280 likes | 778 Views
Electrical Wiring in the Dialysis Unit. ‘A Shocking State of Affairs!’ Ian Morgan ART October 2010. Discussion. Are we providing a safe environment in which to dialyse our patients? Are we protecting them from electric shocks?
E N D
Electrical Wiring in the Dialysis Unit ‘A Shocking State of Affairs!’ Ian Morgan ART October 2010
Discussion • Are we providing a safe environment in which to dialyse our patients? • Are we protecting them from electric shocks? • Should we understand the DoH building guidelines covering electrical wiring (Health Technical Memorandum 06-01)? • Are those building our dialysis units interpreting these guidelines correctly?
Hazard: Something that has the potential do harm. Using mains powered electrical equipment in the patient environment. Placing a conductor in or near the heart. Micro shock to the myocardium. Risk: The likelihood of the hazard occurring & the consequence. Prevalence of catheters. High impedance or open circuit in the mains flex earth conductor. Patient touching conductive objects. Ventricular fibrillation leading to death. Hazards & Risks
Leakage current Infusion pump ECG monitor Dialysis chair BP monitor Heater Lamp Infusion pump ECG monitor Dialysis chair BP monitor TV Phone charger Wicker and Svensson, 2006
What is a Safe Current? • The HD machine chassis current in normal operation could be 100μA and up to 500μA if the earth conductor breaks. • The allowable current for equipment connected to the heart is 50μA in a single fault condition. • Even this may be enough to cause VF as Laks et al (1996) commented that the minimum current shown to cause VF in a human was 15μA. • Deller (1979) commented “(HD) Equipment that has been constructed to BS 5724 should not, by its design, give rise to primary electrical hazards”. • Is that statement correct?
Lets try to kill a dog Are you £µçЌing mad! • Historical acceptance of Type B limits for HD equipment is shown in experiments done by Frize et al. (1978), where a leakage current of 85mA was passed through a 14 gauge needle in the left front leg of a dog to earth via the right rear leg & did not induce VF or heart failure; concluding “…50μA, implies a large margin of safety”. Are you mad!
HD Machine Leakage Current? • Jonsson & Stegmayr (2000) measured patient leakage currents of four machines (Gambro AK10, AK100 & Fresenius 2008 (C&E) & 4008E with values from 1 to 20μA in normal condition and 60 to 140μA with a broken earth conductor; measure at dialyser connectors. • Main on applied part produced 3,500μA leakage current. • Jonsson et al, (2005) then measure in vitro patient leakage current in single fault condition at the tip of a vascular catheter primed with blood; median 68μA (range 35-118μA ). • When mains was applied to the catheter tip a dangerous leakage current was measured; median 610μA (range 441-662μA).
Assessing the Risks • “At least 65% of patients presenting more than three months before initiation of dialysis should start HD with a usable native arteriovenous fistula” Clinical Practice Guidelines Vascular Access for Haemodialysis (2007). • 20-35% of patients are dialysed with a central venous catheter. • Roy (2001) concluded that 10% of deaths on dialysis are most likely caused by technical failure. • “Faults in detachable power cords account for 80-90% of all earth bond failures as … moulded power cables are prone to stress when … dropped” Backes (2006). • Many non-medical electronic devices are present in the patient environment; TV’s PC’s etc.
Assessing the Risks Unacceptable, stop activity and make immediate improvements Tolerable, look to improve within specified timescale Adequate, look to improve @ next review Consequence Acceptable, no further action but ensure controls are maintained Likelihood
Assessing the Risks • The consequence is life threatening and the likelihood is not insignificant. • “This (micro) shock causes no tissue damage, which makes it unlikely that the true cause of death is found in autopsy” Polaschegg • Law requires that we put in place control measures to reduce the risks – so far as is reasonably practical.
Risk Reduction Hierarchy • Eliminate the hazard- This cannot be achieved, the potential to electrocute the patient will always be present. • Reduce the hazard- Electricity will always be in the patient environment. Increase the number of fistulas and grafts used as access. • Prevent contact with the hazard- Bond all conductive surfaces to earth & remove faulty equipment. • Introduce safe system of work- Ensure devices are tested and potential equalisation leads are used at all times. • Protective equipment- Ensure earth equalisation sockets & leads are available.
