Review of Analytical Methods Part 2: Electrochemistry. Roger L. Bertholf, Ph.D. Associate Professor of Pathology Chief of Clinical Chemistry & Toxicology University of Florida Health Science Center/Jacksonville. Analytical methods used in clinical chemistry. Spectrophotometry
E N D
Review of Analytical MethodsPart 2: Electrochemistry Roger L. Bertholf, Ph.D. Associate Professor of Pathology Chief of Clinical Chemistry & Toxicology University of Florida Health Science Center/Jacksonville
Analytical methods used in clinical chemistry • Spectrophotometry • Electrochemistry • Immunochemistry • Other • Osmometry • Chromatography • Electrophoresis
Electrochemistry • Electrochemistry applies to the movement of electrons from one compound to another • The donor of electrons is oxidized • The recipient of electrons is reduced • The direction of flow of electrons from one compound to another is determined by the electrochemical potential
Electrochemical potential • Factors that affect electrochemical potential: • Distance/shielding from nucleus • Filled/partially filled orbitals
Relative potential Zn Cu • Copper is more electronegative than Zinc • When the two metals are connected electrically, current (electrons) will flow spontaneously from Zinc to Copper • Zinc is oxidized; Copper is reduced • Zinc is the anode; Copper is the cathode e-
mV Zn0 Cu0 Zn2+ Cu2+ e- e- e- Zn0 Zn2+ + 2e- Cu2+ + 2e- Cu0
The Nernst Equation Where E = Potential at temperature T E0 = Standard electrode potential (25ºC, 1.0M) R = Ideal gas constant F = Faraday’s constant n = number of electrons transferred
mV Zn0 Cu0 Zn2+ Cu2+ Elj Zn0 Zn2+ + 2e- E0= +(-)0.7628 V Cu2+ + 2e- Cu0 E0 = +0.3402 V
Electromotive force Ecell = Ecathode + Elj - Eanode Ecell = ECu(II),Cu + Elj – EZn(II),Zn Ecell = (+)0.340 + Elj – (-)0.763 Ecell = (+)1.103 + Elj G = -nFEcell
Would the reaction occur in the opposite direction? Ecell = Ecathode + Elj - Eanode Ecell = EZn(II)Zn + Elj – ECu(II) Cu Ecell = (-)0.763 + Elj – (+)0.340 Ecell = (-)1.103 + Elj
How do we determine standard electrode potentials? • Absolute potential cannot be measured—only the relative potential can be measured • Standard electrode potentials are measured relative to a Reference Electrode • A Reference Electrode should be. . . • Easy to manufacture • Stable
mV Test electrode The Hydrogen Electrode H2 gas 2H+ + 2e- H2 E0 = 0.0 V
mV Test electrode Calomel paste (Hg0/Hg2Cl2) Saturated KCl Liquid junction The Calomel Electrode Hg2Cl2 + 2e- 2Hg0 + 2Cl- E0 = 0.268V
mV Test electrode Silver wire Saturated KCl + AgNO3 Liquid junction The Silver/Silver Chloride Electrode AgCl + e- Ag0 + Cl- E0 = 0.222V
mV Ion-selective Electrodes Ecell = ERef(1) + Elj – ERef(2) Ref1 Ref2
mV Ecell EISM Ref 1 Ref 2 + + + + + + + Ion-selective membrane + + + + Typical ISE design Ecell = ERef(1) + Elj – ERef(2)
Activity and concentration • ISEs do not measure the concentration of an analyte, they measure its activity. • Ionic activity has a specific thermodynamic definition, but for most purposes, it can be regarded as the concentration of free ion in solution. • The activity of an ion is the concentration times the activity coefficient, usually designated by :
The activity coefficient • Solutions (and gases) in which none of the components interact are called ideal, and have specific, predictable properties • Deviations from ideal behavior account for the difference between concentration and activity • Dilute solutions exhibit nearly ideal behavior (1)
Types of ISE • Glass • Various combinations of SiO2 with metal oxides • Solid-state • Involve ionic reaction with a crystalline (or crystal doped) membrane (example: Cl-/AgCl) • Liquid ion-exchange • A carrier compound is dissolved in an inert matrix • Gas sensors • Usually a combination of ISE and gas-permeable membrane
mV pH electrode Shielded connecting cable Non-conducting glass body External reference electrode Internal reference electrode H+-responsive glass membrane
mV NaHCO3/H2O External reference electrode CO2(g) pCO2 electrode Electrode assembly Gas-permeable membrane (silicone rubber) CO2 + H2O HCO3- + H+ Flow Cell
mV NH4Cl/H2O External reference electrode NH3(g) NH3 electrode Electrode assembly Gas-permeable membrane (PTFE) H2O + NH3 NH4+ + OH- Flow Cell
Other glass electrodes • Glass electrodes are used to measure Na+ • There is some degree of cross-reactivity between H+ and Na+ • There are glass electrodes for K+ and NH4+, but these are less useful than other electrode types
mV Lipids, proteins Na+ Plasma Aqueous phase Cells (45%) Whole blood The Sodium Error(or, direct vs. indirect potentiometry) Since potentiometry measures the activity of the ion at the electrode surface, the measurement is independent of the volume of sample.
