An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Molecular Luminescence Spectroscopy Lecture 32

  2. Molecular Luminescence Spectroscopy (molecular emission spectroscopy) Luminescence spectroscopy is a technique which studies the phenomenon types of chemical systems: 1- Fluorescence: (needs source for excitation) 2- Phosphorescence: (needs source for excitation) 3- Chemiluminescence (chemical reaction take place): no source needed • The analyte or its reaction product needs to be luminescent. • Very sensitive up to 10-15 M for UV-Vis up to 10-8 M. • The relative luminescence intensity is related to analyte concentration as will be seen shortly.

  3. Singlet and Triplet States • Electrons in molecular orbitals are paired, according to Pauli exclusion principle. • When an electron absorbs enough energy it will be excited to a higher energy state; but will keep the orientation of its spin. Singlet transition (state): The molecular electronic state in which electrons are paired. Triplet transition (state): The molecular electronic state in which the two electrons are unpaired. • The triplet state is achieved when an electron is transferred from a singlet energy level into a triplet energy level, by a process called intersystem crossing; accompanied by a flip in spin.

  4. Multiplicity : (number of lines observed when the molecule is placed in a magnetic field) is related number of unpaired spins in the molecule (S):. M = 2•S + 1. Singlet

  5. In a singlet state: The spins of the two electrons are paired and thus exhibit no magnetic field and called diamagnetic. Diamagnetic molecules, containing paired electron, are neither attracted nor repelled by a magnetic field. In the triplet state: Molecules have unpaired electrons and are thus paramagnetic which means that they are either repelled or attracted to magnetic fields. The terms singlet and triplet stems from the definition of multiplicity where: Multiplicity = 2S + 1 Where, S is the total spin.

  6. For singlet state: The total spin for a singlet state is zero since electrons are paired which gives a multiplicity of one (the term singlet state): Multiplicity = (2 * 0) + 1 =1 For a triplet state: The total spin is one (the two electrons are unpaired) and the multiplicity is three: Multiplicity = (2 * 1) + 1 = 3 • It should also be indicated that the probability of a singlet to triplet transition is much lower than a singlet to singlet transition. • Therefore, the intensity of the emission from a triplet state to a singlet state is much lower than emission intensities from a singlet to a singlet state.

  7. Energy Level Diagram for Photoluminescent Molecules The following diagram represents the main processes taking place in a photoluminescent molecule when it absorbs and emits energy. The different processes will be discussed below:

  8. Excitation

  9. Optical Atomic Spectra We have briefly described in an introductory chapter that atomic spectra are usually line, rather than band, spectra due to absence of vibrational and rotational levels. The existence of quantized electronic energy levels explains the origin of the observed line spectra and exact locations of possible lines

  10. Absorption (Excitation) 1st step • The absorption of UV-Vis radiation is necessary to excite molecules from the ground state to one of the excited states. • Absorption of radiation promotes electrons in chemical bonds to be excited. • However, we have seen earlier that not all transitions have the same probability and while certain transitions are practically very important, others are seldom used and are of either no or marginal importance. • Therefore, from information we have discussed in Chapter 14 we concluded that there are four different types of electronic transitions which can take place in molecules when they absorb UV-Vis radiation. • A s-s* and a n-s* are not useful while • The n-p* transition requires low energy but the molar absorptivity for this transition is low and transition energy will increase in presence of polar solvents.

  11. The most frequently used transition is the p-p* transition for the following reasons: a.   The molar absorptivity for the p-p* transition is high allowing sensitive determinations. b. The energy required is moderate, far less than dissociation energy. c.  In presence of the most convenient solvent (water), the energy required for a p-p* transition is usually smaller. • Therefore, best molecules that may show absorption are those with pbonds or preferably aromatic nature as discussed earlier. • Absorption to higher excited singlet states requires a very short time (in the range of 10-14s).

  12. Vibrational Relaxation • Absorption of radiation will excite molecules to different vibrational levels of the excited state • This process is usually followed by successive vibrational relaxations (VR) as well as internal conversion to lower excited states (as small amount of heat).. • In cases where transitions occur to the first excited state, vibrational relaxation to the main excited electronic level will take place and/or an intersystem crossing (ISC) to the triplet state can occur.

