An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. SRRp 2010, Bucuresti, 8 octombrie 2010 Implicarea Laboratorului de Metrologia Radionuclizilor din IFIN-HH in imbunatatirea calitatii masurarii 99mTc in unitatile de medicina nucleara M.Sahagia*, A. Luca, A.Antohe, A.C.Watjen, C.Ivan, Institutul National de C&D pentru Fizica si Inginerie Nucleara Horia Hulubei, IFIN-HH Str. Atomistilor nr.407, 077125, Magurele, Jud. Ilfov Corespondenta*: msahagia@nipne.ro

  2. Rezultate recente ale Laboratorului de Metrologia Radionuclizilor (LMR) din IFIN-HH, suport pentru medicina nucleara Asigurarea lantului de trasabilitate prin: • Etalonarea absoluta a radionuclizilor utilizati pentru radiofarmaceutice • Participarea la comparari internationale • Implementarea Sistemului de Calitate, aprobat de Technical Committee – Quality (TC-Q) al EURAMET • Echivalenta internationala este recunoscuta prin : 34 fisiere, Calibration and Measurement Capability (CMC) publicate in CIPM-MRA, Annex C, Data Base, www.bipm.org, in 2008. • Obtinerea acreditarii RENAR pentru LMR: - Etalonare in Activitate Certificat : LE 013/2009; la p. 9 figureaza Etalonarea calibratoarelor - Incercare in Masurarea Activitatilor Joase Certificat: LI 084/2009 • Desemnare CNCAN ,2009, Laborator de Etalonare Notificat: LE 05/2009

  3. Cadrul National si International • Legislatia de Metrologie din Romania: Calibratoarele de Radioizotopi din spitale nu sunt in lista mijloacelor de masura supuse controlului legal. Ele trebuie sa fie reetalonate dupa un program • Agentia Internationala pentru Energia Atomica a condus: Coordinated Research Project (CRP) E2.10.05 “Harmonization of quality practices for nuclear medicine radioactivity measurements” 2004 – 2009. Rezultatele CRP: - Elaborararea de documente • Technical Report Series 454 • Protocols for participation of the national laboratories at international comparisons • Protocols for organization of the national comparisons with nuclear medicine units . - Organizatrea de comparari internationale : 131I si 57Co - Organizarea de comparari nationale: 131I si 99mTc • LMR a participat la toate actiunile CRP.

  4. Publicatii : International Atomic Energy Agency (IAEA)-Vienna, Austria - Technical Report Series 454 (TRS 454), 2006, ISBN 92-0-105306-1, ”Quality Assurance for Radioactivity Measurements in Nuclear Medicine” p.1-96, Contributors to drafting and review: M.Dondi, C.P.Herbst, A.Iwahara, M.Morengo, S. Mather, J.P. Norenberg, V.Olsovcova, P.Oropesa Verdicia, P.Otiz-Lopez, L.Joseph, M.Ghafoori, J.A.Onto, N. Ramamoorthy, M.Sahagia, K.R. Shortt, S.Tastan, P.Vincze, M.J.Woods, B.E.Zimmerman Olsovcova, V., Iwahara, A., Oropesa, P.,Joseph, L., Rabindra, A., Ghafoori, M., Son, H-K., Sahagia, M., Tastan, S., Zimmerman, B. ”National comparisons of I-131 measurements among nuclear medicine clinics”. Applied Radiations and Isotopes 6(2010)1371-1377 Zimmerman, B., E., Megzifene, A., Short, K.R. ”Establishing measurement traceability for national laboratories: Results of an IAEA comparison of 131I”. Appl. Radiat. Isot, 66, 6-7 (2008)954-959 Sahagia, M., Wätjen, A.C., Ivan, C. «Progress in organizing national and international comparisons for nuclear medicine measurements”. Rom. J. Phys.54, 7-8(2009)619-627

  5. Lucrari recente ale LMR in cadrulCRP - E2.10.05 • Participarea la o comparareorganizata de AIEA pentru Co-57, 2008. • Recunoscuta ca o compararesuplimentara a CIPM-MRA, CCRI : CCRI(II)-S6.Co-57 • Motive pentruorganizareacompararii: - Tc-99m esteutilizatpescaralarga in clinici de diagnostic • Nu se puteaorganiza o comparareinternationala de Tc-99m datoritavietii sale scurte (6,007 ore) • Pregatireauneicompararirealenationale de Tc-99m

  6. Participarea la compararea de 57Co. - Solutia a fost distribuita de QSA Global GmbH-Germania - Etalonarea absoluta s-a facut prin metoda coinicidentelor 4πPC-γ. - S-a reetalonat camera de ionizare CENTRONIC IG 12/20A, Etalon Secundar de Activitate al LMR. Rezultatele au fost trimise la AIEA, fiind publicate in Rapoartele Internationale.

