An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Operational Training on HPGe Detection System Bruker Baltic HPGe and MiniMCA-166 Training

  2. The HPGe Spectrometer • Based on: • Portable HPGe detector (Bruker Baltic) • Multi Channel Analyzer: MiniMCA-166 • Notebook • Software: separate programs to record and evaluate spectra • WinSPECi – detector setup and acquiring spectra • Identify – isotope identification • MGA – burn up evaluation of Plutonium • MGAU – enrichment of Uranium

  3. Get started for Gamma Spectroscopy • Step 1: fill detector • Step 2: connect detector with MMCA and PC • Step 3: switch on MMCA • Step 4: start software WinSPEC for Inspectors (WinSPECi) • Step 5: select or create setup file

  4. Step 1: cooling detector • Replace screw from detector with funnel • Put detector in upright position • Fill the detector (several shots) with liquid nitrogen (LN2) • Detector is full if drain releases LN2 • Wait 2 hours until detector is cooled down completely • Detector is cool for several hours • Replace funnel with screw (wait until layer of ice is gone) 1st fill 2 hours No topping Top (2nd fill) Cool for 10 hours Cool for 20 hours

  5. Step 1: cooling detector LN2, several shots Funnel Detector in upright position Drain with LN2 dropping out

  6. Step 1: cooling detector Hint to increase the resolution Fix the screw (safety valve) Turn detector for 180 degrees, remain for 5 -10 seconds and turn back

  7. Step 2: connecting equipment • Set of cables to connect HPGe’s Preamplifier with MMCA • High Voltage cable (HV), type SHV • Signal cable (OUT), type BNC • Preamplifier power cable (DC12V), type serial RS232 • Signal test cable (TEST), type BNC (not connected) OUT DC12V HV

  8. Step 3: switch on MMCA • Connect MMCA with the USB cable to the PC • Switch on MMCA – safety switch with lift function • Green LED is blinking in a high frequency • Detectors green LED should be on – indicates that detector is cool

  9. Step 4: start WinSPECi software • Open program WinSPECi • Three selections before program starts • Customized preferences (select no) • Connecting MMCA to serial port* (select search and offered COM port) • Retry to connect MCA (select yes) * Physical address of the USB port

  10. Step 4: start WinSPECi software

  11. Step 4: start WinSPECi software

  12. Step 5: detector setup SETUP Create setup file Open stored setup file Open spectrum recorded with the system Use old setup, stored in MMCA’s memory:DO NOT RESET MMCA at the start sequence

  13. Step 5: create setup • To work with the system the following parameters have to be implemented • Detectors High Voltage (see manual) • Measurement time • MCA setup: amount of channels, thresholds • Amplifier: polarity, shaping time, pole zero cancellation, gain adjustment • Energy calibration to change the x-axis from channels to Energy [keV]

  14. Step 5: create setup a) High Voltage for detector • Menu\setup\high voltage • High voltage: up to 3000 V • Polarity depending on hardware (HV-module) • Inhibit signal if interruption hardware is used (select Canberra HPGe)

  15. Step 5: create setup b) Preset for measurement time • Menu\Setup\Presets • Measurement time in Real Time, Live Time or None • Measures until certain area or integral value Select like here

  16. Step 5: create setup c) Multi Cannel Analyzer • Menu\Setup\MCA • Channels 128 – 4096 • Threshold: set to 0 % • Upper Level Discriminator (ULD) and Lower Level Discriminator (LLD) – don’t change Select like here

  17. Step 5: create setup Step by step 1.) Shaping time: 2μs 2.) Input polarity: positive 3.) Pole Zero Cancellation: • Select PZC adjustmen • Place source in front of detector • Select Automatic PZC by offset minimization 4) Gain adjustment • Select visual amplifier adjustment 4. 1. 2. 3. 3a.

  18. Step 5: create setup Gain adjustment: • Select the energy range for your purposes • Standard: 3000 keV with 4096 channels • Select a high energetic source, like Co-60 • Select visual amplifier adjustment and place source in front of detector • Calculate position for 1332 keV Co-60 peak: 4096 channel = 3000 keV ? channel = 1332 keV  1818 • Move cursor to the position in channel 1818 • Adjust gain (coarse and fine) until recorded Co-60 peak and cursor overlapping, quit with OK (F11)

  19. Step 5: create setup Fine adjust with F4 or F5 – until overlap of peak and cursor Peak is far to the right – press F2 several time

  20. Step 5: create setup • Energy calibration (fast method): X-axis is now in channel units, but we need energy units [keV] • Record a spectrum of the Co-60 source • Enter Menu/Analysis/Energy calibration and select “channel/energy pair input” and select ” Do energy calibration” • Channel 1: 0, Energy 0 • Channel 2: 1818, Energy 1332

  21. Step 5: create setup Co-60 spectrum X-axis in channels

  22. Step 5: create setup

  23. Step 5: create setup • energy calibration with 2 sources will follow – it is more complicated by exacter

  24. Error – high count rate • The red LED on the detectors preamplifier gives signal if the count rate is so high that the detector cannot work properly – move back from source until red light goes off • In case of bad vacuum in the detector the humidity inside raises and leads to a lot of signal (like high count rate) when detector fired with HV • Try to work with less HV • Send it for repair