  13. Fluorescence • After vibrational relaxation to first excited electronic level takes place, a molecule can return to the ground state by emission of a photon, called fluorescence (FL). • The fluorescence lifetime is much greater than the absorption time and occurs in the range from 10-7 to 10-9s. • As the lifetime in the excited state is increased, the probability of fluorescence will be decreased since radiationless deactivation processes may take place. • However, not all excited molecules can show fluorescence by returning to ground state and most return to ground state by losing excitation energy as heat or through collisions with other molecules or solvent.

  14. Internal and External Conversion Internal conversion (IC): Is a radiationless deactivation process whereby excited molecules return to the ground state without emission of a photon. This process lacks rigid understanding but seems to be the most efficient deactivation process in luminescence spectroscopy, since most molecules do not show fluorescence.

  15. However: Molecules with close electronic energy levels, to the extent that their vibrational energy levels of ground and excited states are overlapped, are believed to cause efficient internal conversion (radiationless deactivation).

  16. Dissociation and predissociation • Internal conversion can result in a phenomenon called predissociation (PD) (indirect): Where an electron relaxes from a higher electronic state to an upper vibrational energy of a lower electronic state. When the vibrational energy is large enough and is greater than the bond energy, bond rupture occurs in a process called predissociation. • Dissociation (Direct): should be differentiated from predissociation where dissociation involves absorption of high energy so that the molecule is directly promoted to a high energy vibrational level where bond rupture directly occurs.

  17. External conversion (EC): (also radiationless relax.) Is a process whereby excited molecules lose their energy due to: • Collisions with other molecules. or • By transfer of their energy to solvent or other unexcited molecules. Therefore, external conversion is influenced by: 1- Temperature (directly prop.). 2- Solvent viscosity (inversely prop.). 3- Solvent composition.

  18. Intersystem Crossing Electrons present at the first excited electronic level can follow one of three choices including: 1) Emission of a photon to give fluorescence. 2) Radiationless deactivation to ground state. 3) Intersystem crossing (ISC). • The process of intersystem crossing involves transfer of the electron from an excited singlet to a triplet state. • This process can actually take place since the vibrational levels in the singlet and triplet states overlap. • However, crossing of the singlet state to the triplet state involves a flip in electron spin in order to satisfy the triplet state.

  19. Intersystem crossing is facilitated by: 1) Presence of nonbonding electrons. 2) Presence of heavy atoms. 2) Presence of paramagnetic atoms or species also enhances intersystem crossing. • An electron in the triplet state can also cross back to the singlet state and can result in a photon as fluorescence but at a much longer time than regular fluorescence. • This process is termed delayed fluorescence and has the same characteristics as direct fluorescence except for the large increase in lifetime (approximately as phosphorescence).

  20. Phosphorescence Electrons crossing the singlet state to the triplet state with a flipped spin can also follow one of three choices including: • Returning to the singlet state (including a flip in spin). • Relax to ground state by internal or/and external conversion. Or • Lose their energy as a photon (phosphorescence, Ph) and relax to ground state with a second flip in spin to satisfy the singlet ground state. • As can be rationalized from the processes involved in collecting phosphorescence photons, this involves an intersystem crossing and two flips in spin. • This, in fact, requires a much longer time than fluorescence (10-4s to up to few s). • Therefore, the probability of phosphorescence, and hence the intensity of the phosphorescence spectrum, is very low due to high possibility of radiationless deactivation.

  21. Quantum Yield and Efficiency The quantum yield or efficiency of fluorescence is: The ratio of the number of fluorescing molecules to the total number of excited molecules. • For highly fluorescent molecules, a quantum efficiency approaching one can be obtained. fluorescein  = 1 means that every excited molecules deactivates by emitting a photon – such a molecule is considered a very good fluorophore . nonluminesceing molecule  = 0. • The quantum efficiency can be represented by the relation: F = kFL/(kFL + kISC + kIC + kEC + kPD + kdiss) - kEC = external conversion (S1 S0), - kIC = internal conversion (S1 S0) - kISC = intersystem crossing (S1 T1), - kPD = predissociation - kdiss = dissociation K = rate constant Phosphorescence quantum efficiency is defined in the same manner

  22. The transitions most important in luminescence spectroscopy are n-p* and p-p*. • However, fluorescence is encountered more often in molecules having p-p* transitions since this transition has a higher quantum efficiency in terms of: • a higher molar absorptivity and a shorter lifetime (10-7 – 10-9s) so small deact. occurs. • The n-p* transition which has a longer lifetime (10-5 – 10-7s) so large deact. occurs. • Once again, a longer lifetime: means a lower luminescent probability due to increased possibility of radiationless deactivation.