mV Na+ The Sodium Error(or, direct vs. indirect potentiometry) In indirect potentiometry, the concentration of ion is diluted to an activity near unity. Since the concentration will take into account the original volume and dilution factor, any excluded volume (lipids, proteins) introduces an error, which usually is insignificant.
So which is better? • Direct potentiometry gives the true, physiologically active sodium concentration. • However, the reference method for sodium is atomic emission, which measures the total concentration, not the activity, and indirect potentiometry methods are calibrated to agree with AE. • So, to avoid confusion, direct potentiometric methods ordinarily adjust the result to agree with indirect potentiometric (or AE) methods.
Then what’s the “sodium error” all about? • When a specimen contains very large amounts of lipid or protein, the dilutional error in indirect potentiometric methods can become significant. • Hyperlipidemia and hyperproteinemia can result in a pseudo-hyponatremia by indirect potentiometry. • Direct potentiometry will reveal the true sodium concentration (activity).
Na+ 140 mM Na+ 138 mM Na+ 130 mM Na+ 140 mM Sodium error
But. . .why does it only affect sodium? • It doesn’t only affect sodium. It effects any exclusively aqueous component of blood. • The error is more apparent for sodium because the physiological range is so narrow.
Solid state chloride electrode • AgCl and Ag2S are pressed into a pellet that forms the liquid junction (ISE membrane) • Cl- ions diffuse into vacancies in the crystal lattice, and change the membrane conductivity
Liquid/polymer membrane electrodes • Typically involves an ionophore dissolved in a water-insoluble, viscous solvent • Sometimes called ion-exchange membrane electrodes • The ionophore determines the specificity of the electrode
K+ K+ ion-selective electrode Valinomycin is an antibiotic that has a rigid 3-D structure containing pores with dimensions very close to the un-hydrated radius of the potassium ion. Valinomycin serves as a neutral carrier for K+.
di-p-octylphenyl phosphate Ca++ ion selective electrode PVC membrane Ca++
Neutral carrier Ca++ ion selective electrode Inert membrane Ca++
Amperometry • Whereas potentiometric methods measure electrochemical potential, amperometric methods measure the flow of electrical current • Potential (or voltage) is the driving force behind current flow • Current is the amount of electrical flow (electrons) produced in response to an electrical potential
Amperometry Limiting current Current (mA) Half-wave potential Applied potential (V)
2•C0 C0 0.5•C0 Amperometry Current (mA) Half-wave potential Applied potential (V)
-0.65V Reference electrode (anode) Platinum wire (cathode) Gas-permeable membrane Flow cell O2 Oxygen (pO2) electrode
Reaction at the platinum electrode • The amount of current (e-) is proportional to the concentration of O2
Glucose H2O2 + Gluconic acid oxidase 2e- O2 electrode (+0.6 V) O2 + 2H- The glucose electrode • Note that the platinum electrode now carries a positive potential Glucose + O2