An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. CAPE Technologies High Performance Dioxin/Furan Immunoassay Kit Application Note AN-008 Rapid Extraction, Cleanup, and EIA Analysis of Soils at Low to Mid-pg/g Levels Illustrated Guide to Lab Setup, Sample Preparation, and EIA Protocol Revision March 31, 2005

  2. This is a training presentation only. It is designed to illustrate the materials, supplies, and procedures involved in the modification of EPA Method 4025 described in CAPE Technologies Application Note AN-008. It does not provide comprehensive protocol information. For detailed information on equipment, supplies, and protocol, refer to the appropriate CAPE Technologies document: AN-008 for sample preparation protocol IN-DF1 for immunoassay kit protocol EL-001 for equipment used for immunoassay EL-002 for equipment used for sample preparation The following pages contain references to specific sections of these 4 documents so you can connect the illustration to the detailed information from that document.

  3. Facilities Required Performing this method using the CAPE Technologies DF1 Immunoassay Kit requires a modest laboratory facility with capacity for 4-8°C storage of EIA kits. This facility can be a small mobile or stationary field laboratory. Electrical refrigeration, a sink, and running water are helpful, but not absolutely required. A fume hood is required for the sample preparation work, which uses hexane, acetone, and toluene.

  4. Equipment Required for Sample Preparation(not available from CAPE Technologies; see EL-002) tabletop centrifuge with buckets for 40 mL vials and 16 x 125 mm tubes orbital platform or similar shaker for extraction vortex mixer 0.1 g balance

  5. Equipment Required for Sample Preparation(not available from CAPE Technologies; see EL-002) small chemical fume hood sample evaporation system using clean nitrogen or compressed air

  6. Equipment Required for EIA (not available from CAPE Technologies; see EL-001) positive displacement micropipettors (adjustable 20 to 100 µL essential; 1-10 µL very helpful) Repeater pipettor Differential Photometer

  7. Supplies Required for Sample Preparation (see AN-008 for full information) SP3 Sample Preparation Kit (extraction materials not shown) CAPE Technologies Kits SP2-ST Sample Preparation Starter Kit SP2-RK rack for column procedure

  8. Supplies Required for Sample Preparation (not available from CAPE Technologies; see AN-008) HPLC grade or better hexane, acetone, and methanol ultrapure toluene (such as B&J) glass Pasteur pipets and bulbs glass pipets or glass syringes for pipetting solvents waste capture basin for column procedure borosilicate glass tubes (≈20 mL, such as 16 x 125 mm)

  9. Supplies Required for EIA Analysis (see IN-DF1 for full information) CAPE Technologies DF1 EIA Kit (DF1-60 shown) waste dump basin or wash aspiration system reagent or bottled distilled water

  10. AN-008 Method Set-up open SP2-ST Kit and remove materials glass reservoirs bag of stopper/stopcock assemblies bag of 10 and 20 mL polypropylene syringes Luer plugs

  11. AN-008 Method Set-up open SP3 Kit and remove materials 40 mL extraction vials, wooden spatulas, stainless steel mixing beads acid silica columns, carbon mini-columns, bulk acid silica for pretreatment (packaging may vary)

  12. AN-008 Protocol if sample is visibly wet, pre-drying is needed blot on absorbent paper and air dry soil “pancake” after overnight air drying in hood, between layers of paper towel blotting of wet soil samples using several layers of paper towel place 40 mL extraction vial onto balance and tare; using wooden spatula, weigh out 5.0 g of soil

  13. AN-008 Protocol weigh out 20 g of sodium sulfate and add to same 40 mL vial optional: 5 g sand can also be added for better dispersal of samples (especially clays) add 3 mixing beads to extraction vial

  14. AN-008 Protocol add 20 mL of 1:1 hexane:acetone to extraction vial cap vials tightly and lay on their sides in the SP3 Kit box or other box with padding to hold them tightly in place shake at 350 rpm for 2-4 hours if soils are high in clay, examine after 15-30 min to be sure soil lumps are adequately dispersed and vials are not leaking

  15. AN-008 Protocol remove from shaker and centrifuge 5 to 15 minutes at 1000 x g or less using Pasteur pipet, remove supernatant to storage vial (optional: can be left as supernatant over soil pellet in original extraction vial) soil samples after extraction and centrifugation, including method blank at left note wide range of possible appearances

  16. AN-008 Protocol place chosen volume of extract (see AN-008 for guidance) into evaporation tube with approx. 0.25 - 0.5 mL of tetradecane or other hydrocarbon keeper (exact volume is unimportant) evaporate hexane:acetone completely (caution: acetone interferes with the oxidative portion of the cleanup and can lead to false positive results if not completely removed) dilute residue in approx. 5 mL hexane If highly colored, pre-treat with bulk acid silica