  7. 2. Preliminary results of 57Co comparison RAPOARTE INTERNATIONALE : 1. B.E.Zimmerman,S.Palm. Results of an international comparison of 57Co. Appl. Radiat.Isot. 68(2010)1212 - 1220 Draft A Report of CCRI Comparison CCRI(II)-S6.Co-57 B. E. Zimmerman1, S. Palm2, and G. Ratel3 1Physics Laboratory, National Institute of Standards and Technology, Gaithersburg, MD USA 2 Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, Vienna, Austria 3 Bureau International des Poids et Mesures, Sévres, France

  8. Radioactive concentration of the 57Co solution

  9. Radionuclidul Tc-99m - Solutiile au fost preparate prin elutia unor generatori comerciali, din solutie de Mo-99. - Tc-99m este un radionuclid dificil de etalonat absolut - Prima etalonare a camerei CENTRONIC s-a bazat pe metoda spectrometriei radiatiilor gama. - In 2005 a fost elaborata o metoda originala de etalonare prin metoda coincidentelor; a fost reetalonata camera CENTRONIC. Cele doua valori au avut o concordanta foarte buna. Sahagia, Maria. Standardization of 99mTc. Appl. Radiat. Isot. 64(2006)1234-1237

  10. Organizarea compararii de Tc-99m • LMR aelaborat Procedura de Lucru, Cod AC-PL-LMR-07, ”Organizarea de comparari interlaboratoare asupra masurarii activitatii surselor solide si a solutiilor radioactive” • Documente de referinta: (i) ISO/IEC 17043, (ii) ILAC-G13:2000 „Guidelines for Requirements for the Competence of Providers of Proficiency Testing Schemes” (iii) EAL-P7, 1996 „Interlaboratory Comparisons of the European Cooperation for Accreditation of Laboratories”

  11. Desfasurarea compararii • Conceptia a fost adaptata unui radionuclid de viata scurta • Participantii si-au eluat proprii generatori, - au masurat fiola cu calibratoarele proprii, - au completat formularul de comparare - au trimis formularul completat si fiola la IFIN-HH • DRMR: compararile de I-131 si Tc-99m - parte a procesului de acreditare RENAR pentru CPRLAB

  12. Date raportate • Activitatea, AR , exprimata in MBq, sau mCi la momentul masurarii • Incertitudinea statistica combinata, uc2 % , k = 1 • Alte date: - Tipul echipamentului - Situatia etalonarii/verificarii metrologice • Incercari suplimentare: - masurarea fondului, - masurari cu sursele test: Co-57, Ba-133, Co-60, Cs-137, - corectii de dezintegrare

  13. Masurari la IFIN-HH, LMR • Fiolele au fost transportate la IFIN-HH • Activittatea a fost masurata cu camera CENTRONIC IG12/20A, Etalon Secundar de Activitate, calibrata cu solutie etalon. • Activitatea este ACA, Activitate Conventional Adevarata, sau Activitate de Referinta, incertitudinea statistica, uc1 %, k = 1.

  14. Prelucrarea datelor la LMR: • Corectii de dezintegrare • Calculul parametrilor relevanti: • R = AR/ACA, raportul dintre valoarea raportata si cea conventional adevarata; • Diferenta; • Deviatia normalizata la incertitudinea statistica:, k = 2, • Rezultatele sunt declarate acceptabile, daca:

  15. Results

  16. Concluziile compararii: Din 7 valori raportate, 5 sunt in limita diferentei < 5 % , iar 7 sunt in limita< 10 % -Toti participantii dse incedreaza in limita impusa Farmacopeea Europeana: incertitudine maxima de masurare in spitale de 10 % Valoarea medie R = AR/ACA , pentru cele 7 valori raportate este: R = 1.012 (Sn-1 = 4.9 %). Rezultatele sunt aproximativ normal distribuite Criteriul En<1 , k=2, - participantul L2b a subevaluat incertitudinea si este eliminat, desi diferenta sa este <5 %

  17. CONCLUZII • LMR, IFIN-HH a asigurat intregul suport laboratoarelor de medicina nucleara din Romania, prin: 1. Recunoasterea internationala a echivalentei etalonului romanesc de activitate prin comparari internationale: radionuclidul 57Co. 2. Obtinerea acreditarilor nationale necesare pentru asigurarea trasabilitatii nationale, organizarea unei comparari nationale de 99mTc si etalonarea calibratoarelor utilizate in unitatile de medicina nucleara.