Peter Mortier Stent cell sizes: Important data for bifurcation stent selection (EuroIntervention, Accepted for publication)
During provisional stenting, these cells are often enlarged by balloon dilatation Courtesy El-Jack et al
In a perfect world, all stent struts should be in contact with the arterial tissue • Why? • To reduce the risk of stent thrombosis • To facilitate subsequent stent insertion in the side branch • To reduce the disturbance of the blood flow • To optimize drug delivery • If the stent cells are too small, this is not possible!
New method to determine cell size based on in-house developed micro-CT imaging facilities Cypher stent (Cordis, J&J) Equivalent diameter
Overview of investigated stents Endeavor (Medtronic) PRO-Kinetic (Biotronik) Promus (Boston Scientific) Taxus Liberté (Boston Scientific)
Cell sizes differ considerably All stents have a nominal diameter of 3 mm * The platform of the Promus DES is identical to the platform of the Xience V (Abbott Vascular) DES
Size of the ostium 90° 45°
Ostium circumference calculation • The circumference of the ostium (red line) can be calculated in the idealized case of two intersection cylindrical tubes • The ostium circumference depends on three parameters: • Main branch diameter • Side branch diameter • Angle of intersection
The ostium circumference increases rapidly for smaller bifurcation angles • For a 3 mm main branch 3 different side branch diameters
Combining information of the ostium and stent cell size • Stent cell circumference should preferably be as large as the ostium circumference • Applying this criterion does not guarantee that all struts will be in contact with the tissue! • But, it helps to select stents which have at least the potential to be sufficiently enlarged
Cypher stent • For a 3 mm main branch
Promus/Taxus Liberté • For a 3 mm main branch
Limitations • Real-life bifurcations have a much more complex geometry • The criterion is too strict in case of stents partially deploying within the side branch • Stent selection depends on a large number of aspects and cell size is just one of them (e.g. closed cell design may provide better scaffolding)
Model results correspond well with data provided by the manufacturer Data manufacturer Simulation results
Simulated stent shape corresponds with experimentally observed shape
These computer models can also be used to study bifurcation stenting Mortier et al, Medical Engineering and Physics, accepted for publication
These models can accurately predict stent distortion Multi-Link VisionTM stent 2.5 mm balloon MultiLinkTM stent 3.0 mm balloon Ormiston et al., 1999
Conclusions • Cell sizes • Stent cell sizes differ considerably • The proposed criterion helps to select stents which have cells that can be sufficiently enlarged • However, this does not guarantee optimal strut positioning • Virtual tools • Computer models can accurately predict stent deployment/distortion • These models may help to investigate different bifurcation stenting techniques • Such models provide a platform to design and to optimize dedicated bifurcation stents