1 / 39

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology. RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. Part 19.03: Optimization of protection in Mammography Practical exercise. Overview.

kamin
Download Presentation

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION INDIAGNOSTIC ANDINTERVENTIONAL RADIOLOGY Part 19.03: Optimization of protection in Mammography Practical exercise

  2. Overview • To be able to apply quality control protocols to mammography equipment • To measure the breast entrance surface dose and determine the average glandular dose • Interpretation of results 19.03 : Optimization of protection in Mammography

  3. IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology Part 19.03: Optimization of protection in Mammography Topic 1: Entrance surface dose (measurements on patients)

  4. Entrance surface dose (ESAK) • The Entrance Surface Air Kerma (ESAK) free-in-air, i.e., without backscatter, has become the most frequent used quantity for patient dosimetry in mammography • ESAK can be determined under reference conditions either with AEC or manual exposure 19.03 : Optimization of protection in Mammography

  5. Entrance surface dose (I) (measurements on patients) • This simple method of determining ESAK requires dosemeters from a central laboratory to provide first line information on the level of radiation dose being delivered to patients. • It is intended for facilities where local resources or expertise are not available. 19.03 : Optimization of protection in Mammography

  6. Entrance surface dose (II) (measurements on patients) • Select 10 patients with a compressed breast thickness in the range 4 to 6 cm • For each patient and for each projection, position one calibrated TLD on the upper inner quadrant of the breast and x-ray the patient normally • Remove the TLD and keep it away from radiation 19.03 : Optimization of protection in Mammography

  7. Entrance surface dose (III) (measurements on patients) • Complete the questionnaire for dosimetry on patient (provided by the central laboratory) • When all 10 TLDs have been used, return them together with the completed questionnaire to the issuing dosimetry laboratory • Compare the mean value of ESAK with the reference value of 10 mGy. 19.03 : Optimization of protection in Mammography

  8. Calibration of output (I) (Purpose) • This method of determining ESAK relies on the calibration of the radiation output of the X-ray machine together with a recording of the tube loading on a series of patients • It is suitable for X-ray machines which have an AEC system and a post-exposure display or for units with manual exposure control only 19.03 : Optimization of protection in Mammography

  9. Calibration of output (II) (Test equipment) • Select a dosimeter with a dynamic range from at least 0.5 to 100 mGy • Accuracy > ± 10% • Precision > ± 5 • Calibrate the dosimeter in terms of air kerma free-in-air at an HVL as close as possible to 0.4 mm of Al. • Use post-exposure technique readouts (mAs) 19.03 : Optimization of protection in Mammography

  10. Calibration of output (III) (test method - patient measurements) • Select 10 patients with a compressed breast thickness in the range of 4 to 6 cm • For the same view on each patient, record the mAs after exposure value or the value set on units with manual exposure control 19.03 : Optimization of protection in Mammography

  11. Calibration of output (IV) (test method - output measurements) • Place the dosemeter at the reference point (45 mm above the cassette table, 60 mm from the chest wall side and laterally centred). The compression plate should be in place • Set manual exposure mode (same kV, anode, filtration, etc.) as used clinically • Record the air kerma readings per mAs for the exposure that covers the range of values recorded on patients. 19.03 : Optimization of protection in Mammography

  12. Calculations (I) • Calculate the mean value of the mAs for the 10 patients • Determine the output, i.e., the air kerma/mAs • Calculate the mean value of ESAK corresponding to the mean value of tube loading • Calculate the mean ESD by multiplying the mean ESAK by the appropriate backscatter factor if the HVL is known 19.03 : Optimization of protection in Mammography

  13. Calculations (II) If the information on the HVL is lacking, apply a backscatter factor of 1.09 Backscatter factor as a function of HVL (Jansen et al. 1994) 19.03 : Optimization of protection in Mammography

  14. Assessment of results • Compare the mean value of ESAK with the reference value of 10 mGy • If the mean value of ESAK exceeds 10 mGy per view, it is necessary to investigate the reason and corrective actions must be taken 19.03 : Optimization of protection in Mammography

  15. IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology Part 19.03 : Optimization of protection in Mammography Topic 2: Entrance surface dose (measurements with phantom)

  16. Entrance surface dose (measurements with phantom) • This method is an alternative to the method of placing TLDs on patients where this is not deemed to be acceptable due to interference with patient examination • It is intended for facilities where local dosimetry resources and expertise are not available 19.03 : Optimization of protection in Mammography

  17. Tests equipment (provided by a central laboratory) • TLDs: calibrated in terms of air kerma free-in-air at a HVL as close as possible to 0.4 mm Al. • Standard phantom: PMMA (thickness 45  0.5 mm, 150 x 240 mm2) • Developed film strip with labelled optical densities • Questionnaire 19.03 : Optimization of protection in Mammography

  18. Test method (1) • Set up the x-ray equipment for a cranio-caudal view with the compression plate present and a cassette loaded in the bucky • Position the phantom on the breast table, make sure that it completely covers the AEC device • Place the TLD at the reference point 19.03 : Optimization of protection in Mammography

  19. Test method (2) • Expose the phantom to the same conditions as used clinically for a standard-sized breast • Process the film in the normal way • Verify that the film density (base + fog included) at the reference point is in the optical density range of 1.20 to 1.80 (densitometer measurement or visual comparison with the film strip provided by the laboratory) 19.03 : Optimization of protection in Mammography

