step

  30. Measurement Results • Choice of dtD is a compromise between eliminating noise and responsiveness

  31. Lab 9 Transient Thermocouple Response T Environment Temperature TF Faster Slower TC Initial Error EI = TF – TI Error = E = TF – T ≠ 0 TI TI t t = t0 • At time t = t0 a small thermocouple at initial temperature TI is put into boiling water at temperature TF. • How fast can the TC respond to this step change in its environment temperature? • What causes the TC temperature to change? • What affects the time it takes to reach TF? TF T(t)

  32. VI

  33. Lab 9

  34. Initial Transient Time

  35. Transform Trick Small Same for air & water

  36. Lab 9 Find h in: Boiling Water Room Temper Air and water Why does h vary so much in different environments? Water, Air What does h depend on? T TF T r D NuD≡ Nusselt number