Matrix Modifiers. Used For Furnace Analyses. MATRIX MODIFIERS. Help prevent loss of analyte in ashing stage They alter the behaviour of analyte or matrix May: - Render analyte less volatile - Help volatilise matrix - Convert analyte to single compound.
E N D
Matrix Modifiers Used For Furnace Analyses
MATRIX MODIFIERS • Help prevent loss of analyte in ashing stage • They alter the behaviour of analyte or matrix • May: • - Render analyte less volatile • - Help volatilise matrix • - Convert analyte to single compound
MATRIX MODIFIERS1) Render analyte less volatile e.g. (a) Pb Cl2 + (NH4)2HPO4 +→PbHPO4 + 2NH4Cl Permits ashing to 850C PbCl2 would be lost above 450
e.g. (b) Palladium • Add Pd salt to sample soln • Pd on heating • Cd + Pd Cd-Pd Also for Pb, As • Many elements form stable alloys or other compounds with Pd Can ash at a higher temp • Can also add reagents to help reduce Pd salt Pd metal
MATRIX MODIFIERS2) Help volatilize Interfering Matrix e.g. high [NaCl] Causes severe interference & excessive background absorption NaCl boils at 1400C - would lose some analyte
To solve problem: • Add excess NH4 NO3 • NaCl + NH4NO3 NH4Cl + NaNO3 • Both products volatilize at lower temps than NaCl
3) Convert analyte to single compound before Atomisation • e.g. • PbCl2 + Pb(NO)2 +(NH4)2HPO4 → PbHPO4 + salts Conversion During Ashing • Benefit - More Uniform • Volatilisation