580 likes | 774 Views
Clearview COMPLETE HIV 1/2 Rapid Test How to Run and Read. Overview. Background Presentation D emo of the test You will practice running 2 tests We will not be testing anyone in the room Proficiency Test - You will run 5 tests without assistance R un, read, and document 5 tests
E N D
Overview • Background Presentation • Demo of the test • Youwill practice running 2 tests • We will not be testing anyone in the room • Proficiency Test - You will run 5 tests without assistance • Run, read, and document 5 tests • Read results of 12 test images • Pass / fail course, must get 100% to pass
Intro to Clearview COMPLETE Rapid Test • One-step test • Visual interpretation • Uses blood • Looks for HIV antibodies • Results within 15 minutes
In California…. • Individuals who have been trained by CDPH/OA and are working in a OA-funded HIV testing site can run HIV/Hep C rapid tests. OR • Are working in an HIV testing site that meets these two criteria: • Utilizes HIV counseling staff who are trained by OA or its agents and • Has a quality assurance plan (QA) approved by the local health department in the jurisdiction where the site is located and has HIV testing staff who comply with specific QA requirements.* • They must also be certified to perform finger sticks (or be occupationally exempt, like nurses) * QA requirements are specified in Section 1230 of Title 17 of the California Code of Regulations
California State Law • State law allows certified HIV test counselors to conduct finger sticks for the purpose of collecting blood samples for HIV and hepatitis C testing • This law does not allow you to do venipuncture
Package Insert • Contains instructions • CLIA requires following the manufacturer’s instructionsto the letter!!
Subject Information Notice • Contains information about the test that is relevant to the client • Package Insert says, “All subjects must receive the Subject Information Notice” • Check in with your site to see how this is handled
Test Accuracy • We’re going to talk about how well the test works • You do not need to memorize what I’m saying - only need to know test is very accurate • There are two components of test accuracy: • Sensitivity • Specificity
Specificity • When a test finds something, it should be the right thing • Tests ability to detect a true negative
SpecificityTable 6 page 16 of package insert 1 time Clearview COMPLETE read (+) when client was (-) 9 times lab EIA tests read (+) when client was (-) Calculated specificity of Clearview COMPLETE assay in these studies is 99.9% EIA was 99.3%
Sensitivity • The ability of a test to find what it’s looking for and not miss anything. • Test’s ability to detect a true positive
SensitivityTable 1 page 9 of package insert Clearview COMPLETE missed 2 positives EIA missed 0 positives (however EIA read 5 positives that were actually negatives) Calculated sensitivity of the Clearview COMPLETE assay in these studies is 99.7%.
Combining Specificity and Sensitivity • An ideal test would find the right thing (specificity) and not miss anything (sensitivity)
Think of a Tuna net…. A net that gets all the tuna and none of the dolphins
Because Test is Highly Sensitive… • We do not need to confirm negatives • We do, however, always need to confirm reactive results
2 Different Tests ClearviewCOMPLETE 1stTest _ + + Negative, please come back and see us in 6 months OraQuick Test Reactive, 2ndTest _ + Both tests were reactive your next step is to see a doctor and we can help you with that Lab Testing
With Clients… • Emphasize that the test is extremely accurate • De-emphasize statistics and percentages • For example “This test is highly accurate”
Quality AssuranceRequirements • QA are practices and procedures which ensure that every client receives an accurate test result • QA reduces human error as much as possible
Components of Quality Assurance • Personal and logistical characteristics • External controls • Lab space • Universal precautions
Must Have… Steady hand Good eye sight Organizational skills Adequate lighting • Full/bright light, task light • Do not use a flash light
External Controls • fluids made from human plasma • Biohazards – wear gloves! • Negative, Positive for HIV-1, Positive for HIV-2
External Controls are Used for: • Training, we will use them today • Determining if test is working properly • Determining if lighting is adequate • Determining if the test reader has sufficient eye site
HIV 1 vs HIV 2 • Two different strains • Both are transmitted the same way • HIV-2 is less infectious • HIV 2 progresses more slowly • HIV 2 found in West Africa and is rarely found elsewhere
When to Run External Controls • New operator • New test kit lot or shipment • Two invalid results in a row • Out of temperature range in testing area or in storage area • Every 40 tests or once a month (whichever comes first)
How to Run Controls Use Accessory Kit (#92115) • Use a fresh pipette, collect control from vial • Transfer control to a fresh weigh boat • Collect control from weight boat using sampling tip of device
Temperature Control • Perform test, • 64-86 degrees • If testing temperature is out of required temperature range stop testing • If out of temp range, run controls before proceeding • Store test: 46-86 degrees
Lab Space • A space for undisturbed test processing • Your site is a lab • Separate from counseling area • No smoking, eating or drinking
Universal Precautions The universal practice of avoiding contact with patients' bodily fluids, (blood) by means of the wearing of nonporous articles such as medical gloves
Gloves • Wear them when handling blood or blood products • TODAY – whenever you touch the control fluid vials!!!! • How often do • you change them? • How do you remove them?
Sharps: Handling & Disposal • Medical instruments that are used to puncture the skin (syringes, lancets, needles) • Dispose of sharps immediately, • in a hard redplastic bio bin! • Do NOT dispose of sharps • in a red bio hazard bag
Biohazard Bags Handling & Disposal • Bandages, used cotton and gauze, and gloves with body fluids on them are bio hazardous waste • By law, if fluid cannot be squeezed out of the cotton, gauze, etc., the waste item can be disposed of in regular trash.
For Today’s Training • Place used test kits in the sharps container • Place gloves and all other trash in the brown paper bags
Paperwork Needed • Expanded Checklist • Short Checklist (Competency Assessment Test) • Rapid testing log or lab slip • Lab stickers • HIV Testing Form • These forms help make sure samples are not mixed up
TestingDevice • Control area • Test area • Blue line • Test stand • Buffer cap
Remove Buffer Cap • Remove buffer cap from test device • Place buffer cap in stand
Collect Blood Sample • After performing finger stick, wipe away first drop of blood. • Touch device’s blood sample tip to second drop of blood • Blood will travel into the sample tip
Push Device Tip into Buffer Cap • You will feel 3 snaps: • Snap 1: through foil • Snap 2: into cap • Snap 3: fully seated (blue line is visible through clear area of the stand)
Confirm Device is Fully Seated If you do not see pink/purple flow within 3 minutes, Push Again! (then start timer)
Trainers Demonstration • Trainer Reads • Other Trainer preforms test • Participants follow along (Expanded Checklist)
Participants’ First Practice • No food or drink • Listen to detailed steps: • Only do what we tell you to do • Do all steps in order • If you finish step quickly, wait for the next step
Reading Time • Negative results in 15-20 minutes • Reactive results as soon as clear • test line and control line develop
What causes invalids? • Human error (e.g., no specimen) • Unknown • Manufacturer error • If you ever have an unusual result, do not deliver it
What does it mean if a client has an invalid test result?What does the darkness of the lines mean? NOTHING
Participants’ Second Practice • Pair up • One person run a test at their own pace, using the short checklist • Partner observe them and help only as needed (e.g., if you see any steps that were missed, say something) • Switch • Use the checklist!!!!
Internal Control – The “C” Line • The control line is the “C” line • Internal control tells us: • Specimen was adequately applied • Proper hydration • Migration of reagents past the “T” zone. Internal & external controls are standard lab practice – not a sign of test kit unreliability