Risk Reduction • Dialysis machines are constructed to meet Type CF requirement. Unlikely in the near future. • Exclude all non-medical electrical equipment from the patient environment. Impractical. • Supply power in the patient environment through an isolated power supply. Impractical & costly. • Use earth potential equalisation leads on all medical devices. • This reduces the risk of a fault in the mains power cord but not excessive patient auxiliary current from non-medical devices & other conductive surfaces.
Isolated Power supply Leakage current Dialysis Machine L X N X
Creating a Safer Environment • All medical devices are constructed to comply with the Medical Devices Directive (93/42/EEC): Procurement. • Routinely perform earth bonding, earth leakage and patient leakage current tests: Renal Technologist/MEP. • Routinely perform portable appliance electrical safety tests on non-medical devices: Estates/Facilities. • The electrical distribution within the clinical environment is designed to comply with DoH Estates & Facilities Division, Health Technical Memorandum 06-01 (HTM 06-01) published 2007: Estates/Facilities. • Routine inspection (6 monthly) and service (2 yearly) of earth & power distribution system (HTM 06-01): Estates/Facilities. • Connect all medical equipment to an earth potential equalisation socket: End users.
HTM 06-01 Electrical service supply & distribution • Provides guidance for all works on fixed wiring within healthcare premises. • Provides guidance to managers on how European and British Standards relating to electrical safety can be used to fulfil their duty relating to the H & S at Work Act 1974. • Understanding & ownership of risk.
Reverse Mains Dialysis Machine L N
Engaging the Design Team • Make your self familiar with the relevant sections of HTM 06-01. • Ask if those designing & modifying your dialysis units are familiar with it as the previous version HTM 2007 published in 1993 was very vague. • Do they understand dialysis and the risks? • Do they realise that providing earth potential equalisation sockets is mandatory?
The Future • More dialysis equipment manufacturers are recommending the use of earth potential equalisation when dialysing patient with a catheter. • The MHRA are looking into providing additional guidance on this subject similar to the MEIGaN (Medical Electrical Installation Guidance Notes) they produced for diagnostic imaging & x-ray installations.
References • Backes J 2006 A Practical Guide to IEC60601-1. Rigel Medical • Deller AG 1979 Electrical safety in dialysis J Med Eng Technol 3(4):186-191 • Department of Health 2007 Health technical Memorandum 06-01: Electrical service supply and distribution. https://estatesknowledge.dh.gov.uk/index.php?option=com_documents&Itemid=1&sdocid=49(registration required) • Frize M, Scott J, Durie N, Park G (1978) Fibrillation caused by leakage current from dialysis machines – What is the danger? Med. & Biol. Eng. & Comput16:124 • Johnsson P, Eliasson G, Stegmayr B (2005) Blood lines conduct leakage current during haemodialysis: a potential safety risk during first failure, especially for patients with dialysis catheter as access. Med. Biol. Eng. Comput43:731-38 • Jonsson P, Stegmayr B (2000) Current Leakage in Hemodialysis Machines May Be a safety Risk for Patients. Artificial Organs24(12):977-87 • Laks M, Arzbaecher R, Bailey J, Geselowitz D, Berson A (1996) Recommendations for Safe Current Limits for Electrocardiographs Circulation93:837-39 • MHRA 2007 Medical Electrical Installation Guidance Notes http://www.mhra.gov.uk/home/idcplg?IdcService=GET_FILE&dDocName=CON2018069&RevisionSelectionMethod=Latest&noSaveAs=0&Rendition=WEB • Polaschegg H-D Neglected Safety Aspects in Hemodialysis Machines and Their Related Problems. http://www.aami.org/publications/hh/Neglected.Polaschegg.pdf Accessed 20 July 2010 • Roy T 2001 Patients’ safety and haemodialysis devices Nephrol Dial Transplant16:2138-2142 • Wicker PO, Svennson R (2006) Is potential equalization in HD treatments necessary? J Ren Care32(3):136-40