1 / 12

Quick & Safe Techniques for AcrySof Delivery

‘Turn, Shoot & Unturn’ for the Multipiece ‘Shoot and Turn’ for the Single Piece. Quick & Safe Techniques for AcrySof Delivery. Dr. Suven Bhattacharjee , MS, DO, DNB, FRF. Complete care Eye Clinic Kolkata, INDIA. suvenb@gmail.com. No Financial Interest. Purpose.

johana
Download Presentation

Quick & Safe Techniques for AcrySof Delivery

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. ‘Turn, Shoot & Unturn’ for the Multipiece ‘Shoot and Turn’ for the Single Piece Quick & Safe Techniques for AcrySof Delivery Dr. Suven Bhattacharjee, MS, DO, DNB, FRF. Complete care Eye Clinic Kolkata, INDIA suvenb@gmail.com No Financial Interest

  2. Purpose • Estimate‘CapsuloZonular stress’ caused by AcrySofMultipiece IOLwhen injected without any rotation of the Injector (Shooter / Monarch / Royale). • Recommend a safe technique for Implantation for AcrySofMultipieceIOL. • Analyze the unfolding motions of the AcrySof Single piece IOL. • Recommend easy technique for delivery of AcrySof single piece IOL in bag. CapsuloZonular Stress - Indicators • Available space in Ant Chamber • Mean Anterior Chamber Depth (2.93 mm)1 + Mean thickness of Crystalline lens(4.2 mm)2 = 7.13 mm • Tanja M. Rabsilber, MD, Ramin Khoramnia, Gerd U. Auffarth, MD: Anterior chamber measurements using Pentacam rotating Scheimpflug camera: J Cataract Refract Surg 2006; 32:456–459 • Vargas LG, Peng Q, Escobar-Gomez M, Apple DJ. Overview of modern foldable intraocular lenses and clinically relevant anatomy and histology of the crystalline lens.Int Ophthalmol Clin. 2001;41(3):1-15. If the Space Required for Unrestricted Unfolding is > 6mm Leading haptic has to be bent against PC for safe unfolding in AC / Cap bag • CZ stress is directly proportional to: • Space required for unrestricted unfolding and • Force required to bend the leading haptic.

  3. Materials & Methods AcrySof MA60AC(MP) Mutipiece & SN60WF (IQ) Single piece IOLs Synchronized Surgical videos - Eye and Injector AcrySof MP ‘Direct Shoot’ - 20 eyes ‘Turn, Shoot & Unturn’ - 10 eyes AcrySof IQ ‘Direct Shoot’ - 18 eyes ‘Shoot & Turn’ - 12 eyes

  4. Laboratory Video Set up Top Cam Microscope head • IOLs injected into transparent plastic chamber filled with OVD. • Operating microscope used to capture top view. • Detached Slit lamp head used for front view. Front Cam Slit Lamp head Calibration & Measurements • Magnification Correction • Caliper jaws set to 1 mm & Immersed in OVD chamber • Still Image extracted & Scale with smallest marking corresponding to distance between caliper jaws Superimposed • Used this as a ‘mm’ scale for measurements.

  5. Haptic bending force measurement – Set up • IOL partially delivered till stage requiring maximum vertical space • Haptic pushed on the pan of a precision laboratory weighing scale to bend on itself till 3 mm • Force required for every mm of bending noted.

  6. Observation - AcrySof Multipiece MA60AC Vertical space required for unrestricted unfolding 1 8 4 2 7 3 4 5 6

  7. Leading haptic bending force - AcrySof Multipiece Leading haptic - Vertical Leading Haptic - Horizontal • 1st mm: 30 + 7 mg • 2nd mm: 94 + 13 mg • 3rd mm: 168 + 21 mg 1st mm: 399 + 11 mg 2nd mm: 855 + 19 mg 3rd mm: 1420 + 72 mg Leadinghaptic orientation – Prior to Unfolding Leading haptic points to the right within cartridge barrel Therefore, To ensure that the leading haptic points left and down – ‘Injector’ needs to be turned clockwise by 160° before beginning delivery of IOL

  8. Observation -‘Arc of Sweep’ of Leading hapticAcrySof MA60AC - Multipiece • Relevant only till the final stages of optic delivery • IOL injected without any rotation of delivery system • Superimposition of 1st frame over last frame during unfolding • shows ‘arc of sweep’ of leading haptic 170° Angle between haptic position at onset and completion of unfolding = 170 ̊ clockwiseTherefore, Angle by which Injector is to be rotated to maintain constant haptic position throughout Unfolding = 170 ̊ Anticlockwise

  9. Vertical Space required – Direct Shoot – AcrySof IQ IOL rotates clockwise 90̊ - trailing haptic delivers outside capsular bag 1 8 8 2 7 3 6 5 4

  10. ‘Shoot & Turn’ Technique - AcrySof IQ Hapticscolour tagged for better visualization Injector rotates anticlockwise 90̊ as IOL is delivered - Trailing haptic unfolds at same plane as leading haptic - ‘Reduced space & In the bag’ 1 9 8 2 8 7 3 6 5 4

  11. New Techniques for Acrysof IOL Delivery Multipiece - Turn, Shoot and Unturn • Insert Barrel of cartridge into AC – bevel down - Hapticpointing right • Rotate (Turn) Injector Clockwise by 160 ̊ • Advance rod by pushing (Shoot) or screwing to deliver IOL. - haptic emerges pointing left and down • Rotate Injector anticlockwise (Unturn) by 170 ̊ as the IOL is delivered to maintain the haptic horizontal throughout. Watch Video! Single piece - Shoot and Turn • Insert Barrel of cartridge into AC – bevel down • Advance rod by pushing (Shoot) or screwing to deliver IOL • As IOL delivers into capsular bag - Rotate Injector anticlockwise (Turn) by 90 ̊ • Trailing haptic of IOL is also delivered into capsular bag.

  12. Results: Surgical video analysis • Time from emergence of leading haptic till IOL delivery • AcrySof MP (Direct shoot) – 8 sec ( Turn, shoot & Unturn) – 9 sec • AcrySof SP (Direct shoot) – 6 sec ( Shoot & Turn) – 4 sec Trailing haptic placement & IOL delivery • AcrySofIQ (Direct Shoot) 18 eyes • All eyes • Optic delivered obliquely • Trailing hapticmanipulated into bag • AcrySof IQ (Shoot & Turn) 12 eyes • All eyes - Optic delivered Completely in the bag • 11 Eyes -Trailing haptic into the bag • Leading haptic placement & IOL delivery • AcrySof MP (Direct Shoot) 20 eyes • 8 eyes – leading haptic into bag directly • 12 eyes – Leading haptic orientation corrected in AC • and then placed in the bag • 1 eye - haptic stuck on Iris, endoth touch • 6 eyes – OVD top up during delivery • AcrySof MP (Turn, Shoot and Unturn) 10 eyes • All 10 safely in the bag in controlled manner Conclusion • The ‘Turn, Shoot & Unturn’ technique reduces CZ Stress of AcrySof MP IOL by keeping the leading haptic horizontal and reducing the vertical space required for unfolding. • The ‘Shoot & Turn’ technique delivers the AcrySof SP IOL in the capsular bag in a single action by countering the rotation of the IOL as it unfolds.