TIPS: STIS Report

SPACE TELESCOPE SCIENCE INSTITUTE. TIPS: STIS Report. Operated for NASA by AURA. Paul Goudfrooij. Unusual Target ACQ Failures: Update & Resolution Calibration of CTE loss in Spectroscopic Modes

ivrit
Download Presentation

TIPS: STIS Report

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. SPACE TELESCOPE SCIENCE INSTITUTE TIPS: STIS Report Operated for NASA by AURA Paul Goudfrooij • Unusual Target ACQ Failures: Update & Resolution • Calibration of CTE loss in Spectroscopic Modes • Full story available under http://www.stsci.edu/hst/stis/training/team/activities/lectures.html • Also STIS ISR 2003-03 • New “Pseudo-Apertures” (if time available)

  2. 1 2 1 2 3 3   Recent Target ACQ Failures(with L. Dressel, R. Pitts, T. Wheeler) • Two recent ACQs failed (March 2, April 6) due to No Flux in the Lamp Image • All mechanisms show nominal telemetry • ACQ macro used 3.8 mA setting for HITM1 lamp, much below ‘default’ 10 mA • 3.8 mA setting originally put in place to allow wavecals for the most sensitive MAMA settings, and to save lamp life time • Contacted manufacturer + their consultant • Sputtered material forms a ring inside glass envelope around cathode. If set at (too) low current, electrons may flow to sputtered ring rather than to cathode • Conclusion: Lamp did not fire • Since 5/12/03, ACQs use 10 mA setting. All ACQs taken so far are OK.

  3. Amp D Nominal Readout Direction Axis2 (Y) Nominal Clocking Direction 1 2 (fluxD / fluxB) Y CTI = Parallel (virtual) overscan Amp C Amp A Amp B Axis1 (X) Serial overscan Serial overscan Correcting CCD Spectroscopy for CTE Loss (with R. Bohlin) STIS CCD: • 4 Readout Amps (1 / corner) • Bi-directional Clocking yields CTI  1 – CTE: Sensitive Region (1024x1024 pix) Measured using “Sparse Field Tests”

  4. (i) “Internal” Sparse Field Test “Sparse Field” Tests • Sparse fields to ensure that sources do not overlap, in which case (e.g.) PSF wings could fill traps for sources along the readout direction • Two varieties: • Annual series of lamp images through narrow slits, projected at 5 positions along columns (or rows) • Designed to represent “worst–case” point source spectroscopy (should be no background to fill traps)

  5. “Sparse Field” Tests • (ii) “External” sparse field test (annually) • A. Imaging: • Sparse outer field in NGC 6752 • CVZ target (‘cheap’ observing time; yields range of backgrounds) • 3 exposure times; 50CCD mode • B. Spectroscopy: • Young open cluster NGC 346, in nebulosity • CVZ target • Slitless; 3 exp. times; G430L • [O II] 3727, H, [O III] 5007 lines in nebulosity provide three convenient, ~constant “sky” levels per spectrum

  6. Spectroscopy Imaging CCD Column Number CCD Row Number CTI Parametrization:Imaging vs. Spectroscopy • Dependence on signal & background levels to be done separately for imaging and spectroscopy

  7. bck signal External Sparse Field Test: Imaging CTIAnalysis Clear dependence on background level (“sky”) • Slope systematically flatter with increasing flux • “Sky” presumably fills traps in bottoms of potential wells, mostly affecting transfer of small charge packets. • Suggests CTI  exp –

  8. The Strong Effect of Background:Gain=1 vs. Gain=4 • Background level in spectroscopy mode typically low, dominated by dark current • Also need to account for spurious charge of the STIS CCD CCD Readout flush CCD

  9. Functional Dependence on Signal and Background Levels • Iterative Process for Spectroscopy • Parameter space covered by ESF test at a given epoch is limited • Sensitivity monitor: good coverage of signal levels, but not of sky • G230LB data allow suitable cross-comparison with MAMA G230L AGK+81D266,G230LB

  10. Time Constant of CTI Evolution • Need several datasets, each with same signal & background level • Need datasets covering long baseline in time  ISF data • Have to correct for signal & background dependence prior to fitting 60 e– CTI = CTI0 + { 1 + 0.243 [± 0.016] (t – t0) } (with t in yr) CTI data points from Tom Brown 120 180 500 3400

  11. 0.17 bg GROSS Final CTI Correction Formula (For Point-Source Spectroscopy) • Define background (sky) and epoch parameters: yr = (MJD – 51765.25) / 365.25 (i.e., relative to 2000.6) bg = max(BACKGROUND,0) + 0.5 for CCD Gain = 1 + 5.0 for CCD Gain = 4 • Functional form producing best fit to the data: ) ( CTI = 0.0467 GROSS–0.720  exp –3.85  (1 + 0.243 yr) • Implementation into the pipeline: • Formula parameters into CCD table reference file (new columns) • 1-D extraction step (x1d) will correct for CTI by default for CCD data (CTE correction step switchable) • For Cycle 12 Phase II, provided downloadable IRAF script to calculate correction factor for a given net & background level.

  12. Quality of CTI fit CTI Correction good to 7%  Spectrophotometry good to 1%

  13. New “Pseudo-Apertures” • FUV-MAMA first-order spectroscopy at detector location with low dark • ~ 2’’ above bottom of detector • Reduction of dark current by factor of 5 • 52x0.05D1, …, F25QTZD1 • Improvement of Fringe Flats at E1 positions • Important to align fringes in flat with those in target spectrum • 52x0.1 slit (best for defringing) location is offset in dispersion direction from wider slits • New ‘E2’ positions will place target slightly off-center in slits  0.2 arcsec wide • New WEDGEA0.6 position for 50CORON • Provide POS TARGs to GOs in Phase-II Update; Apertures to be implemented in next APT build. nominal new