An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. The Problem of Secondary Fluorescence in EPMA in the Application of the Ti-in-Zircon Geothermometer And the Utility of PENEPMA Monte Carlo Simulations John H. Fournelle Eugene Cameron Electron Microprobe Lab Department of Geology and Geophysics University of Wisconsin Madison, Wisconsin

  2. Zircon…takes a licking, keeps on ticking… Zircon (ZrSiO4) is a small mineral highly resistant to chemical changes; this robustness has led it to be used for estimating earth conditions eons ago. It has been used to date geologic events, measuring its radiogenic Pb, U and Th isotopes. The oldest dated mineral on Earth is a zircon from Australia, and its oxygen isotope value suggests Earth’s crust was cool and wet as long ago as 4.3 billion years.

  3. Zircon…takes a licking, keeps on ticking… • ZrSiO4, small ~100-200 microns, accessory mineral (~granites, rhyolites) • highly resistant to chemical changes • Th, U and Pb present: radiogenic decay • used for estimating earth conditions eons ago • oldest dated mineral on Earth is a zircon from Australia • its oxygen isotope value suggests Earth’s crust was cool and wet as long ago as 4.3 billion years.

  4. Taking the Earth’s temperature Scan slide of ? At Kileaua re Therm • geologists can directly measure temperature of geothermal springs and lava flows • but direct methods are not possible for most of Earth conditions … • use various geochemical evidence -- element partitioning between coexisting minerals -- to infer conditions deep within the earth. • these geothermometers and geobarometers have been developed with, and utilize, electron and ion probes.

  5. A zircon thermometer? Watson and Harrison* and Watson et al** experimentally determined that the amount of Ti incorporated in zircon (~1 to 1000s of ppm), coexisting with a high-Ti mineral (e.g., rutile TiO2 or ilmenite FeTiO3), was proportional to the temperature at which the zircon crystallized and could be used as a geothermometer. * Watson and Harrison, 2005, Science, 308, 841 ** Watson, Walk and Thomas, 2006, Contrib Mineral Petrol, 151, 413

  6. EPMA SIMS Measuring tiny levels of Ti ….. Where the level of Ti is very low (1-100 ppm), the preferred method is ion probe (SIMS). However, there are situations where EPMA is used: for original validation of Ti in synthesized zircons (i.e., for SIMS standards),

  7. EPMA SIMS Measuring tiny levels of Ti ….. Where the level of Ti is very low (1-100 ppm), the preferred method is ion probe (SIMS). However, there are situations where EPMA is used: for original validation of Ti in synthesized zircons (i.e., for SIMS standards), by geologists who do not have ready access to SIMS, or

  8. EPMA SIMS Measuring tiny levels of Ti ….. Where the level of Ti is very low (1-100 ppm), the preferred method is ion probe (SIMS). However, there are situations where EPMA is used: for original validation of Ti in synthesized zircons (i.e., for SIMS standards), by geologists who do not have ready access to SIMS, or where the large SIMS spot size (~25 microns) is prohibitive.

  9. Secondary Fluorescence may become an issue • the electron beam’s interaction volume is small relative to the volume excited by both characteristic and continuum x-rays generated • what is the amount of secondary fluorescence (SF) that is measured during EPMA??? • critical for trace element EPMA measurements

  10. One alternative: model the effect The PENEPMA Monte Carlo program, based upon PENELOPE, has been shown to accurately predict the extent of secondary fluorescence and has been recently modified to explicitly state SF intensity.

  11. 2 Cases Modeled with Penepma Both 15 kV, 40° take off angles, with only continuum secondary fluorescence Rutiles (TiO2) in experimental runs for Ti in zircon solubility Zircons in rock samples with nearby ilmenite and biotite (= Ti-bearing minerals)

  12. Watson et al (2006) experimentally grew zircons where rutile (TiO2) were also present, and tried to account for secondary fluorescence….

