1 / 12

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology. RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. Part 19.10: Optimization of protection in mammography Practical exercise. Contents.

kieran-west
Download Presentation

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION INDIAGNOSTIC ANDINTERVENTIONAL RADIOLOGY Part 19.10: Optimization of protection in mammography Practical exercise

  2. Contents • To be able to apply quality control protocol to mammography equipment • To measure the AEC performance • Optical density control setting • Short term reproducibility • Long term reproducibility • Object thickness compensation • Tube voltage compensation 19.10: Optimization of protection in mammography

  3. IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology Part 19.10: Optimization of protection in mammography AEC (Automatic Exposure Control)

  4. AEC system • The performance of the Automatic Exposure Control (AEC) system can be described by the reproducibility and accuracy of the automatic optical density control under varying conditions, e.g., different object thicknesses and tube voltages. • An essential prerequisite for these measurements is a stable operating film-processor and the use of the reference cassette. 19.10: Optimization of protection in mammography

  5. Equipment needed • PMMA: plates 20 x 300 x 300 mm • Densitometer • Reference cassette • Dosimeter 19.10: Optimization of protection in mammography

  6. Optical density control setting: central value and difference per step • Produce exposures with a breast equivalent phantom material with varying positions of the optical density control selector • A target value for the mean optical density should be established according to local preference. 19.10: Optimization of protection in mammography

  7. OD control setting: central value and difference per step Limiting value : The film density (base + fog included) at the reference point should remain within ± 0.20 in density of the target value. Target value is typically in the range: 1.30-1.80 (base + fog included) Thedesirablecontrol step difference is 0.10; £ 0.20 per step isacceptable Adjustable range > 1.00 Frequency :Step size: every six months Central density: daily 19.10: Optimization of protection in mammography

  8. (AEC) - Short term reproducibility • Position the dosimeter in the x-ray beam but without covering the AEC-detector. • The short term reproducibility of the AEC system is the deviation of the radiation dose for five exposures with a breast equivalent phantom Limiting value : The deviation must be < ± 5%; desirable would be < ± 2%. Frequency : Every six months 19.10: Optimization of protection in mammography

  9. (AEC) - Long term reproducibility • The long term reproducibility can be assessed from the measurement of the film optical density resulting from the exposure of a breast equivalent phantom Limiting value : < ± 0.20 optical density acceptable; < ± 0.15 desirable Frequency : Daily 19.10: Optimization of protection in mammography

  10. Object thickness compensation (AEC) • Compensation for object thickness is measured by exposures of PMMA plates in the thickness range 20 to 40 mm, using the clinical setting for the AEC Limiting value : All density variations must be within ± 0.20 of the target density. Desirable : ± 0.15 Frequency : Weekly 19.10: Optimization of protection in mammography

  11. Tube voltage compensation (AEC) • Compensation for tube voltage is measured over the clinical range of kV, filters, and target materials Limiting value : All density variations must be within ± 0.15 of the target density; Desirable : ± 0.10 Frequency : Every six months 19.10: Optimization of protection in mammography

  12. Where to Get More Information European protocol for the quality control of the physical and technical aspects of mammography screening. http://euref.org/index.php?option=com_phocadownload&view=category&id=1&Itemid=8 American College of Radiology Mammography Quality Control Manual, Reston VA, 1999. 15.3: Optimization of protection in radiography

More Related