Irradiation in Karlsruhe. The Karlsruhe-Probestations Procedure of Irradiation Results Future annealing studies. The Probestation. Cold measurements. The Software. Procedure. Measurements of sensors @ 20°C @ rel. humidity < 30% IV & CV Ileak Rpoly (2V) CC (100Hz)
E N D
Irradiation in Karlsruhe The Karlsruhe-Probestations Procedure of Irradiation Results Future annealing studies Irradiation Workshop Karlsruhe
The Probestation Irradiation Workshop Karlsruhe
Cold measurements Irradiation Workshop Karlsruhe
The Software Irradiation Workshop Karlsruhe
Procedure • Measurements of sensors • @ 20°C • @ rel. humidity < 30% • IV & CV • Ileak • Rpoly (2V) • CC (100Hz) • Pinhole (10V) • Rint (0-0,5V) >10G • Cint (1MHz) Irradiation Workshop Karlsruhe
Procedure • Bias connections for irradiation (conductive rubber & bonds) • Irradiation • 34 MeV-protons • Fluence • Annealing • @ 60 °C & 80 min • Mistake: annealing of sensors in the envelopes • Same measurements as before irradiation, but • @ -10°C & dew point << -10°C (39°C) Irradiation Workshop Karlsruhe
Alpha @ -10°C From literature Irradiation Workshop Karlsruhe
Leakage current before and after irradiation • Guardring connected (for defined Volume) • Increase of leakage current as expected • Breakdown voltage >1200V Irradiation Workshop Karlsruhe
Depletion voltage before and after irradiation • Annealed @ 60°C and 80 min • But in the envelopes ! Irradiation Workshop Karlsruhe
Leakage current after Irradiation Increase of leakage current depends on fluence as aspected Low Fluence High Fluence Irradiation Workshop Karlsruhe
Rpoly after irradiation 100V 0V 1V No dependence on bias conditions ! Irradiation Workshop Karlsruhe
CC after Irradiation 0V 100V 1V No dependence on bias conditions ! Irradiation Workshop Karlsruhe
Cint after irradiation No increase of Cint seen 1V Irradiation Workshop Karlsruhe
IQC-ramps Irradiation Workshop Karlsruhe
Annealing curves Irradiation Workshop Karlsruhe
Annealing @ 50°C ? • Problem of annealing studies @ 60 °C • high stability of temperature needed • Effects of heating up and cooling down become important • Problem of annealing studies @ 40 °C • long time constants Irradiation Workshop Karlsruhe
Time constants @ 50°C • Fit from values in the Moll-Thesis: • Beneficial annealing: 65 min • Reverse annealing: 5162 min Annealing steps: 15 min ? Irradiation Workshop Karlsruhe
Conclusions • Measurements @ -10°C after irradiation • Annealing without envelopes ! • No dependence of stripvalues from bias conditions • Next irradiations without bias ? Irradiation Workshop Karlsruhe