1 / 4

XTAX Repairs

XTAX Repairs. Bilan dosimetrique. Rappel : Doses individuelle maximale : 420uS Dose individuelle max actuelle : 382uS . Bilan reparation table aval : Ambiance dans le bunker : Dose ambiante dans le bunker : Entre 7 et 12uS/h Dose au contact : 20uS/h

khanh
Download Presentation

XTAX Repairs

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. XTAX Repairs Bilandosimetrique

  2. Rappel : Doses individuellemaximale : 420uSDose individuelle max actuelle : 382uS

  3. Bilan reparation table aval : Ambiance dans le bunker : Dose ambiante dans le bunker : Entre 7 et 12uS/h Dose au contact : 20uS/h Dose au contact du moteur 450uS/h Dose au contact des vérins 750uS/h Remplacementmoteur Mecaniciens : 10uS Coach : 10uS Remplacementsoufflet : 2 Personnes : 2-3uS/personne/soufflet Total : 48uS (estimation ALARA 36.4uS) Bilan reparation table amont : Ambiance dans le bunker : 40uS/h Remplacementmoteur Mecaniciens : 13-15uS Coach : 16-17uS Remplacementsoufflet : 2 Personnes : 2-3uS/personne/soufflet Total : 53uS (estimation ALARA 300.4uS) Evacuation des soufflets in-situ : Environ 60uS pour les 3soufflets. (estimation ALARA 166.6uS)

More Related