1 / 9

Prepared for ASTM F04.30.06 May, 2007

This study presents finite element analysis (FEA) predictions of limit conditions for axial fatigue loading of generic coronary stent designs. It focuses on the impact of geometry and dimensional correction on fatigue behavior, stress distribution, and effective stress during extension. Residual stress effects are also examined. The analysis provides valuable insights for improving the fatigue resistance of coronary stents.

heltons
Download Presentation

Prepared for ASTM F04.30.06 May, 2007

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Fatigue to Failure:FEA Predictions of Limit Conditions for Axial Fatigue Loadingof Generic Coronary Stent DesignsPart 2: Geometry 1 with EP correction Prepared for ASTM F04.30.06May, 2007

  2. Dimensional Correction 0.085mm 0.11mm

  3. Expansion Analysis Stress at max expansion Stress after recoil Plastic strain due to expansion

  4. Expanded Diameter Expansion Minimum ID Recoil Maximum OD

  5. Location of Max Mean and Alternating Stress Mean Stress Alternating Stress

  6. Effect of Residual Stress Effective stress during extension with residual expansion stress Effective stress during extension without residual expansion stress

  7. Fatigue Scenario Studies

  8. Fatigue Analysis ±.70 % is roughly Here ±.50 % is roughly Here ±.33 % is Here ±.33 % without residual stress is Here