Wavelength Standards for IR Spectrographs at ESO Supporting high resolution spectroscopy Florian Kerber, ESO
Wavelength Calibration - Near IR • Night Sky OH Meinel Bands (Abrams et al., 1994; Rousselot et al. 2000) • Present on science exposure • Number of lines at given wavelength and resolution • Variability • Stability at high resolution ? (< 25 m s-1) • Lack of good sources • Lamps - Ne, Kr • Gas cells ESAC, March 22nd, 2007
Th-Ar hollow cathode lamp Anode Kerber et al. 2007, ASP Conf. Series. 364, p.461 Cathode ESAC, March 22nd, 2007
Hollow cathode lamp • Introduced by Paschen (1916) • Study of hyperfine structure • Atomic Absorption (AA) Spectroscopy (Walsh 1950s) • Lifetime - Getter • Quantitative chemical analysis • Great variety of materials • Mass production for commercial market ESAC, March 22nd, 2007
Th-Ar lamp:Visible and Near IR • Why Thorium ? • Nature: 232Th - no isotope splitting • zero nuclear spin - no hyperfine structure • Established standard source in Visual • Palmer & Engleman (1983) 278 - 1000 nm • FEROS, FLAMES, HARPS, UVES, X-shooter ESAC, March 22nd, 2007
CRIRES • Cryogenic High Resolution InfraRed Echelle Spectrometer (CRIRES) at VLT • 950 - 5000 nm, • Resolution / 100,000 • Project to establish wavelength standards (NIST) ESAC, March 22nd, 2007
Th-Ar Lamp Project • Physical Modelling Group (ESO INS) • F. Kerber, P. Bristow • NIST Atomic Spectroscopy Group • J. Reader, G. Nave, C.J. Sansonetti • Space Telescope European Coordinating Facility (ST-ECF) • M.R. Rosa • ESO La Silla-Paranal Observatory • G. Lo Curto ESAC, March 22nd, 2007
Th-Ar lamp:Visible and Near IR • Th-Ar is work in progress for near-IR • Hinkle et al. (2001), 1000 - 2500 nm, 500 lines • Gaps in wavelength coverage • FTS and grating spectrometer data • Engleman et al. (2003), 1000 - 5000 nm, > 5000 lines • High current source, rich Th spectrum • Differences in spectrum but … ESAC, March 22nd, 2007
Visible/IR FTS at NIST ESAC, March 22nd, 2007
NIST Experimental Work • 2-m UV/Visible/IR FTS • CaF2 beamsplitter, silver coated mirrors, InSb detectors • Two different settings optimized for wavelength • 800 - 2000 nm, resolution 0.01 cm-1 • > 2000 nm, resolution 0.005 cm-1 • Data acquisition times: up to 20 hours • Radiometric standard lamp ESAC, March 22nd, 2007
Th-Ar ESAC, March 22nd, 2007
CRIRES FTS ESAC, March 22nd, 2007
Th-Ar at NIST: Results • Total ~2500 lines between 750 nm and 4800 nm • accuracy ~ 0.001 cm-1 for strong lines • Absolute accuracy in wavelength: ~1.4 x 10-8 • calibration using laser measurements of Th lines (DeGraffenreid & Sansonetti 2002) • 5 x 10-8 1/100 of a pixel in CRIRES • Engleman et al. (2003) and our work agree: • Th: (6 ± 2) x 10-9 (weighted mean) • Ar: (-2 ± 0.2) x 10-8 ESAC, March 22nd, 2007
Results • The absolute accuracy in wavelength is about 1.4 10-8 • calibration using laser measurements of Th lines (DeGraffenreid & Sansonetti 2002, JOSA, B19, 1715) • Engleman et al (2003) and our work agree: • Th: 6 x 10-9 (weighted mean) • Ar: 2 x 10-8 • CRIRES wavelength calibration: 5 x 10-8 corresponds to ~1/100 of a pixel • Th-Ar lamp is able to support CRIRES science ESAC, March 22nd, 2007
At ESO Additional Measurements FTS Parabolic Mirror Integrating Sphere Elliptical Mirror Sample Compartment ESAC, March 22nd, 2007
Spectrum - Operating Current ESAC, March 22nd, 2007
Gas Cells - Status and next Steps • CRIRES commissioning: efficient use of the N2O gas cells (l=18 cm) as calibration source: ~2200 - 4100 nm • primary calibration molecule from Heterodyne frequency measurements at NIST (Maki & Wells) • Fundamental mode: 523 – 2845 cm -1 (1912 – 3515 nm) • Characterize spectrum of the CRIRES gas cell (P,T) Q2 2007 • optimal choice of fill gas pressure - measurements using ESO’s FTS • Wavelength Standards for first overtone transitions of N2O • Possible use of OCS (carbonyl sulfide) as calibrator (Q2 07) • separate gas cell or in a mixture with N2O • Long-term Stability of spectrum (Q4 2007) ESAC, March 22nd, 2007
Gas Cells - N2O ESAC, March 22nd, 2007
Summary: CRIRES Calibration • Calibration Reference Data • Established Wavelength Standards Th-Ar: 950 - 2200 nm, ~5000 lines; N2O: 2200 - 4100 nm • Laboratory “ground truth” Accuracy few parts in 10-8; ≤ 1/100 pixel • Properties of Sources • Optimize use: Th-Ar - current; N2O - gas pressure • Paradigma for wavelength calibration in near IR • very similar to the UV-visible region ESAC, March 22nd, 2007
Outlook • CRIRES Science Verification: • Nissen et al. astro-ph/0702689: Sulphur abundances • CRIRES operational: April 1st, 2007 • ESO archive • Spectral Data Model - Simple Line Access Protocol • Publish IR data set for Th: VOTable - metadata • Q2/Q3 2007 • linking astronomical and atomic physics communities ESAC, March 22nd, 2007