Emmanuel Angulo (CSU Fresno) Weekly Meeting ATLAS Muon Micromegas 7/8/2012

Homogeneity of Small Chambers T4, T6, T7, T8 and Gas Gain Measurements of Frascati and Roma Chambers . Emmanuel Angulo (CSU Fresno) Weekly Meeting ATLAS Muon Micromegas 7/8/2012. Cu X-Ray Measurements. 9 0 Positions (0.9cm x 1 cm ) HV of Strips @ 52 0 HV of Drift @ 300

jag
Download Presentation

Emmanuel Angulo (CSU Fresno) Weekly Meeting ATLAS Muon Micromegas 7/8/2012

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Homogeneity of Small Chambers T4, T6, T7, T8 and Gas Gain Measurements of Frascati and Roma Chambers Emmanuel Angulo (CSU Fresno) Weekly Meeting ATLAS MuonMicromegas 7/8/2012

  2. Cu X-Ray Measurements • 90 Positions (0.9cm x 1cm) • HV of Strips @ 520 • HV of Drift @ 300 • X-Ray with 1mm collimator and cu foil cover @ 100000 event count, and @ 10mm distance from chamber

  3. T4 (10mm to 5mm Drift) Unifortmity

  4. T6 Uniformity Connector 2 Connector 2 Connector 1 Connector 2 Middle point 6cm x 5cm Corner points 1cm x 1cm Connector 2 Connector 1

  5. T6 Uniformity

  6. T7 Uniformity Connector 2 Connector 1 Connector 1 Middle point 4cm x 5cm Corner points 1cm x 1cm Connector 2 Connector 1

  7. T7 Uniformity

  8. T8 Uniformity Connector 2 Connector 1 Connector 1 Middle point 4cm x 5cm Noise up to Ch 143 (Bad signal with connector 2) Corner points 1cm x 1cm Connector 2 Connector 1

  9. T8 Uniformity with 55Fe • 25 Positions (1.8cm x 1.8cm) • 3mm Collimator Hole @ 50000 event count • Strip HV @ 520 • Drift HV @ 300

  10. Roma and Frascati Chambers Gas Gain 55Fe Measurements X-Strips 1 2 • 5 Positions (1.8cm x 1.8cm) • 3mm Collimator Hole @ 20000 event count • Strip HV from 440-560 • Drift HV @ 300 Y-Strips 3 4 5 HV Connectors

  11. Frascati Gas Gain

  12. Roma Gas Gain