HAMILTON Software VENUS

HAMILTON Software VENUS. VENUS 96 Head Strategies. HAMILTON Software VENUS. Using the 96 Head to pipette less than 96 samples. 2 Strategies “Potato printing” – removing some tips then using the remaining ones Shifted tip pick up – picking up tips with the head over only part of the rack.

lynley
Download Presentation

HAMILTON Software VENUS

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. HAMILTON Software VENUS VENUS96 Head Strategies

  2. HAMILTON Software VENUS Using the 96 Head to pipette less than 96 samples • 2 Strategies • “Potato printing” – removing some tips then using the remaining ones • Shifted tip pick up – picking up tips with the head over only part of the rack

  3. HAMILTON Software VENUS Potato Printing • Create a rack that contains just the tips you want to use • Pickup some tips with the channels and throw them away • Pickup tips with the channels and use them (e.g. to add controls) then throw them away • Pick up tips with the channels and put them in an empty rack Details

  4. HAMILTON Software VENUS Potato Printing • Pick up the remaining tips with the 96 head This sequence must include the whole rack of tips Sample Tracking

  5. HAMILTON Software VENUS Shifted Tip Pickup • There must be space around the rack For clear tip racks, you will need a tip support (p/n 182040) Nestable tips should be placed on an NTR1 module (p/n 191420)

  6. HAMILTON Software VENUS Shifted Tip Pickup • Ensure the tip sequence is correctly sorted To pick up columns, sequence sorted in columns To pick up rows, sequence sorted in rows Either order will work for single tips

  7. HAMILTON Software VENUS Shifted Tip Pickup • First, move the tips to the correct position Use a CORE 96 Head Tip Pick Up (single step) to pick up the tips from a standard rack and the Tip Eject step to put them in the support Move your nestable tip rack from a stack to this position (CORE grippers or iSWAP)

  8. HAMILTON Software VENUS Shifted Tip Pickup • Picking up one or more columns of tips Mode is (3) Columns(s) Select the number of columns here – this is view of the head from above Ignore this – it is incorrect

  9. HAMILTON Software VENUS Shifted Tip Pickup • Picking up one or more columns of tips • The channel pattern in the previous image will result in this pickup Take care that there is nothing here that the 96 head will hit tip rack 96 head Multiple Cycles

  10. HAMILTON Software VENUS Shifted Tip Pickup • Picking up one or more rows of tips Tip sequence must be sorted in rows

  11. HAMILTON Software VENUS Shifted Tip Pickup • Picking up one tip Choose which corner of the head