An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Filter Set Selection

  2. Filter Set Selection • It is best to follow this presentation with a Fluorophore excitation/emission maxima chart to hand.

  3. Preliminary • The MRC series has two dichroics • They are numbered 1and2

  4. Preliminary • Dichroic number 1 is to the rear of the scan head • It filters the excitation beam • They are identified with names such as K1 or I1 etc. Also VHS/BHS

  5. Preliminary • Dichroic number 2 is to the front of the scan head • It filters the emissions from the sample • It is identified with a name such as K2 or I2 etc.

  6. The MRC1000/1024 • The MRC1000/1024 has extra filters • They have excitation filters and emission filters as well as dichroics • The dichroics are changed manually • The extra filters are computer controlled

  7. The Radiance • The Radiance has similar filters to the MRC1000/1024 series • They are all computer controlled. • No manual filter changes are necessary • Instead of changing dichroics, dichroic filter wheels are rotated by computer control.

  8. To select the Correct Filter Set • First identify the fluorescent dye used in your sample

  9. Spectra • The excitation and emission wavelengths of fluorophores are not single wavelengths • They have spectra

  10. An example. Imaging CY3

  11. An example. Imaging CY3 • From the spectra, the useful excitation and emission ranges are seen to be: • Excitation = 470 - 580 nm • maximum at 550 nm • Emission = 540 - 660 nm • maximum at 565 nm

  12. An example. Imaging CY3 • To successfully image CY3, first look to the left hand side of the chart. • Locate CY3 on the chart.

  13. Excitation Filter Selection • This shows the excitation maximum

  14. Excitation Filter Selection • Identify the laser line which is closest to and (if possible) has a shorter wavelength than the dye excitation maximum.

  15. Excitation Filter Selection • Project the chosen laser line horizontally across the chart to the RIGHT hand side

  16. Excitation Filter Selection • Note the dichroic number 1 it intercepts. • This determines which laser excitation filter and dichroic number 1 are best

  17. Emission Filter Selection • From the excitation maximum, follow the solid white line on the chart from left to right • This will identify your emission maximum

  18. Emission Filter Selection • Continue the line horizontally to the right of the chart • Note the emission filter and the dichroic number 2 it intercepts

  19. Emission Filter Selection • Ensure the emission filter is NOT intercepted by the chosen laser line

  20. Emission Filter Selection • If it is, you must choose an emission filter which passes longer wavelengths • Or a laser line which has a shorter wavelength • Or both

  21. Conclusion • Once these conditions are met, you have successfully chosen the best filter set for your sample