1 / 26

Supplemental O2 Respiration Device

Supplemental O2 Respiration Device. Team members Advisor Jarrod long dr. Dorin BOLDOR Melissa wason Jim presley Shreya purohit. Sponsors. Mr. Schwartz

halden
Download Presentation

Supplemental O2 Respiration Device

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Supplemental O2 Respiration Device Team members Advisor Jarrod long dr. Dorin BOLDOR Melissa wason Jim presley Shreyapurohit

  2. Sponsors • Mr. Schwartz ( Superior Offshore International) • Mr. Ed Collins • Louisiana State University College of Agriculture

  3. Overview • Problem statement • Background • Design • Testing • Conclusion

  4. Problem Statement • To design a single use, portable oxygen dispenser that releases oxygen for 5 minutes at 6 liters/min

  5. Background Information

  6. Design • Cylinder • Regulator • Retention Bag • Mask • Case

  7. Cylinder • Cylinder & Retaining Valve • Manufacturer: Cliff Impact Division of Catalina Cylinders • Light weight, aluminum high pressure gas cylinder • Contains 42 liters of compressed oxygen gas at max pressure of 2000 psi • Cylinder specifications: • 2.5 inch diameter, 5.5 inch height, weighs 0.5 lbs

  8. Cylinder • Custom fit brass retaining valve • Manufactured by InterMed Gas Products Corporation • Model: CGA 870 (oxygen grade) • 1 inch diameter, 3.5 inch height, weighs 1.75 lbs

  9. Regulator • Brass-Sleeved 50 psi oxygen regulator • Manufactured by Inter Med Gas Products Corporation • Model: CGA 870, variable flow: 0-8 L/min • 1.5 inch diameter, 3.5 inch length

  10. Retention Bag • Clear, inflatable plastic bag (oxygen grade) • 8 inch height x 4 inch width (flat/no air) • Direct Inflow from oxygen cylinder • One-way valve exit to mask • Connections: oxygen grade, clear, flexible, plastic tubing

  11. Mask • Clear, flexible plastic • Adjustable aluminum nose piece • Fit with 2 custom one-way valves • Valve Functions: • Creates vacuum upon inhalation (opens one-way valve on retention bag) • Allows entrance through oxygen tube • Allows escape of CO2 during exhalation

  12. Case • Light weight aluminum: prototype • Future: plastic • Main function: • Ergonomics and aesthetics • Easily portable: shoulder straps, compact case • Contains vital parts to design, prevents loosening of connections

  13. Case • Other functions: • Prevent direct contact of cylinder, valve, or regulator with ground upon impact. • Increases safety of using the device

  14. Regulator One-Way Valve Case Retention Bag Mask Cylinder One Way Valves Completed Design

  15. Testing • Sensor Calibration • Respiration • Retention Bag • Valves • Flow Rate as Pressure Changes

  16. Percent Oxygen Inside Mask • Objective • 50-70% Oxygen • Oxygen Sensor • Pasco Industries • PasPort Interface • Data Studio • Oxygen specific • 0 – 100% O2 • Precision of ± 1% • Resolution of 0.024%.

  17. Oxygen vs. Atmospheric Air • Objective: prove oxygen sensor accuracy and % oxygen content in cylinder

  18. Respiration:Oxygen vs. Atmospheric Air • Objective: • Determine if oxygen reaches mask • Determine percent oxygen available to user • Hypothesis Testing • % oxygen from cylinder > % oxygen in air • 30 to 40% is lower than our ideal % oxygen

  19. Alterations • Valves • Bag

  20. With Retention Bag vs. Without Retention Bag • Objective: determine whether or not the retention bag increased the percentage of oxygen available

  21. Retention Bag on Mask vs. in Case • Objective: determine whether or not placing the bag inside the case decreased the percentage of oxygen available • Hypothesis Testing • Slight Difference • Acceptable for aesthetic reasons

  22. Valves • Objective: determine best valve configuration

  23. Flow Rate as Pressure Changes • Objective: Flow Rate must stay consistent – 6 L/min • Dependency of % O2 consumption rates • Concerns: • Initial pressure = 2000 psi • Final pressure = 0 psi • Flow Meter • Dwyer Instruments • Oxygen flow meter • Scale: 1.0-10 LPM

  24. Flow Meter Results • Objective: to determine if the regulator produces the required 6 L/min flow rate for 5 minutes

  25. Conclusion

  26. Questions? ?