1 / 10

HPLC and UHPLC Part 1

The presentation discusses a set of findings from the 1990's that can affect the stability of RPLC columns at a high

macmod
Download Presentation

HPLC and UHPLC Part 1

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. HPLC and UHPLC Separations at Mid to High pH: Part 1

  2. Introduction • Reviewed findings published by Kirkland, Claessens and co-workers in the 1990s regarding the factors that can affect the stability, performance, and lifetime of silica-based RPLC columns at mid and high pH • Many experts have recommended that reversed-phase method development begin using low pH mobile bases

  3. Introduction • Many basic compounds • carry a full or partial positive charge at low pH, • will often be poorly retained, • have poor peak shape, or • may even be excluded (elute before void) on most modern type B silica based packings • It is an advantage to be able to use mobile phases with pHs at or above 7.0 for basic compounds

  4. Advantages of Working at Mid or High pH • Some analytes are insoluble or insufficiently soluble at low pH • Some analytes are unstable at low pH • Some analytes are not adequately retained or cannot be separated from closely eluting compounds at low pH • Improvement in response and signal-to-noise ratio is needed for analysis of basic analytes by LC-MS vs low pH

  5. Advantages of Working at Mid or High pH • Analytes are not adequately retained or cannot be separated from closely eluting compounds at low pH • Can observe improvement in LC-MS response and signal-to-noise ratio for basic analytes at mid- or high pH vs. low pH

  6. Column Stability • Degradation of silica-based columns at intermediate or high pH • largely a function of dissolution of the silica support • Rather than a loss of bonded phase due to hydrolysis (cause of column degradation at low pH)

  7. Column Stability • Rate of silica degradation due to dissolution is affected by: • Choice of mobile phase organic modifier • Type of silica support • Type and bonding density of bonded phase • Type and concentration of buffer salts in mobile phase (NH4+, Na+, K+; PO4-3, CO3-2, acetate, formate, etc.) • Column temperature • Silica support solubility in the pH 6-8 range is greatly increased in the presence of phosphate and carbonate buffers

  8. Column Stability • Silica-based columns are rapidly degraded when carbonate and phosphate buffers are used at pH 10 • Silica dissolution for untreated silica at pH 9-10 was reduced in high concentrations of organic modifier

  9. Column Stability • Precolumns of unmodified silica greatly extend the lifetime of columns used at high pH • Bonded-phase precolumns were much less effective

  10. Column Stability • Longer chain bonded phases are more stable than shorter chain phases • Certain silica-based, bonded-phase columns can be used routinely for long periods to at least pH 9-10 http://www.mac-mod.com

More Related