An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. PolarisCdZnTe Gamma-Ray Imaging Spectrometer Systems On behalf of the Orion group Zhong He ISOE International ALARA Symposium Tokyo, Japan, 27-29 August, 2013

  2. Polaris Systems Eighteen 221.5 cm3 CdZnTe detectors (108 cm3, 648 grams = 1.43 lb)  Performance GoalsE/E  1% FWHM (at 662 keV)Real-time  Imaging + isotope I.D.   E2 E1 Number of photons: 18 97 2033 3 4 1 2

  3. Polaris 1.1 (GMI ASIC) – August 2010

  4. 137Cs Energy Spectra of the 1stPolaris system #1.1(From all 18 detectors of Polaris, 24oC, uncollimated 137Cs) • 1-pixel events • 2-pixel events • 3-pixel events • 4-pixel events 662 keV 1.44% FWHM All events Counts Energy (keV)

  5. 2nd-GenerationPolaris System #2.1 (BNL ASIC)(From all 18 detectors, room-temperature, uncollimated 137Cs) 662 keV Single-pixel events 0.61% (4.0 keV) FWHM

  6. 2nd-GenerationPolaris System #2.1 (BNL ASIC)(From all 18 detectors, room-temperature, uncollimated 137Cs) 662 keV All events – no selection July 2012 0.86% (5.7 keV) FWHM Peak-to-Compton ratio 20.5

  7. Next-Generationdigital3-D CdZnTe detectors (VAD_UM digital ASIC, Room-temperature, uncollimated 137Cs) 662 keV Energy resolutions (FWHM) of all events measured on best detectors 0.46% FWHM Counts E/E is stilllimited by ASIC noise ! Peak-to-Compton ratio 22.5 ( 25.1 if no event selection) Energy (keV)

  8. 228Th Energy Spectra Polaris-2 D.E. 2614keV S.E. Counts Energy (keV)

  9. Yuefeng Zu Willy Kaye He Weiyi Wang 22Na 60Co Chris Wahl 137Cs 137Cs Target specific -Spec. Polar Angle 23 min. data EIID Azimuthal Angle

  10. Real-Time Combined Coded Aperture and Compton Imaging for Locating Sources From all18 detectors Isotope selected: 57Co,22Na,137Cs 122 keV 662 keV 511 keV 1275 keV

  11. Detection Sensitivity of Polaris Image of 650  E  674 keV SBP with 2 – 4 pixel events 208Tl 583.2 Source Present 214Bi 609.3 • Background • 137Cs present A 30 Ci Cs-137 source behind a wall at ~4m 40K 1461 A 12-hour “background” Measurement All events

  12. Today and Tomorrow