Electron Microscopy: Lecture 1: Introduction to the Transmission Electron Microscope (TEM). School of Life Sciences, University of Sussex. MSc Imaging in Biomedical Research, October 18 th 2011. Julian Thorpe The Sussex Centre for Advanced Microscopy. Why use electron microscopes (EMs)?.
E N D
Electron Microscopy:Lecture 1: Introductionto the Transmission Electron Microscope (TEM) School of Life Sciences, University of Sussex MSc Imaging in Biomedical Research, October 18th 2011 Julian Thorpe The Sussex Centre for Advanced Microscopy
Why use electron microscopes (EMs)? • Transmission electron microscopes • (TEMs) utilise electrons as • their source of illumination • which gives much improved resolution over a light microscope (around a thousand-fold better: c. 0.2nm compared with 0.2mm) • this is mainly because the effective wavelengths • of accelerated electrons are extremely shorter • than those of light
Transmission EMs As the name suggests, the electron beam is transmitted through the sample (normally a thin section of tissue or cells or a particulate sample such as viruses or proteins) so that the fine structure of the specimen may be observed (e.g. cellularultrastructure)
An electron beam has wave-like properties In 1923 de Broglie showed that an electron beam has wave-like properties…… ……thus pointing the way forward to the possible development of electron microscopes
Early development of EMs 1928-1931: Knoll and Ruska, in Berlin, began development of electron lenses and built a prototype EM 1937: Metropolitan Vickers Company (Manchester, UK) supply first commercial EM(to Louis Martin at Imperial College, London), but its resolution was no better than that of a LM Late 1930s: a resolution of about 7nm is achieved
Early development of EMs 1948-1953:The ultramicrotome was developed, allowing cutting of ultrathin (60-100nm) sections This was important, as electrons have limited energy and cannot pass through sections of more than a few hundred nm (except for high voltage TEMs) Gas or water molecules would also obstruct the passage of electrons down the ‘column’ of the TEM, thus they operate under a high vacuum
Effective wavelengths in the TEM l= (1.5/V)1/2nm where V = the accelerating voltage of the electron beam Effective wavelength decreases with increased accelerating voltage
Resolution in the TEM Resolution (nm) = 0.61 X l/N.A. where N.A. = the numerical aperture of the objective lens Resolution improves with increased accelerating voltage (and associated shorter effective wavelength)
Resolution in a ‘standard’ TEM vs LM Comparing a ‘standard’ TEM of 100kV accelerating voltage with a light microscope using UV illumination and optimal objective lens numerical aperture Although the wavelength of the illumination source in the TEM is 5 orders of magnitude shorter, numerical apertures of LM lenses are much greater
Life Sciences TEM: Hitachi-7100 (<125kV; resolution = 0.204nm)
A Million Volt TEM (resolution = 0.07nm)
3 Million Volt Hitachi: the most powerful TEM ever made (resolution = 0.04nm)
3 Million Volt Hitachi: the most powerful TEM ever made (resolution = 0.04nm) Operator
The way TEMs are going? HT7700 120 kV biomedical TEM from Hitachi 100% integration of all functions into the graphical user interface http://www.youtube.com/watch?v=h6VkvseFkzQ
Resolution in a ‘standard’ 100kV TEM Optical Microscope resolution c. 200nm TEM resolution c. 0.2nm
Resolution in the TEM 500nm
Resolution in the TEM 200nm
Resolution in the TEM 100nm
The electron beam source Theelectron beam is routinely derived from a thin hairpin filament of tungsten wire housed in a gun assembly A high accelerating voltage is used to boil electrons off the tip of the tungsten wire by thermionic emission and these are fired down the column of the EM and focused by electromagnetic lenses
The electron source (Cathode Gun Assembly) Figure c/o: http://en.wikipedia.org/wiki/Transmission_electron_microscopy
The electron source A high accelerating voltage is supplied to the filament Figure c/o: http://en.wikipedia.org/wiki/Transmission_electron_microscopy
The electron source Electrons are boiled off the tip of the filament by ‘thermionic emission’ Figure c/o: http://en.wikipedia.org/wiki/Transmission_electron_microscopy
The electron source The ‘Wehnelt cylinder’ has a higher –ve charge than the filament Figure c/o: http://en.wikipedia.org/wiki/Transmission_electron_microscopy
The electron source The ‘Wehnelt cylinder’ has a higher –ve charge than the filament……. and thus focuses the electrons Figure c/o: http://en.wikipedia.org/wiki/Transmission_electron_microscopy
The electron source Electronsare attracted to the positively charged ‘anode plate’ and pass through an aperture within it Figure c/o: http://en.wikipedia.org/wiki/Transmission_electron_microscopy
Transmission EMs The TEM ‘column’ stands vertically with a cathode gun assembly at the top housing the tungsten filament Electrons are boiled off the tip of this filament by ‘thermionic emission’ when a high voltage is applied Cross-Section through the Column of a TEM (side view)
Transmission EMs Beneath this is an anode plate, to which electrons are attracted and an aperture allows their passage down the TEM column Cross-Section through the Column of a TEM (side view)
Transmission EMs The TEM column is maintained under a high vacuum as electrons have insufficient energy to pass through gas and water molecules Cross-Section through the Column of a TEM (side view) This vacuum is achieved usually via oil diffusion pumps, backed up by rotary pumps
