250 likes | 270 Views
Scout Pro Training. The Ohaus Scout Pro. Introducing the Scout Pro Portable Balance: 7 Models From 200g x 0.01g to 6000g x 1g 5 Application Modes Optional USB or RS232 Interface Accessory w/cable Density/Specific Gravity Determination with Below Balance Hook
E N D
The Ohaus Scout Pro • Introducing the Scout Pro Portable Balance: • 7 Models From 200g x 0.01g to 6000g x 1g • 5 Application Modes • Optional USB or RS232 Interface Accessory w/cable • Density/Specific Gravity Determination with Below Balance Hook • Battery or AC Power (Power Pack Included) • Integral Security Bracket • Programmable Energy Saving Auto Shut-off
Shipping lock Secures load cell Unpack & Setup • Battery Installation • 4 AA batteries provide 40 hours of operation
AC plug Unpack & Setup • Security bracket
Unpack & Setup • USB Accessory • RS232 Accessory
Unpack & Setup • Rotate square pans counter clockwise • Insert round pans
Turn balance on Scout Pro is shipped ready to operate in all available weighing units, parts counting and percent weighing 0.00g ON/ZERO Off 0.00g ON/ZERO Off PRINT Unit Operation PRINT Unit • Short press to print. • Long press to switch weighing unit.
Weigh in any unit OFF ON/ZERO Off PRINT Unit Operation • Hold Off until “OFF” is displayed, then release to turn off.
With balance off, hold ON/ZERO until MENU is displayed MENU ON/ZERO Off PRINT Unit C.A.L. ON/ZERO Off PRINT Unit C.A.L. S.E.t.U.P. U.n.I.t. E.n.d. Menus and Calibration • Release ON/ZERO to enter the menu
PRINT Unit PRINT Unit PRINT Unit PRINT Unit C.A.L. S.E.t.U.P. U.n.I.t. E.n.d. Auto-Off Linearity cal Mode - Display Hold - Totalize - Off End Menu System Calibration g kg oz lb PC - Count - Cnt. Enh - Off % End
Access calibration ON/ZERO Off ON/ZERO Off Calibration C.A.L. - C - C 2 0 0 g • Add cal weight - C - d o n E
Parts Counting & Parts Counting Setup • Mode: Parts Counting • This mode allows the user to count based on the average weight of an original reference sample • This mode is enabled by setting “PC” to “Count” in the “.U.n.i.t.” part of the menu • To set-up mode: • Enter Menu Mode, press PRINT/Unit to go to “.U.n.i.t.” and then press ON/ZERO/Off • Press PRINT/Unit to go to “UnitS PC” and then press ON/ZERO/Off • Press PRINT/Unit to go to “Count” and then press ON/ZERO/Off • Press PRINT/Unit to go to “End” and then press ON/ZERO/Off • Press PRINT/Unit to go to “.E.n.d.” and then press ON/ZERO/Off • Sample Sizes -- • The Scout Pro allows the user to change the sample size with a press of the Print/Unit key
EnhancedParts Counting • Mode: Enhanced Parts Counting • APW Enhancement -- • The Scout Pro allows the user to turn-on automatic recalculation of the average piece weight (APW Optimization) -- this is the same function as found in Trooper Count and Ranger Count • Without Optimization, the APW is the average of the weight of the sample, divided by the number of pieces in the sample -- if the user needs a sample of 100, they have to count by hand 100 pieces! • With Optimization, the APW is the average of the weight of the initial sample as well as any subsequent samples -- the user can reach a sample of 100 by adding parts on the pan 10 at a time, increasing accuracy!