  17. AN-008 Protocol Note deciding whether to pre-treat with bulk acid silica (based on extract color) pre-treatment needed pre-treatment not needed

  18. AN-008 Protocol Note example of pre-treatment with bulk acid silica (done in 16 x 125 mm evaporation tubes after acetone removal and dilution with hexane) Is pretreatment sufficient? The supernatant should be clear (no particulates) and colorless or very lightly colored

  19. Prepare coupled column system (method b) add 10 mL hexane to top of column use one acid silica column per sample (include QA samples) remove both end caps and put column into rack

  20. Prepare coupled column system (method b) when hexane begins dripping from tip, rinse outside of tip with 1- 2 mL hexane attach carbon mini-column with no air bubbles (very important to see AN-008 for details)

  21. Prepare coupled column system (b) insert stopper/stopcock assembly into top of acid silica column pressurize to allow hexane to flow through column into waste basin release pressure to stop flow when hexane level reaches 1-2 cm above bed of acid silica column remove stopper/stopcock assembly

  22. Extract Cleanup transfer hexane diluted extract to top of prepared acid silica column if pre-treated with bulk acid silica, then transfer both hexane and acid silica to column replace stopper/ stopcock assembly, and pressurize as before

  23. Extract Cleanup release pressure to stop flow when level reaches 1-2 cm above top of acid silica column (or top of bulk acid silica from pretreatment- note columns at right with dark acid silica on top ) complete sample loading onto carbon mini-column by adding 30 mL hexane wash (in 2 or 3 separate portions) and pressurize as before

  24. AN-008 Protocol Note example of pre-treatment with bulk acid silica (coupled column systems just before end of hexane wash) acid-neutral silica boundary should be difficult to see (arrows)

  25. Extract Cleanup at end of 30 mL wash, release pressure to stop flow when air begins to penetrate bottom (neutral silica) portion of column remove carbon mini-column and place on clean reservoir from SP2-ST Kit discard used acid silica column

  26. Extract Cleanup place carbon column and reservoir in rack and add 6 mL of 1:1 toluene:hexane to reservoir pressurize as before and allow solvent to flow through to waste release pressure to stop flow when level reaches tip of reservoir

  27. Extract Cleanup reverse carbon mini-column on same reservoir add 12 mL toluene pressurize as before and capture eluate in glass evaporation tube

  28. Extract Cleanup/Solvent Exchange add keeper (stock vial from DF1 Kit, diluted with methanol), according to IN-DF1 Table 3 place tubes in sample evaporation system and evaporate toluene

  29. Method Summary (IN-DF1 step J1-J2) warm EIA kit reagents prepare EIA wash 1

  30. Method Summary evaporate sample tubes until only a small amount of viscous residue remains centrifuge sample tubes 2 minutes at 1-2000 x g a small amount of keeper/sample residue should be visible in the bottom (the residue volume is only 20% of the original keeper volume); the residue should be homogeneous with no color or precipitate visible

  31. Method Summary (IN-DF1 step J3-J5) label EIA tubes and prepare for EIA by rinsing with water use Repeater pipettor to add 0.5 mL Sample Diluent to each tube

  32. Method Summary (IN-DF1 step J6) add standards (in keeper) gently mix each tube immediately

  33. Method Summary (AN-008 step F12) add 50-120 µL of methanol (depending on IN-DF1 Table 3 protocol) to each evaporation tube and mix 15 seconds using vortex mixer if protocol B, then let tube stand for 30-60 sec so liquid can drain back to bottom of tube examine tubes before proceeding- no color or precipitate should be visible in liquid

  34. Method Summary (IN-DF1 step J7) transfer 50 µL sample to EIA tube (33-80% of sample in evaporation tube) gently mix each tube immediately incubate 2-24 hrs at room temp.

  35. Method Summary (IN-DF1 step J8) tap rack inverted on paper towel to remove excess wash solution pour out contents and wash each tube 4x with 1 mL of EIA wash 1 (0.01% Triton in water) dump washes into waste basin

  36. Method Summary (IN-DF1 step J9) use Repeater pipettor to add 0.5 mL Competitor-HRP Conjugate to each tube incubate 15 min at room temp.

  37. Method Summary (IN-DF1 step J10) tap rack inverted on paper towel to remove excess wash water pour out contents as before and wash each tube 4x with 1 mL of bottled distilled water dump washes into waste basin

  38. Method Summary (IN-DF1 step J11) use Repeater pipettor to add 0.5 mL HRP-Substrate solution to each tube incubate 30 min at room temp. blue color should be visible in at least some tubes within the first few minutes

  39. Method Summary (IN-DF1 step J12-J13) read OD value in Differential Photometer add 0.5 mL Stop solution to each tube wipe each tube dry