The procedure for noise rate measurement of BBC’s PMTs

The procedure for noise rate measurement of BBC’s PMTs. Hiroyuki Harada Hiroshima University Mail : harada@hepl.hiroshima-u.ac.jp. Purpose of this measurement.

liza
Download Presentation

The procedure for noise rate measurement of BBC’s PMTs

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. The procedure for noise rate measurement of BBC’s PMTs Hiroyuki Harada Hiroshima University Mail : harada@hepl.hiroshima-u.ac.jp Hiroyuki Harada

  2. Purpose of this measurement • The purpose of this measurement is to understand the condition of PMTs, which good or bad, by measuring and checking noise rate of PMTs. Thus, For comparing condition of PMTs, we must take a data of noise rate under same assumption. Hiroyuki Harada

  3. Equipment for set up of the measurement • Pocket pulsar (for rate of decrease by the cycle) • Some cables • Discriminator • Fan in/out • Oscilloscope • Scalar Hiroyuki Harada

  4. The cycle of general measurement for PMT’s noise rates Only this part is essentially different in case of measurement in IR. To next page. • For all measurement is the same conditions, the discriminator’s threshold values, which 30, 60, 200 and 1000 mV, must be set to a value shortly after PMTs output the signals. • After HV was turned on, we must wait for measurement more than 5 minutes for warming PMTs. Fan in/out Discriminator Scalar PMT oscilloscope The wave form for PMT’s signals was taken a photo by oscilloscope. Hiroyuki Harada

  5. The measurement for it in IR • Rate of decrease for signal by cable (PMT-before FEM) is 20%. Thus the discriminator’s threshold values, which 30, 60, 200, and 1000 mV, must be set more lower value than them of general measurement. • Air Flow -> On • HV of all channels -> On Hiroyuki Harada

  6. Standard of good or bad PMTs Scale -> 10mVΩ, 4ns, Threshold -> -10mV, Participant -> 10s Good bad Broad pulse? Discharge pulse? Raise time about 4ns? High rates about 1000 Hz on Th60mV? Hiroyuki Harada

  7. Real data http://www.hepl.hiroshima-u.ac.jp/phx/bnl/pmt/All_PMTList.xls Hiroyuki Harada

  8. Final check list • Did you measure under rate of decrease of cables? • Did you measure the same condition, which Air Flow – on and HV of all channels – on, in case at IR? • Did you take a respectively photo by oscilloscope? • Did you update the data xls file on back page? Then, Did you link the photo? • Did you compare with a data before measurement? Hiroyuki Harada