Motion Correction on Dual ECG and Respiratory Gated 3D Cardiac PET/CT Data

Motion Correction on Dual ECG and Respiratory Gated 3D Cardiac PET/CT Data. Mohammad Dawood 1,3,4 , T Kösters 2 , M Fieseler 1,3 , F Büther 1,4 , X Jiang 3 , KP Schäfers 1,4 1 European Institute of Molecular Imaging, University of Münster

loman
Download Presentation

Motion Correction on Dual ECG and Respiratory Gated 3D Cardiac PET/CT Data

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Motion Correction on Dual ECG and Respiratory Gated 3D Cardiac PET/CT Data Mohammad Dawood1,3,4, T Kösters2, M Fieseler1,3, F Büther1,4, X Jiang3, KP Schäfers1,4 1European Institute of Molecular Imaging, University of Münster 2Institute for Computational and Applied Mathematics, University of Münster 3Institute of Computer Science, University of Münster 4University Hospital Münster

  2. Problem: Motion during data acquisition PET, Minutes CT, Seconds

  3. Problem:1. Attenuation correction CT: Static PET: Respiratory phases Expiration Inspiration

  4. Problem: 2. Image blur Extent of motion Superimposed image

  5. Clinical relevance • Motion artifacts lead to errors: • wrong „staging“ of the tumors (Osman et al 2003, Erdi et al 2004) • „Uptake Values“ are incorrect (Nakamoto et al 2004) • small tumors may remain undetected (Papathanassiou et al 2005) • Solution? • Gating + Motion Correction

  6. Method: Amplitude- vs. Time-based gating P1 Amplitude based P1 Time based P2 Amplitude based P2 Time based

  7. Solution: Part 2 - Optical flow Brightness consistency constraint I(x,y,z,t) = I(x+∂x,y+∂y,z+∂z,t+∂t) I.V = -It I = Grey value V = Flow x,y,z,t = Position

  8. Local consistency Solution: Part 2 - Optical flow Lucas/Kanade: Optical flow is locally consistent Ix1Vx+Iy1Vy+Iz1Vz = -It1 . . . . . . . . . . . . IxnVx+IynVy+IznVz = -Itn

  9. Global smoothness Brightness consistency Solution: Part 2 - Optical flow Horn/Schunck: Optical flow is globally smooth ∫((I.V+It)+(|u|2+|v|2+|w|2))dxdydz

  10. Local consistency Global smoothness Ψ2 (w,s) Discontinuity preservation Solution: Part 2 - Optical flow Lucas/Kanade + Horn/Schunck + Discontinuity Preservation ∫ (ψ1(I.V+It)+αψ2(w,|u|2+|v|2+|w|2)) dxdydz Dawood et al, IEEE Trans Med Imaging, August 2008

  11. Results: Visual PET-Data before correction -8 gates Remaining Motion ~ 0.3 mm << 1 Voxel PET-Data after correction -8 gates

  12. Respiratory motion, 6 gates Combined motion, 36 gates Cardiac contraction, 6 gates Dual ecg and respiratory gating and motion correction +

  13. Combined motion, 36 gates After motion correction, 36 gates Results: Visual Combined motion, 36 gates, motion corrected

  14. Data-based motion correction and reconstruction Data-based approach outperfoms the image-based motion correction Line of Response Lamare et al, Phys Med Bio, 2007

  15. Attenuation correction: PET-derived dynamic CT-based mu-map Dynamic CT-based mu-maps derived from a static mu-map and motion information from the PET

  16. Target Gate Ungated Motion Corrected Dual ecg and respiratory gating and motion correction

  17. Results: Myocardial thickness One gate 12,5 mm Before motion correction 17,2 mm After motion correction 12,5 mm

  18. Results: Myocardial thickness

  19. Thank you!