An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. More digital 244 wk 12 Perry Sprawls, Ph.D.sprawls@emory.eduProfessor Emeritus Department of Radiology Emory University School of Medicine Atlanta, GA, USACo-DirectorCollege on Medical PhysicsInternational Centre of Theoretical PhysicsTrieste, ItalyDirector Sprawls Educational FoundationMontreat, NC, USA

  2. Three possibilities include processing methods to: Reduce image noise Increase visibility of detail Adjust and optimize the image contrast characteristics. Some type of digital image processing is used with most of the medical imaging modalities. Processing of digital images can be used to change most image characteristics.

  3. There are several digital processing methods that can be used to adjust the contrast characteristics of an image.  The two methods that we will consider here are: Look Up Table (LUT) processing Windowing Both of these are used in digital radiography as well as with many of the other imaging modalities.

  4. pixels

  5. Here we are using a very simple image to show how digital processing can be used to change the image contrast. As we see, the image consist of a background area and a small square object in the center.  In the low contrast image on the left the background area has pixel values of 40 and the object in the center has pixel values of 30. The numerical contrast (the object relative to the back ground) is the difference (40-30=10). Look up tables (LUT) are data stored in the computer that is used to substitute new values for each pixel during the processing. In our example here, we are keeping it simple and working an image with only two pixel values, 40 and 30. As we see here, the processing uses a LUT that substitutes a 90 for a 40 and a 10 for a 30.  The effect of this is to increase the image contrast (90-10=80). As we are about to discover, it is usually possible for the user to select from a variety of LUTs, each one designed to produce specific contrast characteristics.

  6. Let's recall that a LUT indicates what number is to be substituted for each pixel value during the processing of the image. It is very helpful to show this relationship between the original pixel values and the new values with a graph as we see here. Here we are starting with a simple straight-line or linear graph that shows that the substituted number is the same as the original image pixel value. Processing with this LUT does not change the image, it just introduces us to the concept of LUTs.

  7. If LUT processing is to change the contrast characteristics of an image it must substitute numbers that are different from the original pixel values. Here we see that relationship shown as a curve.  Note that the values and positions on the scale for the three pixel values represented by the colors ( yellow, green, and red) are now different in the processed image. You might recognize the shape of this curve as being similar to that of a radiographic film characteristic (H & D) curve.  That is intentional.  By processing a digital radiograph with a LUT with this characteristic it then takes on some of the familiar contrast "look" of a radiograph recorded on film

  8. One of the advantages of digital processing is that the processing parameters (factors) can be selected to produce images with different contrast characteristics. Here we see curves representing three different LUTs. Like the characteristic curve of film, the slope of the curve at every point represents how the contrast will be changed by the processing.  Where the curve has a high slope (steep) the contrast will be increased by the processing.  If the slope is small (less than 450) the contrast will be decreased by the processing. The curve with the greater slope shown here will produce an image with high contrast.  The one with less slope simulates the latitude type film that is used for chest imaging.  We will see the effect of processing with the inverted curve a little later

  9. In the typical digital radiographic system, a variety of LUTs are installed.  The appropriate LUT is then automatically selected to give the desired contrast characteristics to match the type of procedure (chest, extremity, etc) that is designated by the operator.

  10. If he original image recorded with most digital radiography systems is displayed directly, it might have relatively low contrast as we see with the image on the left.  This is because of the wide dynamic exposure range and the linear response of the digital radiographic receptors.  This is actually a desirable characteristic because a wide range of exposure is recorded and the digital receptor does not suppress or cutoff the contrast in some of the exposure range as happens with images recorded directly on film. The usual procedure is to use processing, as illustrated here, to increase the contrast for some selected portion  of the exposure range.  The processing illustrated here produces an image that is similar to one recorded with a high contrast film.  Note: we do not generally use a high contrast film for chest radiography because it produces an image with too much area contrast (like the dark lungs shown here).   

  11. Here we see LUT processing that is more appropriate for chest radiography. In general, it simulates an image recorded on a latitude type film that is used for chest imaging. Compared to the high contrast processing that we have just seen, this produces an image with good object and anatomical structure contrast and visibility throughout most of the image. 

  12. Many radiologist find value in having an inverted (black bone) image in addition to the conventional (white bone) image at the time of viewing. Some objects and structures might be more visible and easier to detect on the inverted image. The inversion can be achieved by processing with a LUT as illustrated here.

  13. The ability to window is a valuable feature of all digital images. Windowing is the process of selecting some segment of the total pixel value range (the wide dynamic range of the receptors) and then displaying the pixel values within that segment over the full brightness (shades of gray) range from white to black. Important point...Contrast will be visible only for the pixel values that are within the selected window.  All pixel values that are either below or above the window will be all white or all black and display no contrast. The person controlling the display can adjust both the center and the width of the window.  The combination of these two parameters determine the range of pixel values that will be displayed with contrast in the image.

  14. One of the advantages of windowing is that it makes it possible to display and enhance the contrast in selected segments of the total pixel value range.  This can be compared to the limitations of images displayed on film where the full range of exposure is displayed in one image and cannot be changed. With windowing we can create many displayed images, each one "focusing on" a specific range of pixel values. As we see here, when the window is set to cover the lower segment of total pixel value range, we see good contrast in the lighter areas like the medistimum. Setting the window to the higher segment produces good contrast in the darker areas like the lungs.

  15. The Blurred Mask Subtraction is often used, especially in digital radiography, to enhance the visibility of detail in certain clinical procedures. Important point....This process does not un-blur an image and recover detail that was completely lost because of blurring from the focal spot, motion, and the receptor.  What it does do is increase the visibility (contrast) of some objects, especially where the visibility of the objects is somewhat limited by large area contrast as in chest imaging. The process consist of two distinct steps.  First, a blurred copy of the original image is produced.  A common form of digital processing that can be used to produce a blurred image is just by replacing each pixel value with the average of  the pixel values in it's neighborhood.  This produces a "blurred mask" image. The blurring removes all detail from the image (that is what blurring does!).  We now have two images.  The original image contains the general large-area contrast background plus some detail.  The blurred mask image contains only the large-area contrast background The final step is where the computer subtracts the blurred mask image (actually some fraction of it) from the original image. This process reduces the large-area contrast background in relation to the contrast of the detail.  The result is that the contrast and visibility of the detail (small objects and structures) in enhanced. Enhancing Visibility of Detail