Air Quality Monitoring

Reading: Chap 7.1 & 7.2. Air Quality Monitoring. ENV 4101/5105 Elements of Air Pollution. http://www.statcan.ca/english/freepub/16-254-XIE/2007000/tablesgraphs/figure3-en.htm. Monitoring Considerations Manual vs. automated (real-time, continuous) monitoring Sampling time

jacob
Download Presentation

Air Quality Monitoring

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Reading: Chap 7.1 & 7.2 Air Quality Monitoring ENV 4101/5105 Elements of Air Pollution http://www.statcan.ca/english/freepub/16-254-XIE/2007000/tablesgraphs/figure3-en.htm Aerosol & Particulate Research Lab

  2. Monitoring Considerations • Manual vs. automated (real-time, continuous) monitoring • Sampling time • Federal reference method (FRM) vs. equivalent method (EM) • Monitoring of Ambient Air Pollutants • SO2, NOx, CO, O3 • Hydrocarbons • PM10, PM2.5 • Source Sampling and Monitoring • Sampling train • Isokinetic sampling • Quality Assurance Programs • Quality Assurance • Quality Control • Air Quality Monitoring Network Air Quality Monitoring What’s the use of ambient air quality monitoring data? Aerosol & Particulate Research Lab

  3. Monitoring Considerations What are the objectives? Examples of criteria for selecting ambient sampling locations? • Sampling location: limited number of fixed site monitors whose locations reflect objectives of air quality monitoring program • Lower limit of detection (LOD): a sufficient amount of pollutant must be collected, f(sampling rate, duration) • Integrated sampling vs. real-time sampling • Area sampling vs. personal sampling • Collection efficiency of the instrument: • Low flow rate for gas-phase contaminants (< 1 L/min) Aerosol & Particulate Research Lab

  4. Differences in averaging times associated with real-time data Which duration should you use? Aerosol & Particulate Research Lab

  5. National Ambient Air Quality Standards Why different durations? http://www.epa.gov/air/criteria.html Aerosol & Particulate Research Lab

  6. Federal Reference Methods for Criteria Pollutants Aerosol & Particulate Research Lab

  7. FRM - Spectrophotometry (pararosanilne method) Air sample  potassium tetrachloromercurate solution  HgCl2SO3-2  react with HCHO and colorless pararosaniline hydrochloride  red-violet product  measured spectrophotometrically  SO2 concentration EM – FT- IR Spectrometry (Absorption of IR by SO2 in the air  SO2 concentration) SO2 http://clu-in.org/programs/21m2/openpath/op-ftir/images/exhibit3.gif Aerosol & Particulate Research Lab

  8. SO2 EM – UV Fluorescence 1) UV light excites SO2 to a higher energy state SO2 + hv1  SO2* 2) Decay of the excited SO2*, emitting a characteristic radiation SO2*  SO2 + hv2 www.cse.polyu.edu.hk/~airlab/so2.jpg Aerosol & Particulate Research Lab

  9. FRM – Gas-Phase Chemiluminescence Chemiluminescence: emission of light from electronically excited chemical species formed in chemical reactions. NO + O3 NO2* + O2 NO2*  NO2 + hv Measurement of NO2: conversion of NO2 to NO, and subsequent measurement by chemiluminescence. 2NO2 + Mo  3 NO + MoO3 Possible interference: N-containing compounds  higher measured NO2 NO – NO2 – NOx EM – FT- IR Spectrometry www.k2bw.com/images/chem.gif Can you design an instrument that can measure the concentration of both NO and NO2 in the air? Aerosol & Particulate Research Lab

  10. FRM – Nondispersive Infrared (NDIR) spectrometry CO strongly absorbs infrared energy at certain wavelengths. Detection device: two cylindrical cells, a sample and a reference cell. Difference in infrared energy in the two cells  concentration of CO CO CO Reference Cell Sample Cell ~ IR ~ Detection Aerosol & Particulate Research Lab

  11. O3 UV light (254 nm) Detector O3 FRM – Chemiluminescence Light emissions produced on reaction of O3 with ethylene (C2H4). C2H4 flammable – replaced by Rhodamine B dye embedded in a disk Rhodamine B does not attain a stable baseline rapidly after exposure to O3 EM – UV Photometry Absorption of UV light (254 nm) by O3 and subsequent use of photometry to measure the reduction of UV energy O3 What would cause an interference on a UV photometry O3 monitor? Aerosol & Particulate Research Lab

  12. FRM – Gas Chromatography - FID Intensity Time NMHCs Stationary and mobile phases GC-Detector: Generate an electronic signal when a gas other than the carrier gas elutes from the column. Aerosol & Particulate Research Lab

  13. NMHC • Mass sensitive rather than concentration sensitive http://teaching.shu.ac.uk/hwb/chemistry/tutorials/chrom/gaschrm.htm • FID – Flame Ionization Detection: • Combustion of organic substances • Positive ions (+) and electrons (-) are formed when burned – change in current Aerosol & Particulate Research Lab

