An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. HPLC DetectorsUV-VisFluorescence Derek Jackson CHM410/1410 October 22, 2009 djackson@chem.utoronto.ca

  2. HPLC Detectors Once a mixture of compounds has been separated by HPLC, how do we detect them? Requirements for an HPLC detector • Good sensitivity (high signal, low noise) • No interference from mobile phase • Must be able to work in a liquid phase environment

  3. HPLC Detectors List of more common HPLC detectors • Refractive Index • UV-Visible • Fluorescence • Conductivity (for ion chromatography) • Mass Spectrometry

  4. Refractive Index Detector • “Legacy” bulk property detector • Almost universal, rugged • Low sensitivity (high ppm), no gradient programs possible

  5. UV-Visible Detector • Most common detection method along with mass spectrometry • Detects solute analytes by their absorbance of light at various wavelengths • More sensitive than refractive index, depends on specific analyte and wavelength • Less sensitive than mass spec, compound must absorb in the UV-Vis, mobile phase cutoff

  6. Molecules and Light Why is our universe coloured? • Absorbance - compounds absorb light of specific wavelengths and reflect or transmit all others • Emission - compounds emit light after converted to a higher energy state (ex: fluorescence, phosphorescence)

  7. Molecular Orbital Theory Molecular orbitals exist at different energy levels; bonding orbitals (sigma/pi), non-bonding orbitals and anti-bonding orbitals Molecular absorption occurs when photonic energy causes promotion of an electron to a higher energy orbital, different types of transitions possible

  8. Molecular Orbital Theory • σ (sigma) – orbital has symmetry about the bonding axes, lowest energy • π (pi) – only one orbital plane passes through both nuclei involved • n (non-bonding) – orbital involved is not involved in bonding, usually a lone pair, higher in energy • σ*, π* (anti-bonding) – nodal planes exist between nuclei, high in energy, usually unpopulated in stable molecules

  9. Molecular Orbital Theory FORMALDEHYDE σπ n

  10. Molecular Orbital Theory BENZENE ππ*

  11. Molecular Orbital Theory Absorption occurs when light of a specific wavelength causes the electronic transition

  12. Molecular Orbital Theory HOMO = highest occupied molecular orbital (σ, π, n) LUMO = lowest unoccupied molecular orbital (π* , σ*) Most transitions we will be concerned with are from HOMO to LUMO The orbital types of HOMO/LUMO partially determine the energy required to make the transition

  13. Formaldehyde UV-Vis Possible Transitions for Formaldehyde π π* at182 nm n  π* at 290 nm But do we see sharp peaks at those wavelengths? Why are electronic transitions broad? Answer: Vibrational transitions combined with condensed phase and solvent effects broaden UV-Vis peaks

  14. Absorption Intensities π π* at182 nm (ε = 10,000 L M-1 cm-1) n  π* at 290 nm(ε = 12 L M-1 cm-1) In formaldehyde, π π* has strong absorption n  π* has very weak absorptions ε= Molar absorptivity Beer’s Law: A = ε c l Hence, UV-Vis can be used to quantify chromatography peaks linearly

  15. After Absorption

  16. Stokes Shift Remember: Emission spectra are redshifted relative to absorption (excitation) spectra

  17. Aromatic Rings • Benzene rings absorb “nominally” at about 254 nm but this can change depending on auxochromes • Absorption bands are redshifted by: • Electron donating groups (OH, NH2) redshift π π* transitions • Extended conjugation (NO2, C=O) which create n  π* transitions at longer wavelengths

  18. Auxochromes 254 nm; ε = 200 270 nm; ε = 1450 280 nm; ε = 1450 269 nm; ε = 7800 (π π*) 330 nm; ε = 125 (n  π*)

  19. Halogens • Halogens redshift UV-Vis spectra in the order F << Cl < Br < I because of polarizability 1: 10 Br 2: 9 Br 3: 8 Br 4: 7 Br 5: 6 Br 6: 5 Br 7: 4 Br 8: Sunlight

  20. Wavelength Selection • For HPLC-UV, want to observe a chromatogram at the longest reasonable wavelength, why? • Signal:noise is usually better at longer wavelengths due to reduction in noise from mobile phase and impurities • Ex: DNPH derivitization of carbonyls

  21. Wavelength Selection Acetone λMAX = 190 nm DNPH-Acetone λMAX = 360 nm

  22. Solvent Cutoffs

  23. Chromatograms Top = 220 nm Bottom = 280 nm CBN CBD

  24. Chromatograms λ = 245 nm

  25. HPLC-UV-Vis • Generally, UV-Vis HPLC detector not too different from a standalone UV-Vis (flow cell instead of a cuvette) • Variable wavelength detector vs. Diode array detector

  26. HPLC-UV-Vis • Variable wavelength detector - monochromator PMT

  27. HPLC-UV-Vis • Diode array detector - polychromator

  28. HPLC-UV-Vis • Variable Wavelength detector • More sensitive due to photomultiplier tube or amplification circuitry • Requires more method development • Diode array detector • Less sensitive due to photodiodes only • Very easy to develop a method

  29. HPLC-UV-Vis

  30. Fluorescence • Example: Highlighter Pens absorb UV and blue light and emit yellow-green

  31. Fluorescence Detectors

  32. Fluorescence Detectors • Greater sensitivity and selectivity over UV-Vis but the analyte must fluoresce! • λflu > λabs What makes a good fluorophore? • High absorbance, aromatic • Fused rings, electron donating groups • Quantum yield (Φ)

  33. Fluorescence Detectors • Need to select an excitation and an emission wavelength

  34. Chromatogram Top: UV-Vis Bottom: Fluorescence Hence, fluorescence is more selective and sensitive due to noise reductions

  35. Summary • Refractive Index Detector • “Legacy” detector, insensitive, no gradients in mobile phase possible • UV-Vis Detector • Detects absorption of chromophoric analytes based on molecular structure • Variable wavelength vs. Diode array detector • Fluorescence Detector • Most sensitive and selective detector