Concentration Matters

10 ng of sample (note: we request 50 ng of DNA in 10ul and run 5ul on the 3130xl). Concentration Matters. Note the areas of low resolution. 20 ng of sample (note: we request 50 ng of DNA in 10ul and run 5ul on the 3130xl).

Download Presentation

Concentration Matters

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. 10 ng of sample (note: we request 50 ng of DNA in 10ul and run 5ul on the 3130xl) Concentration Matters Note the areas of low resolution 20 ng of sample (note: we request 50 ng of DNA in 10ul and run 5ul on the 3130xl) The call reliability has drastically increased at 20 ng in the same region

  2. Note the increased signal intensity and trace scores for 20 ng The Contiguous Read Length is the longest uninterrupted stretch of bases with quality higher than a specific limit. QV20+ value is the total number of bases in the entire trace that have basecaller quality value greater than or equal to 20 (99% probability the call is correct).