Enhanced PartsCounting Setup • Mode: Enhanced Parts Counting • This mode is enabled by setting “PC” to “Cnt.Enh” in the “.U.n.i.t.” part of the menu • To set-up mode: • Enter Menu Mode, press PRINT/Unit to go to “.U.n.i.t.” and then press ON/ZERO/Off • Press PRINT/Unit to go to “UnitS PC” and then press ON/ZERO/Off • Press PRINT/Unit to go to “Cnt.Enh” and then press ON/ZERO/Off • Press PRINT/Unit to go to “End” and then press ON/ZERO/Off • Press PRINT/Unit to go to “.E.n.d.” and then press ON/ZERO/Off • Sample Sizes -- • The Scout Pro allows the user to change the sample size with a press of the Print/Unit key
0 . 0 0 g PRINT Unit PRINT Unit PC ON/ZERO Off ON/ZERO Off CLr.APu Put 5 Parts Counting Operation • Place a container • Press and hold PRINT/Unit key until “PC” is indicated • Display will then show “Clr.APu” • Press Print/Units to use the average piece (APW) stored in memory (if there is one) • To re-establish a new APW, press On/Zero/Off once • Scout Pro will display “Put 10”. To select a different sample size, press Print/Unit • Place displayed number of pieces in container and press ON/ZERO/Off • Scout Pro will calculate APW and display total count • To reset APW, press Print/Unit until “PC” is displayed again 5 PC
Percent Weighing % • Mode: Percent Weighing • This mode allows the user to display the weight of a sample as a percentage of a reference weight • The reference weight is placed on the pan first, and equals 100% • Each reference weight will be a percentage from 0.1% to 99,999.9%, depending on the Capacity X Readability of the scale -- most users will use 1.0% to 200.0% at the most • Applications: Formulation and compounding, weight checking
Percent Weighing % Setup • Mode: Percent Weighing • This mode is enabled by setting “%” to “On” in the “.U.n.i.t.” part of the menu • To set-up mode: • Enter Menu Mode, press PRINT/Unit to go to “.U.n.i.t.” and then press ON/ZERO/Off • Press PRINT/Unit to go to “UnitS %” and then press ON/ZERO/Off • Press PRINT/Unit to go to “On %” and then press ON/ZERO/Off • Press PRINT/Unit to go to “End” and then press ON/ZERO/Off • Press PRINT/Unit to go to “.E.n.d.” and then press ON/ZERO/Off
PRINT Unit PRINT Unit % ON/ZERO Off ON/ZERO Off ON/ZERO Off CLr.rEF PutrEF 123.45 g 1 0 0 % Percent Operation • Press and hold Print/Unit key until “%” is indicated • Display will then show “Clr.Ref” • Press Print/Units to use target weight in memory (if there is one) • To enter a new target weight, press On/Zero/Off once • Scout Pro will display “Put.rEF” • Place target weight on pan, press On/Zero/Off again -- Scout Pro will display target weight value • Press On/Zero/Off again -- Scout Pro will display 100% • To reset target weight, press Print/Unit until “%” is displayed again 2 5 %
Display Hold • Mode: Display Hold • This mode allows the user to hold the highest stable weight on the display until manually cleared • Display Hold can work with any weighing unit or mode -- including % or parts counting -- except Totalization • The Scout Pro will show the current measurement on the pan with a flashing stability indicator “*” • Applications: Postal and shipping 0.00 Display Hold
Display Hold Setup • Mode: Display Hold • This mode is enabled by selecting “HoLd” in the “MOde” part of the “S.E.t.U.P” menu • To set-up mode: • Enter Menu Mode, press PRINT/Unit to go to “S.E.t.U.P” and then press ON/ZERO/Off • Press PRINT/Unit to go to “MOdE” and then press ON/ZERO/Off • Press PRINT/Unit to go to “HoLd” and then press ON/ZERO/Off • Press PRINT/Unit to go to “End” and then press ON/ZERO/Off • Press PRINT/Unit to go to “.E.n.d.” and then press ON/ZERO/Off 0.00 Display Hold
Hold PRINT Unit ON/ZERO Off HLd.OFF Display Hold Operation • To Use Mode: • Press PRINT/Unit once to start Display Hold • “HoLd” will be displayed briefly indicating Display Hold is active • Place weight on pan • “*” will flash indicating highest stable weight value will be held • Take weight of pan • The display will change and then display the Hold value with the flashing “*” • To Clear Display press On/Zero/Off • To Exit Mode hold ON/ZERO/Off until “HLdOFF” is displayed 123.45 g 123.45 g 0.00 Display Hold
Totalization • Mode: Totalization • This mode allows the user to sum a series of weights -- up to the limit of the display (999,999) • Totalization can work with any weighing unit or mode -- including % or parts counting -- except Display Hold • Once you are in Totalization, you cannot change weighing units/modes without exiting Totalization • The Scout Pro will show the current measurement on the pan with a solid “t”, and then the total as a flashing “t” • Applications: Formulation and compounding, parts kitting Σ Totalization
Totalization Setup • Mode: Totalization • This mode is enabled by selecting “totAL” in the “MOde” part of the “S.E.t.U.P” menu • To set-up mode: • Enter Menu Mode, press PRINT/Unit to go to “S.E.t.U.P” and then press ON/ZERO/Off • Press PRINT/Unit to go to “MOdE” and then press ON/ZERO/Off • Press PRINT/Unit to go to “totAL” and then press ON/ZERO/Off • Press PRINT/Unit to go to “End” and then press ON/ZERO/Off • Press PRINT/Unit to go to “.E.n.d.” and then press ON/ZERO/Off Σ Totalization
t o t A L PRINT Unit PRINT Unit ON/ZERO Off tot.OFF Totalization Operation PREPARATION: 2 plastic cups, different colored water samples. Tare an empty cup • Press PRINT/Unit once to start Totalization • “totAL” will be displayed briefly and “t” will be displayed indicating Totalization is active • Place a sample on the pan and press PRINT/Unit to enter the weight into memory • “t” will flash indicating total weight. • Remove the sample. “t” will stop flashing and the balance will return to zero • Repeat the procedure with another sample and read the totalized weight • To exit Totalization, hold down ON/ZERO until “tot.OFF” is displayed 0.00 g t 123.45 g t 0.00 g t Σ Totalization
The Ohaus Scout Pro • Scout Pro USP’s • Mechanical overload and shock load protection • Protects weighing cell 4 times better than Scout II • Simple 2-button operation – 5 applications and 7 weighing units • Optional USB or RS232 interfaces, simply insert the module; cable included