  14. Impaction and filtration are the primary PM collection principles • Measure the weight of exposed and clean filters • High-volume sampler (Hi-Vol) • Typical sampling duration – 24 h Animation – PM10 Impactor • PM10 Sampler • Remove particles > 10 µm by impaction on a greased surface • Particles < 10 µm collected on a quartz glass fiber filter Particulate Pollutants What does the PM10 sampler measure? Number or mass concentration of particles? PM10 sampler with size-selective inlet http://www.recetox.muni.cz/images/airsag/PM10.jpg Aerosol & Particulate Research Lab

  15. Animation – Virtual Impactor TEOM Series 1400abAmbient Particulate Monitor Thermo Electron Co. Dichotomous Sampler • Equivalent method: TEOM • Measure PM10, PM2.5, TSP • Tapered element oscillating microbalance • Real-time measurement of particle mass collected on a filter http://crac.ucc.ie/images/sampler1.jpg How can a Dichotomous sampler measure coarse (PM2.5-10)and fine (PM2.5) particles? Aerosol & Particulate Research Lab

  16. Aerosol flow In Clean air out Animation – Cascade Impactor Cascade Impactor How can we collect different sizes of particles using cascade impactor? Aerosol & Particulate Research Lab

  17. Source Sampling and Monitoring • Stack Sampling of PM • Probe inserted into the stack • Temperature sensor • Pitot tube – gas velocity and flow rate • Two-module sampling unit • Isokinetic Sampling • Particles – inertial forces • Samples must be collected at the same rate of low as the stack gas What’s the use of source emission data? Aerosol & Particulate Research Lab

  18. Isokinetic Sampling Fig 8 -2, Aerosol Measurement, 2nd Edition, 2001 Aerosol & Particulate Research Lab

  19. Continuous Emission Monitoring (CEM) Opacity • Compare plume darkness to Ringlemann chart by trained smoke readers • Averages of measurements of ¼ or ½ minute over an hour • Simple, low cost, legal acceptance Ringlemann chart Certain utilities and industrial sources are required to measure stack emission continuously SO2, NO2, opacity, CO2, TRS, H2S, Hg Aerosol & Particulate Research Lab

  20.  Bias  Precision Relative Error: Coefficient of variation: Accuracy, Precision and Bias Accuracy is a combination of random (precision) & systematic (bias) errors. Which of the 3 cases has the highest accuracy? Why? Aerosol & Particulate Research Lab

  21. Calibration • Often used for adjusting bias-type errors • Measured values are compared to standard reference values (for pollutant concentration) or standard airflow measuring techniques/devices (for volume air flow) • Primary vs. secondary standard for flow: traceable to the National Institute of Standards and Technology (NIST) • Primary: bubble meter • Secondary: wet or dry test meters calibrated by bubble meter • Gas standards: traceable to a NIST reference material • CO, SO2, NO2, NO: available in cylinder gas or permeation tubes • O3: NIST certified O3 generator Aerosol & Particulate Research Lab

  22. Goal: Valid and reliable air quality monitoring data • Quality Assurance (QA) • Setting policy and overseeing management controls • Planning, review of data collection activities and data use • Setting data quality objectives, assigning responsibilities, conducting reviews, and implementing corrective actions • Quality Control (QC) • Technical aspects of data quality programs • Implementation of specific QC procedures: calibrations, checks, replicate samples, routine self-assessment, and audits Quality Assurance Programs It is federal rule to document QA/QC efforts ! Aerosol & Particulate Research Lab

  23. State and Local Air Monitoring Stations (SLAMS) • Highest pollutant concentrations • Representative concentrations in areas of high population density • Impact of major emission sources • Regional background concentrations • Extent of pollutant transport among populated areas • Welfare-related impacts in more rural and remote areas • National Air Monitoring Stations (NAMS) • Urban area, long-term air quality monitoring network • Air quality comparisons and trends analysis Air Quality Monitoring Network • Photochemical Assessment Monitoring Stations (PAMS) • Monitor O3 and photochemical air pollutants • Clean Air Status and Trends Network (CASTNet) • Primary source for rural O3 level and dry atmospheric deposition • National Atmospheric Deposition Program (NADP) • Assess the problem of atmospheric deposition and its effects on aquatic and terrestrial ecosystems (H+, NH4+, SO42-, NO3-, Cl-, Ca2+, Mg2+, K+, Hg) Aerosol & Particulate Research Lab

  24. Monitoring Considerations • Manual vs. automated (real-time, continuous) monitoring • Sampling time • Federal reference method (FRM) vs. equivalent method (EM) • Monitoring of Air Pollutants • SO2, NOx, CO, O3 • Hydrocarbons • PM10, PM2.5 • Source Sampling and Monitoring • Sampling train • Isokinetic sampling • Quality Assurance Programs • Quality Assurance • Quality Control • Air Quality Monitoring Network Quick Reflections Aerosol & Particulate Research Lab