Peptide Sequencing by Matrix-assisted Laser Desorption-ionization Time-of-flight Mass Spectrometry (MALDI-TOF-MS) Combined with In-source Decay (ISD). Libo Cao Ph.D., Analytical (Dr. P. B. Harrington) 19 th April, 2001. Background. Matrix-assisted Laser Desorption-ionization (MALDI)

Download Presentation

Libo Cao Ph.D., Analytical (Dr. P. B. Harrington) 19 th April, 2001

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Peptide Sequencing by Matrix-assisted Laser Desorption-ionization Time-of-flight Mass Spectrometry (MALDI-TOF-MS) Combined with In-source Decay (ISD) • Libo Cao • Ph.D., Analytical (Dr. P. B. Harrington) • 19th April, 2001

  2. Background • Matrix-assisted Laser Desorption-ionization (MALDI) • Time-of-flight MS (TOF-MS) • Delayed Pulsed Ion Extraction (DPIE) • In-source Decay (ISD)

  3. Components of a Mass Spectrometer • Skoog, D. A.; Holler, F. J.; Nieman, T. A. Principles of Instrumental Analysis, 5th Ed.; Saunders College Publishing: Philadelphia, 1998;pp 255 Inlet System Ion Source Mass Analyzer Detector Signal Processor Sample Vacuum System Readout

  4. MALDI • A sample introduction and ionization method • Matrix absorbs strongly at the wavelength of a pulsed laser beam • Analyte exhibits little absorbance at this wavelength Li, L.; Garden, R. W.; Sweedler, J. V. Trends in Biotech.2000, 18 (4), 151-160.

  5. Goals of MALDI • Vaporize and ionize the analytes without fragmentation • Very low background noise • Obtain accurate molecular weight information • Tolerate high level of impurities

  6. Matrices Most Frequently Used for MALDI Together with the Usable Wavelengths Matrix Wavelength (nm) • Nicotinic acid 266, 220-290 • Benzoic acid derivatives: • 2,5-Dihydroxybenzoic acid 266, 337, 355 • Vanillic acid 266 • 2-Amino-benzoic acid 266, 337, 355 Karas, M.; Bahr, U. Anal. Chem. 1990, 9, 322.

  7. Fundamental Equations for TOF-MS • KE =zV = ½ m1v12 = ½ m2v22 • L = v t • t = (m/2zV)1/2 L

  8. TOF Mass Analyzer • Cotter, R. J.Anal. Chem. News&Features,1999, 77, 445-450.

  9. Delayed Pulsed Ion Extraction (DPIE) in MALDI -200 V -100v Nitrogen Laser (337.1nm) -100v Brown, R. S.; Lennon, J. J. Anal. Chem. 1995, 67, 1998-2003. Sample Probe Tip Ion Guide (1) Signal XY Deflector Plates Ion Guide (2) -2 kV G1 G2 G1 Voltage = Pulse voltage (0—3 kv max) G2 Voltage = Bias Voltage (24 kv max)

  10. Benefits of DPIE • Kinetic energy focusing - Obtain same initial velocity for ions with the same mass to charge ratio (m/z)

  11. Ion-source Decay (ISD) • The decay of the precursor ions occur in the MALDI ion source with DPIE • Provide specific ions for certain sequencing information Li, L.; Garden, R. W.; Sweedler, J. V. Trends in Biotech.2000, 18 (4), 151-160

  12. a-, b-, c-, x-, y-, z- Series Ions y z x NH CHR CO NH CHR CO a c b

  13. MALDI Fast Metastable Ion Decay Spectrum of Porcine Insulin Obtained with DPIE • Brown, R. S.; Lennon, J. J. Anal. Chem.1995, 67, 3990-3999.

  14. MALDI Fast Metastable Ion Decay Spectrum of Cytochrome C Obtained with DPIE • Brown, R. S.; Lennon, J. J. Anal. Chem.1995, 67, 3990-3999.

  15. Metastable Ions (m*) • Broader peaks occurring at non-integer m/z values in MS • C6H6·+  C6H5·+ + H · m/z 78 m/z 77 • m* =m22/m1= 772/78 = 76.01 m/z

  16. Peptide Sequencing Information Brown, R. S.; Lennon, J. J. Anal. Chem.1995, 67, 3990-3999

  17. Conclusions • Mostly generate specific c-,y- ions • Sequencing of post-translational modifications • Do not need predigestion of proteins • Large peptides and small proteins (1,500 –18,000 Da.)

  18. Future Work • More suitable matrix materials need to be found • Simplify the instrumentation and operation • Potential applications need to be developed - for larger proteins - for mixture proteins

  19. Acknowledgements • Dr. Peter de B. Harrington • Tricia Buxton • Guoxiang Chen • Mingming Xu • Xueqing Liu • Erin Kolbrich • Jeannette Perr • Jennifer Cline • Yuka Minoshima

  20. ********************************

  21. G1 Voltage = Pulse voltage (0—3 kv max) G2 Voltage = Bias Voltage (24 kv max)

  22. Metastable ions (m*) m1v1 = m2v2 + (m1 – m2)v2 zV = ½ m1v2 m2 v2/r = Bzv m*/z= B2r2/2V

  23. Disadvantage * Edman degradation -----Tedious, indirect, labeling process complicated, need digested peptides