An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Pulsed Glow Discharge Mass Spectrometry: An Ionization Source forAerosol Analysis and Laser Sampling Farzad Fani-Pakdel Qualifying Examination University of Florida - Department of Chemistry Division of Analytical Chemistry

  2. Outline Introduction Aerosols Aerosolscharacterization Glow discharge Research objective Experimental Instrumental set-up Particle introduction into system Preliminary Results Laser Sampling Importance and application Cathode design

  3. Aerosols Suspension of fine solid or liquid particles (~3 nm -100mm) in a gas. Examples: • Atmospheric Aerosols: • Dust • Fog • Haze • Smoke • Sea spray particles • Biological Aerosols: • Viruses: 0.01 - 0.3 micron • Bacteria: 0.5 – 10 micron • Pollen: 5-100 micron Significance: • Air pollution (Acidic rain) • Atmospheric reactions • Material transfer • War generated particles • Allergy and sickness • Contagious diseases • Biological warfare

  4. Chemical Characterization of Aerosols Spectroscopy: non-destructive, surface only Nano sensors: non-destructive, selective, surface only Mass spectrometry: destructive, core and surface both possible Importance: chemical contamination or bio hazard in core hv Sampling Ionization Nano device

  5. Ideal Ionization Technique Soft ionization provides surface information, at the molecular level CI Intense ionization Provides surface and bulk Information, at atomic level ICP A moderate ionization like EI will result in fragmentation Pulsed glow discharge is an ionization source capable of ionization at all these levels at the same time.

  6. Glow Discharge Ionization Source Penning ionization by meta-stable argon, Electron ionization Anode (+) Cathode (-) Negative Glow Ar e Ar+ e Ar Ar+ Ar * Cu Cu+ e - +

  7. Pulsed Glow Discharge Fragmentation Plateau High energy electrons (ICP) Pre-peak Penning ionization Soft, ( CI ) After peak V time

  8. Schematics of a Glow Discharge -Time of Flight Instrument Majidi, et al., Anal. Chem., 2003, 75, 1983-1996

  9. Research Objectives: Application of pulsed GD - TOF Mass Spectrometry for characterization of different particles bulk vs. surface chemical analysis Silica nano particles with peptides on surface Gold nano particles with hydrocarbon chains CdSe / ZnS nano particles MgCl2, NaCl mixed nano particles

  10. Experimental Setup GD TOF Figure is modified from the original: Eric Oxley, Ph.D. Dissertation, Department of Chemistry, University of Florida, 2002.

  11. Glow Discharge Source O-ring Cathode Eric Oxley, Ph.D. Dissertation, Department of Chemistry, University of Florida, 2002.

  12. Particle Introduction 1

  13. Particle Introduction 2

  14. Particle Introduction 3

  15. Particle Introduction 2; Cathode Design Cathode base Scale in inches Insulator Anode

  16. Preliminary Results Tap water Na-Mg solution DI water To examine the sample introduction design, a sodium chloride-magnesium nitrate solution(1000ppm of each) was nebulized to produce particles of about 100nm size

  17. Laser Sampling: Introduction of particles /atoms into discharge by laser pulse from the sample’s back side. • Synchronized sample introduction is possible • Study of non-conductive thin films also possible, surface analysis of thin films (brain slice) • Sampling yield improves • Back side sampling (LIFT or LIBA) has the advantage of sampling directionality

  18. Laser Induced Back Ablation Laser mirror Lenz Sample on a quartz window

  19. Pulsed Laser Induced Forward Transfer Laser

  20. Summary Glow discharge time of flight mass spectrometry can be used for surface and core chemical characterization of particles Cathode of the source needs to be modified for sample introduction An orifice in the cathode seems to be a proper way of particle introduction Laser sampling can be used for synchronized introduction of particles and also analysis of thin films (useful for non-conductive samples)

  21. References: • Eric Oxley, “The Microsecond Pulsed Glow Discharge: Developments in Time of Flight Mass Spectrometry and Atomic Emission Spectrometry.”, Ph.D. Dissertation, Department of Chemistry, University of Florida, 2002. • W.W. Harrison, C. Yang and E. Oxley, Anal. Chem., 2001, 73, 480A-487A • C.L. Lewis, M.A. Moser, D.E. Dale, Jr., W. Hang, C. Hassel, F.L. King and V. Majidi, Anal. Chem., 2003, 75, 1983-1996 • A. B. Bullock, P. R. Bolton, Appl. Phys., 1999, 85, No. 1, 460-465

  22. Acknowledgments: • Dr Winefordner • Dr Smith • Dr Kevin Turnery • Dr Omenetto • Dr Harrison • Dr Igor Gornushkin • Committee members • Winefordner-Omenetto- Smith group • Friends from Yost / Martin group • Machine shop • Electronic Shop