Transmission EMs Cross-Section through the Column of a TEM (side view) A series of lead-shrouded and water-cooled electromagnetic lenses make up the bulk of the TEM column
Transmission EMs Cross-Section through the Column of a TEM (side view) Condenser lenses condense and focus the electrons onto the area of the specimen being examined
Transmission EMs Cross-Section through the Column of a TEM (side view) An objective lens surrounding the specimen insertion area primarily focuses and initially magnifies the image
Transmission EMs Cross-Section through the Column of a TEM (side view) Intermediate and projector lenses magnify and project the focused image onto the fluorescent screen (converts electrons to photons) at the base of the column or to a CCDcamera beneath that
Transmission EMs Electromagnetic lens defects are similar to those of optical lenses… • …and these detract from achievement of the maximum • theoretical resolution. They are: • Spherical aberration: • Electrons passing through the lens periphery are • refracted more than those passing along the lens axis • and therefore do not have the same focal point. • Apertures are used in the TEM to limit the peripheral • electrons and minimise this aberration
Transmission EMs Electromagnetic lens defects are similar to those of optical lenses… • 2. Chromatic aberration: electrons of different energies • converge at different focal points and this is essentially • equivalent to chromatic aberration in light microscopy • This can be minimised by: • increasing the accelerating voltage • an improved vacuum • use of the thinnest possible specimen
Transmission EMs Electromagnetic lens defects are similar to those of optical lenses… 3. Astigmatism: occurs when the field within the electromagnetic lens is not perfectly symmetrical. Can be due to imperfect boring of the lens polepieces or contamination of the column, specimen or apertures .....TEMs have astigmatism controls to correct for this
Preparation of biological samples for TEM • Fixation: ‘Greater care’ needed for samples prepared for TEM, owing to the improved resolution and the higher magnifications possible • Samples have to be dry* (as the TEM operates under a high vacuum): therefore samples are dehydrated (* the exception to this are frozen-hydrated samples viewed by cryo-TEM) • Samples have to be ultrathin: this is because of the limited energy of the electron beam • Therefore: • special resins designed to allow cutting of ultrathin (c.50-100nm) sections are used to infiltrate and ‘embed’ samples • Or small particulates may be viewed (after air-drying)
Preparation of biological samples for TEM • Fixation: Normally a double-fixation in buffered glutaraldehyde (c.2-5%; cross-links proteins) and subsequently osmium tetroxide (1%; imparts electron-density to lipidic components) • Dehydrate: in an ethanol series • Resin embedding: infiltration with epoxy resin for a few days and heat-polymerised • Thin sectioning: must be ‘ultrathin’ to allow the electron beam to transmit through the section
Thin sectioning/’Ultramicrotomy’ • Sections of c. 60-100nm are cut on an ultramicrotome and collected on TEM support ‘grids’ <<< 3mm >>>
Staining to Achieve Contrast Contrast in thin sections examined in the TEM is facilitated by the use of heavy metals that can occlude and absorb electrons. These are routinely: • Osmium tetroxide: when used as a secondary fixative imparts electron-density to the lipidic component, especially membranes • Uranyl acetate and lead citrate: are used as ‘post-stains’. The former binds nucleic acids and proteins and the latter subsequently enhances the contrast
‘Negative staining’ of particulate samples • A very simple but effective method to examine particulate samples at high resolution • A drop of the sample is aliquotted onto a coated TEM support grid • A drop of heavy metal stain (e.g. uranyl acetate) is then dropped onto the sample and allowed to dry down around it • When viewed under the TEM the sample appears to be negatively-stained as the heavy metal creates an electron-dense background
Life Sciences TEM: Hitachi-7100 (<125kV; resolution = 0.204nm) High voltage supply cable Electron gun region TEM column Specimen airlock Viewing screen area plus binoculars
Hitachi-7100 TEM • Specifications: • Accelerating voltage range: 25 – 125kV • Magnification ranges: • 50 – 1,000X in ‘low mag’ mode • 1,000 – 600,000X in ‘zoom’ mode • Resolution: 0.204nm (lattice)/0.45nm (particle) • Motorized and tilting specimen stage with ‘memorise’ • and ‘relocate’ specimen positions facilities
TEM image formation • Image and contrast formation results from electrons • that are: • non-transmitted • (occluded by the heavy metal stains) • scattered • (elastic and inelastic) • unscattered • (transmitted)
TEM image formation Elastically scattered electrons (by the nucleus of an atom) contribute mostly to image contrast
TEM image formation Inelastically scattered electrons are concentrated within smaller scattering angles
TEM image formation Unscattered (or transmitted) electrons will pass through the specimen to form the electron-lucent regions of the image
TEM image formation • Higher accelerating voltages result in: • Increased electron speed and a concomitant decrease • in the incidence of inelastic scattering • So, although resolution is improved (because of the • shorter effective wavelengths),contrast is lowered • This can be redressed by the use of smaller apertures • in the objective lens, but at the expense of some lowering • of the resolution
TEM image formation Condenser lens aperture Objective lens aperture Or CCD camera
TEM imaging CCD camera