  23. Variables That Affect Fluorescence and Phosphorescence: Factors affecting fluorescence and phosphorescence include both environmental and structural factors. Some of the important factors are discussed below: Fluorescence and Structure • As indicated earlier, best luminescence is observed for molecules with p bonds and preferably those having aromatic rings due to presence of low energy p-p*. • However, some heterocyclic aromatic rings do not show fluorescence. • These include pyridine, furan, pyrrole, and thiophene

  24. The lack of fluorescence in such molecules is largely believed to be due to: • Existence of a low lying n-p* transition that rapidly converts the excited molecule to the triplet state and prevent fluorescence. • However, fusion of a phenyl ring to any of the above molecules increase the possibility of the p-p* transitions and thus increase the fluorescence quantum efficiency.

  25. Substitution of halogens to the aromatic ring: • Has important influence on the fluorescent signal where a decrease in fluorescence is observed with an increase in the atomic weight of the halogen and a subsequent increase in phosphorescence. • This is referred to as the heavy atom effect where promotion of intersystem crossing takes place. • In addition, substitution of a carbonyl or carboxylic acid groups decreased fluorescence due to enhancement of intersystem crossing.

  26. Effect of Structural Rigidity: • The nature of the chemical structure of a molecule in terms of flexibility and rigidity is of major influence on the fluorescence and phosphorescence signal. • Molecules that have high degree of flexibility will tend to decrease fluorescence due to higher collisional probability. • However, more rigid structures have lower probability of collisions and thus have more fluorescence potential. • Biphenyl has very low fluorescence quantum efficiency (~ 0.2) while fluorine has a quantum efficiency close

  27. In addition, some metal complexes have higher fluorescence efficiency than do the ligands; also as a result of increased structural rigidity. For example: The fluorescence intensity of 8-hydroxyquinoline is increased to a large extent in presence of zinc ions, due to more rigid complex formation.

  28. Effect of Solvent Nature: Solvents characteristics have important effects on luminescent behavior of molecules. Three main effects can be recognized: a. Solvent Polarity: A polar solvent is preferred as the energy required for the p-p* is lowered. b. Solvent Viscosity: More viscous solvents are preferred since collisional deactivation will be lowered at higher viscosities. c. Heavy Atoms Effect If solvents contain heavy atoms, fluorescence quantum efficiency will decrease and phosphorescence will increase.

  29. Effect of Temperature: • Higher temperatures result in larger collisional deactivation due to increased movement and velocity of molecules. • Therefore, lower temperatures are preferred. Effect of pH: • The pH of the solution is a very important factor that influences luminescence. For example: Aniline shows fluorescence while aniline in acid solution (anilinium ion) does not. • Most compounds luminesce in basic or slightly basic solutions while some show fluorescence in acidic medium. • It is therefore important to adjust the pH so that maximum luminescence intensity is obtained • The pH also affects the emission wavelength where usually a longer emission wavelength is observed at higher pH.

  30. Effect of Dissolved Oxygen • Dissolved oxygen largely limits fluorescence since it promotes intersystem crossing because it is paramagnetic. • However, dissolved oxygen affects phosphorescence more than it does to fluorescence. • Although one would think that as far as intersystem crossing is increased in presence of oxygen, phosphorescence is expected to increase. On the contrary, phosphorescence is completely eliminated and quenched in presence of dissolved oxygen. • This may be explained on the basis that the ground state of oxygen is the triplet state and it is easier for an electron in the triplet state to transfer its energy to triplet oxygen rather than performing a flip in spin and relax to singlet state. • Therefore, oxygen will be excited and what we really observe is oxygen emission rather than phosphorescence. • It is for this reason that oxygen should be totally excluded to be able to detect phosphorescence.

  31. Effect of Concentration on Fluorescence It can be undoubtedly said that the fluorescence is directly proportional to the amount of absorbed radiation where: F = k(Po-P) P = Po 10-A Substitution gives: F = kP0 (1-10-A) This relation can be expanded by Mc Lauren series giving: F = kP0 (2.303 A – (2.303 A)2/2! + (2.303 A)3/3! - (2.303 A)4/4! + ….)

