An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Small Animal Ventilator for Hyperpolarized Helium MRI Ashley Anderson III, Micah Brown, Matt Smith, Chris Wegener Client: Dr. Sean Fain Advisor: Dr. Willis Tompkins

  2. Client Requirements • MRI compatible ventilator system • Deliver 20% O2 and 80% He-3 in same breath • Allows constant scanning of the He-3 in MRI rather than every 4 breaths

  3. Background - Helium MRI • Newer technique • Same MRI scanner • Convention is H • Here, tuned to Helium (48.6 MHz) • Breath of Helium • Displays respiratory channels

  4. Project Concept • Each syringe feeds off of independent gas reservoirs • Gas mixing occurs right before entering small animal

  5. First Generation Prototype • Currently in use in animal studies • Only using one syringe • Not compatible with small syringes for mouse studies

  6. Modifications to G1 • Reinforced syringe clips • Larger motor • Eliminated O2 chaser breath

  7. Calibration Results using manometer testing. Graph shows small deviation from entered volumes.

  8. Second Generation Concept

  9. Second Generation Prototype • Created in Solidworks • Smaller components • Interchangeable syringes • Adjustable axle support

  10. Second Generation Prototype • Large motor • Motor shielding • Precision machining • Can be disassembled

  11. From Solidworks to Reality

  12. Does it move?

  13. Did we meet client requirements? • Gear ratios allow the appropriate 80/20 gas output ratio • MRI Safe materials • Motor shielded for RF • Rat/Mouse compatible

  14. Future Work • Calibration testing • Volume delivery • Timing • Software development • Simultaneous 3He/O2 delivery • Inspired time based on %Insp and RR • Valve switching • Mixing testing (imaging)

  15. Questions?