Calibration of a Yankee Environmental Systems UltraViolet Rotating Shadow Band Radiometer at Mauna Loa Hawaii

This presentation discusses the calibration of a UV rotating shadow band radiometer at Mauna Loa Observatory in Hawaii using satellite-derived UV spectrum. The importance of UV, health effects, agriculture, and energy deposition are covered. Methods for measuring UV include satellites and ground stations. Langley analysis is used to determine top of atmosphere values. The daily values are compared to MFRSR spectral response function and SORCE satellite measured spectrum. The results show improved calibration using Langley analysis. Acknowledgements to NASA, Langley, PMOD, and U.S. Coast Guard.

gracied
Download Presentation

Calibration of a Yankee Environmental Systems UltraViolet Rotating Shadow Band Radiometer at Mauna Loa Hawaii

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Calibration of a Yankee Environmental Systems UltraViolet Rotating Shadow Band Radiometer at Mauna Loa Hawaii Presented at the 10th International Pyrheliometer Intercomparison in Davos, Switzerland September 26 to October 14 2005, by Fred Denn Edited by: Jay Madigan

  2. Why is UV important? Health effects Agriculture Energy deposition in the atmosphere How to measure? Satellites Ground stations

  3. Satellite derived top of atmosphere spectral measurements are used to determine global and temporal UV surface irradiance and atmospheric energy deposition. Surface measurements are made, to verify the results of the atmospheric radiative transport models, requiring well calibrated surface instrumentation. A method of calibrating UV-MFRSRs using satellite derived UV spectrum is presented.

  4. The UV-MFRSR is located at the Mauna Loa Observatory, Hawaii for a period of time each year because of the exceptionally clear and stable atmosphere. Langley analysis are performed, on the morning data, to determine top of atmosphere values. Mauna Loa observatory Hawaii

  5. Langley plots were used to select morning data periods used in later analysis. To be considered good the standard deviation of the residuals must be less than 0.006

  6. The daily values were plotted for the entire time period, means and standard deviations were determined. These means will be compared to values derived using the MFRSR spectral response and the SOlar Radiation and Climate Experiment (SORCE) satellite measured extraterrestrial UV spectrum.

  7. Overview of Method Use a spectral line lamp to verify functionality of a monochromator. Determine the normalized functional response of the UV-MFRSR head. Perform a numerical integration using the UV-MFRSR spectral response function and a satellite derived UV spectrum.

  8. Top of Atmosphere MFRSR Measured Values and the Values Determined From the MFRSR Spectral Response Function and the SORCE Spectrum are Compared.

  9. TOA values are determined, and scale factors used to adjust the MFRSR derived values to more true values. Normalized to 1 au Wavelength, nm Scale factor ±U95% TOA, W/m2 ±U95% 300.2 1.11 23.7 0.405 16.8 305.6 0.94 9.62 0.547 6.8 311.7 1.04 7.92 0.641 5.6 317.8 1.11 8.63 0.677 6.1 325.4 1.09 7.21 0.824 5.1 332.5 1.26 8.49 0.947 6.0 368.1 1.05 7.21 1.169 5.1

  10. Verification of New Values Corrected Wavelength SIM derived TOA UVMFR-7 TOA Difference nm W/m2 W/m2 W/m2 ------- ------- ------- ------- 300.2 0.408 0.405 0.003 305.6 0.547 0.547 0.000 311.8 0.642 0.641 0.001 317.8 0.678 0.677 0.001 325.4 0.825 0.824 0.001 332.5 0.948 0.947 0.001 368.1 1.171 1.169 0.002

  11. Conclusion Manufacturer's calibration can be improved by applying Langley analysis to the output values supplied by there software.

  12. Acknowledgements NASA Langley PMOD U.S. Coast Guard