1 / 11

PC 3 Advantages

Thermal stabilization of spray chamber improves long-term stability Reduces effect of laboratory temperature variation on signal stability Reduces solvent loading. Fast rinse-out using PFA-ST nebulizer and O-ring-free quartz cyclonic spray chamber. Reduced oxides

jroney
Download Presentation

PC 3 Advantages

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Thermal stabilization of spray chamber improves long-term stability Reduces effect of laboratory temperature variation on signal stability Reduces solvent loading. Fast rinse-out using PFA-ST nebulizer and O-ring-free quartz cyclonic spray chamber. Reduced oxides Enhances stability with PC3-SSI Chamber: Isotope ratio measurements Other analyses requiring improved precision Additional reduction of oxides vs. standard PC3 Optional oxygen addition port for organic solvents. PC3 Advantages

  2. Default settings of spray chamber temperature (+2 C/ -5 C) for aqueous or organic solvents. Custom temperature settings available upon request. Quartz cyclonic or HF resistant cyclonic spray chambers available. Direct connection to torch injector. Completely o-ring-free for organic solvents analysis. Organic solvent analysis using PFA-50 nebulizer. Aqueous sample analysis using PFA-ST nebulizer. PC3-SSI model for isotope ratio determination or other high precision analyses. Optional mounting tray for x-y-z torch adjustment. Air-cooled. PC3 Features

  3. ELEMENT2 with Standard Scott Spray ChamberFive Orders of Magnitude Rinse-out > 3 minutesPFA-ST Self-aspirating Nebulizer 10 ppb 7 6 5 Log intensity (cps) 4 ~ 0.2 ppt 3 2 0 6 10 14 20 16 18 2 8 12 4 Time (min)

  4. ELEMENT2 with PC3 Inlet SystemFive Orders of Magnitude Rinse-out < 30sPFA-ST Self-aspirating Nebulizer 10 ppb 10 ppb 7 7 6 6 5 5 Log intensity (cps) Log intensity (cps) 4 4 ~ 0.1 ppt ~ 0.1 ppt 3 3 2 2 20 20 18 18 16 16 2 2 6 6 14 14 4 4 0 0 8 8 12 12 10 10 Time (min)

  5. PC3 Peltier Cooled Cyclonic Spray Chamber 4 Hour Stability 1 ppb Tuning Solution ThermoFinnigan ELEMENT2 ICPMS 20000000 18000000 16000000 14000000 7Li 1.8% RSD 12000000 59Co 1.5% RSD 10000000 Integrated Signal Intensity (cps) 115In 1.0% RSD 8000000 138Ba 1.2% RSD 6000000 238U 1.0% RSD 4000000 2000000 0 0 1.2 2.4 3.6 0.24 0.48 0.72 0.96 1.44 1.68 1.92 2.16 2.64 2.88 3.12 3.36 3.84 Time (hour)

  6. PC3-SSI • Peltier-cooled Stable Sample Introduction • Low flow PFA-50 nebulizer • Homogenizes sample aerosol • Reduces interferences • Reduces drift • Improves isotope ratio precision

  7. Short-Term (20 min) Signal StabilityScott vs. PC3 vs. PC3-SSI n=10, 1 min read, 1 min delay, LR 10% mass window

  8. PC3-SSI Short-term Precision

  9. Long-Term (14 h) Signal StabilityScott vs. PC3-SSI n=102, 3 min read, 5 min delay, LR 10% mass window

  10. Pb Isotope Ratio Precision 5 ms dwell time PC3-SSI Sample Inlet System with ELEMENT2 ICPMS 0.180% 0.160% 0.140% 0.120% 204/208 0.100% % RSD 207/208 0.080% 206/208 0.060% 0.040% 0.020% 0.000% 0 2000 4000 6000 8000 10000 10 ppb Pb Number of Passes

  11. Summary: PC3 and PC3-SSI with Finnigan ELEMENT • Rinse-out improved by 2 minutes Scott-type chamber. • Sensitivity enhancement up to 2x or more depending upon nebulizer flow rate. • Reduced metal-oxide formation. • SSI version for improved short-term and long-term signal stability. • Convenient system for organic solvent analysis.

More Related