PolarisCdZnTe Gamma-Ray Imaging Spectrometer Systems On behalf of the Orion group Zhong He ISOE International ALARA Symposium Tokyo, Japan, 27-29 August, 2013
Polaris Systems Eighteen 221.5 cm3 CdZnTe detectors (108 cm3, 648 grams = 1.43 lb) Performance GoalsE/E 1% FWHM (at 662 keV)Real-time Imaging + isotope I.D. E2 E1 Number of photons: 18 97 2033 3 4 1 2
137Cs Energy Spectra of the 1stPolaris system #1.1(From all 18 detectors of Polaris, 24oC, uncollimated 137Cs) • 1-pixel events • 2-pixel events • 3-pixel events • 4-pixel events 662 keV 1.44% FWHM All events Counts Energy (keV)
2nd-GenerationPolaris System #2.1 (BNL ASIC)(From all 18 detectors, room-temperature, uncollimated 137Cs) 662 keV Single-pixel events 0.61% (4.0 keV) FWHM
2nd-GenerationPolaris System #2.1 (BNL ASIC)(From all 18 detectors, room-temperature, uncollimated 137Cs) 662 keV All events – no selection July 2012 0.86% (5.7 keV) FWHM Peak-to-Compton ratio 20.5
Next-Generationdigital3-D CdZnTe detectors (VAD_UM digital ASIC, Room-temperature, uncollimated 137Cs) 662 keV Energy resolutions (FWHM) of all events measured on best detectors 0.46% FWHM Counts E/E is stilllimited by ASIC noise ! Peak-to-Compton ratio 22.5 ( 25.1 if no event selection) Energy (keV)
228Th Energy Spectra Polaris-2 D.E. 2614keV S.E. Counts Energy (keV)
Yuefeng Zu Willy Kaye He Weiyi Wang 22Na 60Co Chris Wahl 137Cs 137Cs Target specific -Spec. Polar Angle 23 min. data EIID Azimuthal Angle
Real-Time Combined Coded Aperture and Compton Imaging for Locating Sources From all18 detectors Isotope selected: 57Co,22Na,137Cs 122 keV 662 keV 511 keV 1275 keV
Detection Sensitivity of Polaris Image of 650 E 674 keV SBP with 2 – 4 pixel events 208Tl 583.2 Source Present 214Bi 609.3 • Background • 137Cs present A 30 Ci Cs-137 source behind a wall at ~4m 40K 1461 A 12-hour “background” Measurement All events