GC Triple Quadrupole MS: A Versatile Tool for POPs

GC Triple Quadrupole MS: A Versatile Tool for POPs. David Steiniger Senior Applications Scientist April 11, 2013. POP’s analysis. POPs analysis consists of:. • Polychlorinated dibenzo -p-dioxins (PCCDs) • Polychlorinated dibenzofurans (PCDFs) • Polychlorinated biphenyls (PCBs)

hall
Download Presentation

GC Triple Quadrupole MS: A Versatile Tool for POPs

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. GC Triple Quadrupole MS: A Versatile Tool for POPs David Steiniger Senior Applications Scientist April 11, 2013

  2. POP’s analysis

  3. POPs analysis consists of: • Polychlorinated dibenzo-p-dioxins (PCCDs) • Polychlorinated dibenzofurans (PCDFs) • Polychlorinated biphenyls (PCBs) • Toxaphenes • Organochlorine pesticides (OCPs) • Perfluorinated compounds • Brominated flame retardants (BFRs) • Chlorinated paraffins

  4. GC-MS triple quad portfolio • Thermo Scientific™ TSQ 8000™ GC/MS • High performance “workhorse” in routine complex matrix and for method consolidation • Thermo Scientific TSQ Quantum™ XLS Ultra GC/MS • Ultimate performance for high end applications • Emerging contaminants

  5. GC-MS/MS – What’s so special? • Low detection limits • Optimized sample preparation • Consolidated analytical methods • Faster, automated data processing ...it’s a high selectivity technique...

  6. Inside the TSQ 8000 3rd Quadrupole “Q3” 1st Quadrupole “Q1” Ion Source 2nd Quadrupole “Q2” Collision Cell Dynode and Electron Multiplier Ion Guide

  7. Step 1: Precursor ion selection Selected analyte precursor ion Matrix ion with same mass number

  8. Step 1: Precursor ion selection Acibenzolar-S-Methyl – 10 ppb in Lettuce Significant contribution from matrix ion causes peak area to be too high

  9. Step 2: Fragmentation in collision cell Fragmentation of both analyte and matrix ion

  10. Step 3: Product ion selection Selection of unique product ion of analyte

  11. Step 3: Product ion selection m/z 182 m/z 153 SRM removes the shoulder from matrix interference & reduces background SRM on m/z 182 > 153 SIM on m/z 182

  12. PCDD/Fs environmental samples • Presented is the determination of PCDD/Fs in sediments, soils, bottom and fly ash (as incineration by-products) using GC-MS/MS. • During this study, instrumental LOQs using GC-MS/MS were calculated in the low fg/µl concentration ranges. This and further analytical performance is discussed alongside GC-HRMS; especially degree of agreement between the techniques in some routine sample batches.

  13. Extraction and clean-up • The standard spiking protocols, and the extraction and clean-up process for all sediment and soil samples were performed using an in-house variation of EPA 1613B. For incinerator ash samples an in-house method based upon EN 1948 was applied. • For both of these methods accelerated solvent extraction (ASE) was used as the preferred technique for extraction after initial sample processing.

  14. PCDD/Fs GC-MS/MS GC parameters

  15. GC-MS/MS measurement • TSQ Quantum XLS Ultra GC-MS/MS

  16. MS/MS SRMs monitored

  17. PCDD EPA1613 CS1 (diluted 10x) 2,3,7,8-TCDD 100fg 1,2,3,7,8 PeCDD 500fg 1,2,3,7,8,9-HxCDD 1,2,3,4,7,8-HxCDD 500fg 1,2,3,6,7,8-HxCDD 1,2,3,4,6,7,8-HpCDD 500fg OCDD 1000fg

  18. EPA 1613 CS1 diluted x 10 tcdd ions

  19. TSQ Quantum XLS UltraLOD study on EPA 1613 CS1 dilutions

  20. TCDD in different environmental samples 321.9>258.9 321.9>258.9 321.9>258.9 2,3,7,8-TCDD 2,3,7,8-TCDD 2,3,7,8-TCDD 319.9>256.9 319.9>256.9 319.9>256.9 Fly Ash Contaminated Soil Contaminated Sediment

  21. GC-MS/MS mean deviation in QC results (3 batches)

  22. GC-MS/MS - comparison of calculated results for soils/sediment

  23. Sensitivity driven by precision Fivefold repeated injection of same sample extract of mixed animal fat 319.9 > 256.9 100 fg 2,3,7,8-TCDD on column • Mean : 0.13 pg/g fat • CV: 12.5 % 321.9 > 258.9 13C TCDD 1,2,3,4 & 2,3,7,8

  24. Solid ion ratios for high confidence QC limits around the theoretical ion abundance ratios: ± 15 % • HyperQuad ion ratio performance against theoretical values for dioxins • Solid performance well within EU guidelines • This in combination with excellent instrument sensitivity allows good quality WHO-TEQ values to be determined.

  25. TSQ Quantum XLS Ultra – food dioxin screening • Courtesy Alexander Kotz, EURL Freiburg, Germany • Presented at the 6th POPs Users Meeting, Niagara-on-the-Lake, Canada, Apr 28/29, 2011

  26. SCCPs chromatogram in full scan mode Short Chain Chlorinated Paraffinsdo not elute in separate peaks, but as a range of isomers The calculation is performed by integrating a time range, not integrating individual peaks.

  27. Linearity and peak at lowest level of calibration curve R2= 0.9956 At 0.5 ng the peak is still abundant in area and can be integrated easily

  28. Repeatability of lab control standard

  29. Mixed halogenated dioxins • Routine monitoring not yet in place • 2,3,7,8 substituted congeners are suggested to be even more toxic than their chlorinated analogues • In order to understand sources, fate, and distribution in the environment; low level analytical techniques must be utilized

  30. Mixed halogenated dioxins – method overview • Soil samples were solvent extracted using toluene • The TSQ Quantum XLS Ultra GC-MS/MS system was installed with a TR-PCB 8MS (60 m × 0.25 mm) • Injector SSL: 280 ⁰C • Flow: 1.2 ml/min • Injection volume: 1 µL

  31. PXDD/Fs at 2.4 pg/µL in soil extract

  32. PXDD/Fs at 2.4 pg/µL in soil extract Good low level precision in matrix with good LOD (1 µL on column)

  33. Flame Retardants – Curve and Peak

  34. House dust sample in NCI (above) and EI-SRM (below) showing BDE47 at a level of 0.07 ppb

  35. House dust sample in NCI (above) and EI-SRM (below) showing BDE183 at a level of 0.17 ppb

  36. Calibration curves of various PAHs

  37. Calibration curves of various OCPs & PCB 153

  38. Peak shapes – OCP, PAH, PCB 2 pg on column

  39. %RSDs in various matrices

  40. In conclusion • Triple quad is a great tool for a wide range of compounds • Provides improved selectivity over other instrumentation • Better signal to noise ratios – lower detection limits • Eliminates matrix interferences through SRM • Combine several methods into one run

  41. Acknowledgements • Thermo Scientific Team: • Inge de Dobbeleer • Paul Silcock • Joachim Gummersbach • Dwain Cardona • Valued Collaborators: • Chris Hunter, John Fardon and David Gardner, UK Environment Agency • Eric Reiner, Ministry of Environment, Canada