1 / 89

TEM EDS X-ray analysis

TEM EDS X-ray analysis. Topics. EDS Hardware TEM EDS vs. SEM EDS Signal Processing and Quantification k-factor determination and uncertainty CM200 and EDAX Genesis software. EDS unit. Oxford Instruments: EDSHardwareExplained.pdf www.x-raymicroanalysis.com.

job
Download Presentation

TEM EDS X-ray analysis

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. TEM EDSX-ray analysis

  2. Topics • EDS Hardware • TEM EDS vs. SEM EDS • Signal Processing and Quantification • k-factor determination and uncertainty • CM200 and EDAX Genesis software

  3. EDS unit Oxford Instruments: EDSHardwareExplained.pdf www.x-raymicroanalysis.com Image courtesy of Oxford Instruments

  4. Si(Li) crystal FET Window Electron trap (SEM only) Collimator EDS detector (SEM) Image courtesy of Oxford Instruments

  5. TEM detector Image courtesy of Oxford Instruments

  6. Silicon Drift Detector Peltier Cooler Heatsink Heat pipe Image courtesy of Oxford Instruments

  7. SDD Window Electron trap FET SDD Collimator Image courtesy of Oxford Instruments

  8. SDD detail Image courtesy of Oxford Instruments

  9. EDS Si(Li) diode detector

  10. EDS Si(Li) diode detector

  11. Ultrathin Window

  12. Signal processing

  13. X-ray Generation and Detection • The number of X-rays generated is the product of • the number of electrons * • number of atoms * • ionization cross-section (s) * • fluorescent yield (w). • The number of X-rays we detect is reduced by absorption of the X-rays in the sample and detector and by the limited solid angle subtended by the detector. (ignoring fluorescence effects)

  14. Ionization cross-section Bethe cross-section model U = (E0/Ec) = overvoltage

  15. Avoid this range of U (mostly SEM issue) Ionization Cross-section vs. Overvoltage Overvoltage values (U) for 120kV beam

  16. Fluorescent Yield (K shell)

  17. Fluorescent Yield

  18. Detector Efficiency

  19. C O Na Detector Efficiency - UTW vs. Be Window

  20. SDD vs. Si(Li) efficiency

  21. Detector Icing and Efficiency

  22. TEM vs. SEM • Higher kV in TEM • Hard X-rays • High energy BSE (can’t use electron trap) • Ill-defined TEM sample geometry • Thin TEM sample vs. thick (semi-infinite)

  23. Relative resolution

  24. Spatial resolution

  25. Revised Goldstein Equation b = broadening (m) E0 = Beam energy in keV Z = Atomic Number Nv = atoms per volume (#/m3) t = thickness (m) W&C eq 36.1

  26. Sample geometry • An SEM sample is usually assumed to be a flat, homogeneous, semi-infinite object. • TEM samples vary in thickness and can be bent.

  27. Pt Obj aperture TEM EDS geometry

  28. Typical TEM sample

  29. Qualitative Analysis • Match families of peaks • Start at higher energies and work down • Be careful of spectral contamination (e.g. Cu from grid) • Be careful when using Auto Peak ID functions.

  30. Quantification • Extract peak intensities • Remove Background • Deconvolve overlapped peaks if needed • Quantify with Cliff-Lorimer equation • Modify results with absorption & fluorescence corrections

  31. Extract intensity • Straight line subtraction • Kramers Law continuum modeling • “Top Hat” filtering

  32. Background subtraction

  33. Average Background

  34. Background modeling • Need to choose background areas • Requires knowledge of detector response (esp. at low keV)

  35. Calculated vs. Observed Continuum

  36. Kramer’s Law continuum

  37. Top Hat Filter

  38. Filtered spectrum

  39. Linear combination Where Si are pure element standard spectra To calculate the ci, perform a multiple least squares fit and minimize the c2 value. You multiply the intensity of the Si by ci to get the peak intensities.

  40. Cliff-Lorimer equation Ca & Cb are in weight percent (a tradition started by Castaing)

  41. Cliff-Lorimer equation • C-L equation describes relative concentrations and gives only n-1 equations for n elements • The nth equation is to require the sum of the concentrations = 1. If you neglect an element, the calculated concentrations will vary from the true concentrations though the relative amounts are OK. • The C-L equation neglects absorption and fluorescence. It is exactly valid only in the thin film limit (i.e. at t = 0).

  42. k-factors • kab-factors are relative sensitivity factors. Always reported with respect to another (usually common) element. • The original reference element was Si because Cliff and Lorimer were geologists. Metallurgists will often use Fe. • We can use theoretical k-factors, but …

  43. Theoretical kTiFe can vary by over 10% Theoretical k-factors

  44. k-factors • We can use theoretical k-factors, but … experimental k-factors are more reliable. • Experimental k-factors will be valid for the instrument and the conditions used for the acquisition ONLY. • A change in microscope or detector parameters will change the k-factors (e.g. detector icing, radiation damage, etc.)

  45. Experimental k-factors • Ia/Ib will vary with thickness due to absorption and fluorescence. • Since the k-factors are exact only in the thin-film limit, we can measure the ratios at various thicknesses and extrapolate to zero thickness.

  46. Intensity ratio vs. thickness

  47. Thickness determination • Can measure crystalline thickness using CBED techniques. • Can use EELS to measure material thickness though the mean-free-path (l) needs to be determined first. • Contamination marks, projections of defects, and stereo imaging have also been used.

  48. van Capellen technique • If you assume a constant beam current and moderate absorption, then the count rate scales with thickness. • Plot the ratio against the sum of the counts rather than thickness. • Extrapolate to zero count rate (i.e. zero thickness). • Do a linear regression to get the Y-intercept. • Most packages will give you a standard estimate of error (i.e. s).

  49. Profile for van Capellan technique

  50. Ni-47Al-3Ti data