An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Design of a K=1600 SC cyclotron for Carbon therapyECPM NiceNovember 2 2006Yves JongenFounder & CRO Ion Beam ApplicationsBelgium

  2. K= 1600 Cyclotron design team • JINR Dubna (Russia): G. Shirkov, E. Syresin, G. Karamysheva, N. Morozov, E. Samsonov, S. Kostromin, V. Alexandrov, N. Kazarinov • IBA Louvain la Neuve (Belgium): D. Vandeplassche, W. Beeckman, W. Kleeven, S. Zaremba, M. Abs, A. Blondin

  3. The market of PT is rapidly expanding! • Over the last decade, and excluding 2006, 14 proton therapy systems (average 1.4 / year) and one carbon therapy system have been ordered to industry • Two carbon therapy systems are under construction by national laboratories • In 2006 only, 4 to 6 proton therapy systems will be ordered to industry

  4. IBA is today the market leader in PT Siemens Optivus Mitsubishi 0 0 1 Accel 2 Hitachi IBA 3 9

  5. Essen, Germany Seoul, Korea Beijing, China MGH Boston Kashiwa, Japan (with SHI) MPRI Wan Jie, China UPHS (UPENN) Hampton Univ. of Florida IBA Particle Therapy Partners

  6. Cyclotrons for Carbon therapy? • In 1991, when IBA entered in PT, the consensus was that the best accelerator for PT was a synchrotron • IBA introduced a very effective cyclotron design, and today the majority of PT centers use the cyclotron technology • Over these 15 years, users came to appreciate the advantages of cyclotrons: • Simplicity • Reliability • Lower cost and size • But, most importantly, the ability to modulate rapidly and accurately the proton beam current

  7. Beam current command & feedback

  8. In less space and cost than a synchrotron: a two cyclotrons phased approach

  9. The IBA proton-carbon facility Proton/Carbon Gantry Room 1 Accelerator 230 MeV Proton Gantry Room FBR Proton/Carbon Gantry Room 2 Accelerator 400 MeV ESS Research lines

  10. The IBA Carbon cyclotron design • Superconducting isochronous cyclotron, accelerating Q/M = 1/2 ions to 400 MeV/U (H2 +, Alphas, Li6 3+, B10 5+, C12 6+, N14 7+, 016 8+, Ne20 10+) • Design very similar to IBA PT cyclotron, but with higher magnetic field thanks to superconducting coils, and increased diameter (6.3 m vs. 4.7 m)

  11. Engineering view of the 400 MeV/u cyclotron

  12. Final TOSCA model

  13. Elliptical gap

  14. Azimuthal field distribution

  15. Superconducting coils • Uses low current density (85 A/mm²) Ni-Ti standard cable from MRI systems • Stored energy 65.2 MJ • Peak field on the conductor: 5.26 T: 3 °K of temperature margin • 170 A, 130 minutes charging time at 100 V • Cooling by helium loop, with 4 external 3 stages 1.5 W recondensers (RDK 415 C from Sumitomo)

  16. H2+ to Alphas correction coils

  17. Resonances at the end of acceleration

  18. Study of 3 Qr = 4 resonance Ar initial =1 mm Ar initial=3 mm Ar initial =2 mm Ar initial=4 mm

  19. Resonances in the K=1600 (1 to 8)

  20. Resonances in the K=1600 (9 to 17)

  21. The RF system • The cyclotron will have two 45° dees, operating on the 4th harmonic mode at 76 MHz (lambda/4 = 98.7 cm) • Each dee will be supported by 2 flat pillars and 2 circular pillars in a half-wave resonator. • Each cavity is powered by a 76 MHz, 100 kW tetrode based amplifier (as used in the current C230) • The dee voltage increases from 100 kV at center to 210 kV at extraction, resulting in an average of 600 kV/turn

  22. Layout of RF cavity

  23. Four ion sources allow quick change of ion Carbon 6+ source Proton (H2 +) source Helium 2+ (alphas) source Spare & research source, switchable to: Li, B, N, O, Ne

  24. Supernanogan ion source from Pantechnik

  25. The spiral inflector k’=0.1; E=20kV/cm

  26. Central trajectories

  27. Extraction • One ESD located in the valley (47° long) • Field 170 kV/cm • Two magnetostatic gradient correctors • One permanent magnets quadrupole • Extraction of H2+ by stripping

  28. Extraction path after ESD

  29. Extraction optics

  30. Proton extraction by stripping of H2+

  31. Agreement with INFN on Catania cyclotron