An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Introduction to Optical Atomic spectrometry 2004. 4 Yongsik Lee

  2. introduction • Three methods for atomic spectrometry • Optical spectroscopy • Mass spectrometry • X-ray spectroscopy

  3. Preparation of atomic samples • In optical methods • Sample is atomized • gas phase atoms or ions • Emission/absorption/luminescence • In mass spectrometry • Cation and m/e measured • X-ray • Atomic intrinsic data • Direct measurement

  4. 8A Optical Atomic Spectra • Energy level diagram • Atomic Line width • Thermal effect • Band/continuous spectrum

  5. Atomic Energy levels • Energy levels • Every elements has unique set of atomic orbital levels • Hydrogen atom has a simple energy levels • One proton + one electron (two body problem) • Other elements have complex atomic orbital (p,d,f...) levels • Levels split by spin-orbit (SO) coupling • Energy level diagram • A picture of relative energy levels of an atom

  6. Energy levels and atomic spectra

  7. 복잡한 “커플링”은 에너지 준위를 degenerate

  8. Spin-orbit coupling • Spin (s) and orbital (l) motion create magnetic fields that perturb each other • if fields parallel - slightly higher energy • if fields anti-parallel - slightly lower energy • Define Spin-Orbit coupling by J • total angular momentum • J=L+S (L = Sl , S = Ss) • positive values only

  9. Vector sum of angular momenta • P.W. Atkins, • Physical Chemistry • Atomic structure and atomic spectra • s electron (l=0, s=+1/2 or -1/2) • J=0+1/2=1/2 • p electron (l=1, s=+1/2 or -1/2) • J= 1+1/2 = 3/2 (higher energy) or • J= 1-1/2 = 1/2 (lower energy)

  10. Electronic Term Symbol • 2S+1LJ • L written as letter (S, P, D...) instead of number! • Li =1s2 2s1 L = 0, S = ±1/2 2S1/2 • Li* =1s2 2p1 L = 1, S = ±1/2 2P1/2, 3/2 • Be =1s2 2s2 L = 0, S = 0 1S0 • Be* =1s2 2s12p1 L =1, S = 1, 0 3P2 , 1P0

  11. Sodium D lines http://radio.weblogs.com/0101365/categories/periodic/2002/11/01.html

  12. Energy Level Diagram with SO coupling

  13. Allowed/forbidden transition • Similar pattern between atoms but different spacing • Spectrum of ion different to atom • Separations measured in electron volts (eV) • Figure 8-2 • Electron volt • Energy unit • 1 eV =1.602x10-19 C X 1V (J/C) = 1.602x10-19 J • 1 eV = 96. 484 kJ/mol • As # of electrons increases • # of levels increases • Emission spectra become more complex • Using selection rule (Quantum mechanics) – allowed/forbidden • Li 30 lines, Cs 645 lines, Cr 2277 lines

  14. Review of atomic spectrometry • Three methods for atomic spectrometry • Optical spectroscopy • Mass spectrometry • X-ray spectroscopy

  15. Three Types of Atomic optical spectroscopy • Atomic emission • Arc or spark (heating) • H2-O2 flame • Atomic absorption • Resonance absorption lines • Atomic fluorescence • Resonance fluorescence • Non-resonance fluorescence

  16. Atomic line widths • Desire narrow lines for accurate identification • Broadened by 4 mechanism • (i) uncertainty principle • Natural line width (about 1/10000 Å) • (ii) Doppler effect (about 1/100 Å) • (iii) pressure broadening (about 1/100-1/10 Å) • (iv) electric and magnetic fields • Zeeman effect • Chapter 9C-1

  17. Uncertainty Principle • Why? • Quantum mechanical idea states must measure for some minimum time to tell two frequencies apart • How? • Shows up in lifetime of excited state • if lifetime infinitely long, DE infinitely narrow • if lifetime short, DE is broadened

  18. Calculation of uncertainty broadening

  19. Pressure broadening • Why? • Collisions with atoms/molecules transfers small quantities of vibrational energy (heat) • ill-defined ground state energy • Effect worse at high pressures • For low pressure hollow cathode lamps • 1-10 torr • Dl = 0.1 – 0.01 Å • For high pressure Xe lamps • >10,000 torr • 100-1000 Å (turnslines into continua!)

  20. Doppler broadening • Change in frequency produced by motion relative to detector • In a flame, Doppler broadening is 100 times of natural broadening • Red shift is observed by Edwin Hubble (1920) • Big bang theory

  21. Compression/expansion

  22. Doppler broadening of Gas • In gas, broadened line is symmetric shape • Average speed of gas atoms • Doppler effect to the detector • Maxwell-Boltzmann Distribution of speed • Super sonic jet nozzle (molecular beam) • Doppler broadening increases with (T)1/2

  23. Sub-millimeter supersonic jet Spectrometer at Texas A&M (Instrumental line width <2 kHz, 6.8 x 10-8 cm-1) http://www.chem.tamu.edu/rgroup/bevan/activity.html

  24. Atomic linewidths • Broadened by 4 mechanism • (i) uncertainty principle • Natural line width (about 1/10000 Å) • (ii) Doppler effect • Flames - about 1/100 Å • (iii) pressure broadening • about 1/100-1/10 Å • (iv) electric and magnetic fields • Zeeman effect • Chapter 9C-1

  25. Other Effects of Temperature on Atomic Spectrometry • Boltzmann equation • Important in emission measurements • relying on thermal excitation • Na atoms at 2500 K, only 0.02 % atoms in first excited state! • Less important in absorption measurements • 99.98 % atoms in ground state!

  26. Best method? • ABS/FL • More sample atoms than Emission • Absorption • Difference measurement (log Po – log P) • Big Background and big signal • Cancels out the advantage • FL • For the same number of atoms, FL is best

  27. 8B Methods for atomizing • Sample must be converted to atoms first

  28. 8C Sample Introduction method • transfer sample to atomizer • Achilles heel of atomic spectroscopy • easy for gases /solutions • but difficult for solids

  29. Sample phase and introduction methods • Solution • Or slurry (suspension) • Solid • Solid or fine powder • Less producible • More error

  30. Solution phase • pnuematic nebulization • Sample in aq solution • Figure 8-9 • Bernoulli aspiration • Supersonic nebulization • 20 kHz – MHz peizoelectric surface • Electrothermal vaporizer (ETV) • Non-continuous signal • Hydride generation

  31. Hydride generation • For As, Sb, Sn, Se, Bi, Pb containing samples to vaporize • Detection limit 10-100 times increase • Some of them are toxic! • Generation method • Prepare acidic sample solution • Add to 1% NaBH4 solution • Metal hydride (g) generated

  32. Solid sample • Direct insertion • Powder on the probe • Metal samples as an electrode • Electothermal vaporization • Spark and arc ablation • Vaporized samples moved by inert gas • Conductive (or mixture of) samples • Laser ablation • Various kinds of sample • Glow discharge (GD) sputtering • Figure 8-10 • Introduction and atomization at the same time • Ar+ ion collision (1-10 torr Ar used for 250-1000 VDC) • Samples on cathode (which contains the sample)

  33. Homework • 8-8, 9, 10