Small Animal Ventilator for Hyperpolarized Helium MRI Ashley Anderson III, Micah Brown, Matt Smith, Chris Wegener Client: Dr. Sean Fain Advisor: Dr. Willis Tompkins
Client Requirements • MRI compatible ventilator system • Deliver 20% O2 and 80% He-3 in same breath • Allows constant scanning of the He-3 in MRI rather than every 4 breaths
Background - Helium MRI • Newer technique • Same MRI scanner • Convention is H • Here, tuned to Helium (48.6 MHz) • Breath of Helium • Displays respiratory channels
Project Concept • Each syringe feeds off of independent gas reservoirs • Gas mixing occurs right before entering small animal
First Generation Prototype • Currently in use in animal studies • Only using one syringe • Not compatible with small syringes for mouse studies
Modifications to G1 • Reinforced syringe clips • Larger motor • Eliminated O2 chaser breath
Calibration Results using manometer testing. Graph shows small deviation from entered volumes.
Second Generation Prototype • Created in Solidworks • Smaller components • Interchangeable syringes • Adjustable axle support
Second Generation Prototype • Large motor • Motor shielding • Precision machining • Can be disassembled
Did we meet client requirements? • Gear ratios allow the appropriate 80/20 gas output ratio • MRI Safe materials • Motor shielded for RF • Rat/Mouse compatible
Future Work • Calibration testing • Volume delivery • Timing • Software development • Simultaneous 3He/O2 delivery • Inspired time based on %Insp and RR • Valve switching • Mixing testing (imaging)