  13. 7 Geometries Modeled

  14. 7 Experimental Geometries Modeled Case 1 examines the potential for SF in experimentally grown zircons that have crystals of rutile either touching them or present in the vicinity. Geometry 1: 30 um zircon, no rutile, only Ti in surrounding glass (6 wt%) => 452 ppm SF Ti

  15. Geometry 2: 30 um zircon, 5 large rutiles 15 um away, Ti in surrounding glass (6 wt%) => 948 ppm SF Ti

  16. Geometry 3: 30 um zircon, 5 large rutiles 15 um away, NO Ti in surrounding glass, with Pb instead => 390 ppm SF Ti

  17. Geometry 4: 30 um zircon, 5 large rutiles 15 um away, in epoxy => 1179 ppm SF Ti

  18. Geometry 5: 30 um zircon, 5 large rutiles 60 um away, in Pb-glass => 25 ppm SF Ti

  19. 7 Experimental Geometries Modeled Geometry 6: 30 um zircon, 5 tiny 4 um rutiles 0-1 um away, in epoxy => 61 ppm SF Ti

  20. Geometry 7: 30 um zircon, 10 tiny rutiles 0 um away, in epoxy => 120 ppm SF Ti

  21. Conclusions from Penepma 7 geometries • easy to get several hundred ppm of Ti if large Ti-rich phases are within tens of microns of analysis point • conceptually, “solid angle” that Ti-rich phases present to analysis point • continuum SF is not insignificant

  22. Case 2 examines the potential for SF in a rock where zircons are surrounded by ilmenite, hematite and biotite (suggested by C. Morisset, Univ. British Columbia). Here it is less easy to model a realistic geometry as the zircons are irregular in shape.

  23. Complications of Secondary Fluorescence in EPMA: The “Size Discrepancy Issue”

  24. Electron beam Electron beam Standard not to scale with unknown, would be much larger if true scale. Difference between small sample and large standard Sample = 10 um polished sphere Cr2O3 embedded in plastic Standard = 2 mm polished sphere PMM (plastic) PMM (plastic) In troubleshooting low totals, the question arose: if there is a several order magnitude size difference between unknowns (small grain separates) and standard (large), what could result?

  25. Electron beam Electron beam Difference between small sample and large standard Sample = 10 um polished sphere Cr2O3 embedded in plastic Standard not to scale with unknown, would be much larger if true scale. Standard = 2 mm polished sphere PMM (plastic) PMM (plastic) If the primary electron “interaction volume” is confined within the material, and therefore the primary x-ray generation is also confined therein, is the lack of “additional” Cr x-ray counts resulting from secondary fluorescence outside the primary electron interaction volume of any importance in “normal” EPMA???

  26. z Sample = 10 um polished sphere embedded in plastic Standard = 2 mm polished sphere Cr2O3 PMM (plastic) Set up a PENELOPE Monte Carlo simulation: Standard of “huge size”, 2 mm Unknowns of much smaller size Accelerating voltage of 20 keV, takeoff angle 40°

  27. 4 Yes, Secondary Fluorescence can cause problems Standard=2000 mm Cr2O3 Unknown = smaller Cr2O3 Electron range (K-O): 1.7 micron Cr Ka X-ray range (A-H): 1.6 micron A 100 mm grain of pure Cr2O3 will have 1% low Cr K-ratio, and a 10 mm grain will have a K-ratio 2.5% low. (plots show K-ratios produced in centers of various discrete sized diameter cut-off spheres imbedded in epoxy simulations)

  28. Conclusion Discrepancies in size between unknown and standard can lead to small, but noticeable errors, because secondary fluorescence yields • additional x-rays beyond the primary electron impact-x-ray production volume in the same phase if the phase is large, • or • a lack of additional x-rays if the phase is small and mounted in epoxy.

  29. PENELOPE • created to model high energy radiation in bodies of complex geometries • simulates x-ray generation and x-ray absorption/secondary fluorescence • a new version developed for EPMA, with EDS-like spectral output • a FORTAN program, runs with G77 compiler under OS X, Linux, Windows • developed by Salvat, Llovet et al. of Universitat de Barcelona … and free