  32. Only the first term is important at the very low concentrations used. The relation simplifies to: F = kPo (2.303 A) F = KPoebc Where K = 2.303k

  33. The relation F = KP0ebc implies the following: a. The fluorescence signal can be increased if the radiant power (P0)of the incident beam is increased. Therefore, always use more intense sources. b. The fluorescence signal is directly related to the molar absorptivity(e) and thus molecules of higher molar absorptivities are better fluorescers. c.  Fluorescence signal is directly proportional to path length. d. Fluorescence signal is directly proportional to concentration. This is different from relation between absorbance and concentration which is logarithmic; A = log Po/P especially at low conc. P=Po. e. The linear correlation between fluorescence and concentration is only true when the absorbance is less than 0.05 (generally working at low conc.).

  34. Deviation from Linearity between Fluorescence and Concentration Negative deviations from the linear relation between fluorescence and concentration may be observed in the following cases: a.   At absorbances higher than 0.05. b.  Self-quenching whereby excited molecules lose their energies by collision with other molecules or solvent. c. Self-absorption which occurs when an emission band overlaps with an excitation (absorption) band. In this case, emitted photons excite other molecules in the ground state which results in no net emission.

  35. Fluorescence Instruments Like any emission instrument, a fluorescence or phosphorescence instrument consists of: • a source. • wavelength selectors. • a sample cell. • a detector. • a signal processor. Two wavelength selectors are used: 1) The first is the excitation filter or monochromator which excites the sample. 2) The other is the emission filter or monochromator that separates the fluorescence or phosphorescence wavelength.

  36. Sources • We have seen earlier that the fluorescence signal is proportional to the radiant power of the source. • Therefore, it is very important to select a source of as high radiant power as possible. • In most instruments, a xenon arc lamp is usually the source of choice. • However, some instruments use lasers.

  37. Xenon Arc Lamp • A quartz envelope hosting two electrodes and xenon at high pressure (5-10 atm). • The discharge which takes place between the two electrodes excites xenon and produces a continuum in the range from 200-1000 nm. • The wavelength maximum occurs at about 500 nm. • The radiant power produced by a xenon arc lamp is very high which makes the lamp suitable for luminescence analysis. • Also, the coverage of the whole UV-Vis range adds to the assets of the lamp. • However, a very good regulated power supply is essential since any fluctuations of the radiant power will be directly reflected on the fluorescence signal.

  38. Lasers • We have seen examples of lasers early in this course. • However, it should be indicated that since lasers may have very high intensities, some molecules, which are otherwise non fluorescents, show good fluorescence and can thus be determined by fluorescence spectroscopy. • This is called laser induced fluorescence (LIF) and is a common technique in fluorescence spectroscopy. • In addition, one should remember that as the radiant power is increased, photodissociation may become a problem especially at shorter excitation wavelengths.

  39. 1. Fluorometers • When the wavelength selectors are filters, the instrument is called a fluorometer. • It is a simple instrument that is usually used for quantitative analysis. • The use of fluorometers implies that the excitation and emission wavelengths are predetermined by other means or are known from literature. • The detector is usually a sensitive photomultiplier tube. • A schematic of a simple fluorometer is shown below:

  40. To absorb any radiation in the same directions (large noises if measured here) Measuring at 90 deg. Give minimum noise

  41. Fluorescent microscope

  42. Unlike the cell used in UV-Vis absorption studies, the cell used in luminescence studies has the two transparent faces perpendicular to each other, rather than parallel. • This is important since luminescence is collected 90o to the incident beam. • This configuration decreases the noise and fully excludes interferences from the incident beam.

  43. 2. Spectrofluorometers • In this case, the excitation and emission wavelengths are selected using dispersive elements like gratings or prisms. • The same instrumental configuration as fluorometers is usually used. • The following schematic represents a basic configuration of a spectrofluorometer:

  44. Dual instrument: to overcome flactuations

  45. Excitation and Emission Spectra • The first step in a successful fluorescence measurement is to determine the excitation and emission wavelengths. • The excitation spectrum is determined by adjusting the emission monochromator at an arbitrary, but well selected, wavelength which is thought to be longer than the anticipated wavelength. • Keeping the emission monochromator at the preset wavelength, the excitation spectrum is recorded by scanning the excitation monochromator in a range less than the preset wavelength of the emission monochromator.