  20. Test method (3) • Replace the TLD with a second one • If the film optical density is in the range 1.20 to 1.80 make a second phantom exposure aiming for a density of 1.50 • Finally process a fresh, unexposed film • Estimate the density of the background film by visually comparing with the density with a densitometer or with the test strip 19.03 : Optimization of protection in Mammography

  21. Test method (4) • Complete the questionnaire provided by the central laboratory • Return the TLDs together with the three films and the questionnaire to the issuing laboratory • Compare the value of ESAK provided by the laboratory to the limiting value of 11 mGy corresponding to a net optical density of 1.00 (for different values of optical density, see table) 19.03 : Optimization of protection in Mammography

  22. Limiting values for ESAK as a function of net optical density 19.03 : Optimization of protection in Mammography

  23. Calibration of output (Purpose) • This method of determining ESAK with a standard phantom gives information on compliance of the technical set-up of the mammographic equipment with guidelines on radiation dose. • It is suitable for X-ray machines which have an AEC system and a post-exposure display or for units with manual exposure control only 19.03 : Optimization of protection in Mammography

  24. Calibration of output(Test equipment) • A dosemeter with a dynamic range from at least 0.5 to 100 mGy • Accuracy > ± 10% • Precision > ± 5 • The dosimeter should be calibrated in terms of air kerma free-in-air at an HVL as close as possible to 0.4mm of Al. • A standard phantom: PMMA block of 45 mm; 150x240 mm2 • A densitometer (resolution of 0.01 density units) 19.03 : Optimization of protection in Mammography

  25. Test method (determination of mAs (1)) • Set up the x-ray equipment for a cranio-caudal view with the compression plate present and a loaded cassette in the Bucky • Position the phantom on the breast table, make sure that it completely covers the AEC device • Expose the phantom at clinically used technique 19.03 : Optimization of protection in Mammography

  26. Test method (determination of mAs (2)) • Record the mAs • Process the film • Measure the optical density and verify that it is in the range 1.20 to 1.80 • If necessary, adjust the AEC density setting to achieve a suitable optical density and repeat the procedure described above 19.03 : Optimization of protection in Mammography

  27. Test method (measurement of output) • Remove the phantom and place the dosemeter at the reference point. The compression plate should be in place • Set manual exposure mode and mAd determined previously • Record the dose measurement. If it is not possible to select the exact mAs, make two exposure that bracket this value 19.03 : Optimization of protection in Mammography

  28. Calculations • Apply the appropriate calibration factor to the dose measurement • Interpolate the results if it was necessary to make two bracketing exposures • Express the result as ESAK 19.03 : Optimization of protection in Mammography

  29. Assessment of results • Compare the mean value of ESAK with the limiting value of 11 mGy corresponding to a net optical density on the film of 1.00. For optical densities deviating from this value see the table • If the mean value of ESAK exceeds 10 mGy per view, it will be necessary to determine the cause and corrective must be taken 19.03 : Optimization of protection in Mammography

  30. Entrance surface dose Limiting value : £10 mGy for 40 mm PMMA, £12 mGy for 45 mm PMMA, £20 mGy for 50 mm PMMA. Frequency : Annually Equipment :Dosimeter, PMMA block 150x240 mm2, densitometer 19.03 : Optimization of protection in Mammography

  31. IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology Part 19.03: Optimization of protection in Mammography Topic 3: The average glandular dose

  32. Determination of average glandular dose (AGD) • The AGD cannot be measured directly. It is derived from measurements of the ESAK (with a standard phantom) and HVL. The ESAK is used with tabulated conversion factors (derived from Monte Carlo calculations) to determine the AGD. 19.03 : Optimization of protection in Mammography

  33. Average Glandular Dose Table 1: Conversion factors gPB for calculating the AGD for a 50 mm « standard breast DGS » from the ESAK (Ka) measured at the technique as for 45 mm standard PMMA phantom (Dance 1990) 19.03 : Optimization of protection in Mammography

  34. Limiting values for AGD as a function of net optical density Table 2 19.03 : Optimization of protection in Mammography

  35. Calculations • Derive the HVL by interpolation. This may be done by plotting the logarithm of the dose measurements against the relevant Al absorber thickness • Derive, if necessary by interpolation, the conversion factor gPB for the measured HVL from table 1. Multiply the measured ESAK, Ka, at the mAs determined for the correct exposure of a standard phantom and the relevant gPB to obtain the standard AGD DGS (breast) = gPB x ka 19.03 : Optimization of protection in Mammography

  36. Calculations (example) • An exposure of the standard phantom in the AEC mode, at 28 kV and Mo-Mo anode-filter combination requires 94 mAs • 0.080 mGy/mAs (average of values measured at 90 and 100 mAs) was determined from the calibration of the output at the reference point • The measured HVL was 0.32 mm Al, yielding a gPB value of 0.187, derived by interpolation from Table 1 19.03 : Optimization of protection in Mammography

  37. Calculations (example) • The standard AGD (DGS) is: DGS = 94 [mAs] • 0.080 [mGy/mAs] • 0.187 [mGy/mGy] DGS = 1.4 [mGy] 19.03 : Optimization of protection in Mammography

  38. Assessment of results • Compare the value of standard AGD with the limiting value of 2.3 mGy, derived from the ESD limiting value for a film optical density of 1.00 • Table 2 provides information for other film densities 19.03 : Optimization of protection in Mammography

  39. Where to Get More Information European protocol for the quality control of the physical and technical aspects of mammography screening. http://euref.org/index.php?option=com_phocadownload&view=category&id=1&Itemid=8 American College of Radiology Mammography Quality Control Manual, Reston VA, 1999. 19.03 : Optimization of protection in Mammography

More Related